5-7
Maintenance
Troubleshooting Common Problems
Otherwise, perform the procedure
“To check the test set” on page 5-7
.
To check the test set
W A R N I N G
Prevent any accidental exposure to the laser beam. Prior to
disconnecting any optical fibers on the TLS, ensure that the TLS is
turned off.
To accurately troubleshoot the system, make sure the TLS is working properly
prior to testing the test set.
Refer to “To check the TLS” on page 5-6
.
1
Use a microscope to check the connections on the test set.
Connections for Accurate Measurements” on page 5-20
for information on
recognizing damaged connectors.
2
Run the self test for the HP 8753E Network Analyzer.
network analyzer self test” on page 5-10
. If the self test passes, troubleshoot the
test set. If the test fails, the network analyzer needs repair
To check the computer
❒
For problems with the personal computer that are unrelated to the rest of the
HP 86037B, refer to the computer manual that was shipped with this system,
or call your local IT department.
❒
Troubleshoot the computer’s connection to the HP 86037B:
a
Make sure that the BNC cable (or a serial cable on the dual band system) is
connected to the test set properly.
b
Ensure that the GPIB and computer addresses are correct. The correct
GPIB and I/O addresses for the computer and system are located in
.
❒
Perform the following steps:
a
Disconnect the BNC cable (or a serial cable on the dual band system) from
the test set.
b
With a voltage meter, measure the voltage from the cable. It should be read-
ing between 2 and 4 volts.
Содержание 86037B
Страница 1: ...User s Guide HP 86037B Chromatic Dispersion Test System ...
Страница 31: ...2 13 Operation The Main Window Figure 2 4 Repeat Measurement Selector ...
Страница 72: ...3 16 Normalization Setting Normalization Preferences ...
Страница 94: ...4 22 Remote Control TES Parameters Main chromatic dispersion screen ...
Страница 100: ...4 28 Remote Control TES Parameters CD and Gain Measurement Settings Dialog Box ...
Страница 102: ...4 30 Remote Control TES Parameters Length Measurement Settings Dialog Box ...
Страница 104: ...4 32 Remote Control TES Parameters Normalization Settings Dialog Box ...
Страница 106: ...4 34 Remote Control TES Parameters New Normalization Button Dialog Box ...
Страница 120: ...4 48 Remote Control Output Parameters ...
Страница 158: ...5 38 Maintenance Hewlett Packard Sales and Service Offices ...
Страница 165: ...6 7 Installation Installation 2 3 4 5 1 8 7 Fan Pack 6 6 Power Distribution Assembly 1 7 Power Distribution Assembly 2 ...
Страница 170: ...6 12 Installation Installation ...
Страница 174: ...7 4 Reference Theory of Operation Figure 7 3 HP 86037B Block Diagram Options 121 and 122 ...
Страница 175: ...7 5 Reference Theory of Operation Figure 7 4 The HP 86037B Option 114 block diagram ...
Страница 190: ...7 20 Reference System Options ...
Страница 191: ...8 Specifications 8 3 Regulatory Information 8 7 Specifications and Regulatory Information ...
Страница 198: ...8 8 Specifications and Regulatory Information Regulatory Information Declaration of Conformity ...
Страница 204: ...Index 6 Index ...