2-18
HP 85106D mm-Wave System
Operation
Measurement System Calibration
Selecting the
Averaging Factor
It is also advisable to turn averaging ON; an averaging factor of 128 is
acceptable and improves the measurement results without substantially
increasing measurement time.
In the RESPONSE [
MENU
] press:
{
AVERAGING ON/restart
} [
128
] [
x1
]
Checking the
Instrument State
Verify that System Z
o
is set to 1. This is used to normalize impedance in
waveguide.
In the [CAL] {MORE} menu press:
{
SET Zo
} [
1
] [
x1
]
Entering the Exact
Offset Delay
For the 1-Port and Full 2-Port calibrations, improve accuracy by entering the
actual the exact length of the waveguide section instead of the nominal
length as provided in the standard calibration kit definition. For TRL 2-port
calibration either the actual or nominal length may be used.
NOTE
The disk file describing the calibration kit definition includes the “nominal”
length of the waveguide section used to make the quarter-wave offset short
and the quarter-wave offset load. The actual physical length of the
waveguide section is printed on the waveguide section or can be found in the
summary data sheet or calibration report. This value can be converted and
entered into the Cal Kit standard definitions thereby improving the accuracy
of the offset calibration. Please refer to the HP 11644 Series Calibration Kit
manual for instructions to enter the actual offset length. You may want to
store this definition onto a disk for future use. For further information on
modifying the cal kit definition see the standard HP 8510C Operating and
Programming manual and Product Note 8510 - 5A.
Содержание 85106D
Страница 11: ...x HP 85106D mm Wave System ...
Страница 95: ...4 2 HP 85106D mm Wave System Service and Troubleshooting Figure 4 1 HP 85106D System Block Diagram ...
Страница 96: ...HP 85106D mm Wave System 4 3 Service and Troubleshooting Figure 4 2 HP 85106D Option 001 System Block Diagram ...
Страница 117: ...4 24 HP 85106D mm Wave System Service and Troubleshooting Check Unratioed Power Levels ...
Страница 139: ...5 22 HP 85106D mm Wave System HP 85104A Test Set Module Troubleshooting Procedures ...
Страница 140: ......
Страница 141: ......
Страница 142: ......
Страница 143: ......
Страница 149: ...5 36 HP 85106D mm Wave System HP 85104A Test Set Module Assembly Procedure Figure 5 12 HP 85104A Assembly Procedure ...
Страница 153: ...5 40 HP 85106D mm Wave System HP 85104A Test Set Module Replaceable Parts Figure 5 13 HP Q85104A Replaceable Parts ...
Страница 156: ...HP 85106D mm Wave System 5 43 HP 85104A Test Set Module Replaceable Parts Figure 5 14 HP U85104A Replaceable Parts ...
Страница 159: ...5 46 HP 85106D mm Wave System HP 85104A Test Set Module Replaceable Parts Figure 5 15 HP V85104A Replaceable Parts ...
Страница 162: ...HP 85106D mm Wave System 5 49 HP 85104A Test Set Module Replaceable Parts Figure 5 16 HP W85104A Replaceable Parts ...
Страница 208: ......
Страница 212: ...6 48 HP 85106D mm Wave System HP 85105A mm Wave Controller Replaceable Parts Figure 6 28 HP 85105A Major Assemblies ...
Страница 214: ...6 50 HP 85106D mm Wave System HP 85105A mm Wave Controller Replaceable Parts Figure 6 29 HP 85105A Semi rigid RF Cables ...
Страница 218: ...6 54 HP 85106D mm Wave System HP 85105A mm Wave Controller Replaceable Parts Figure 6 31 Wire Harnesses ...
Страница 220: ...6 56 HP 85106D mm Wave System HP 85105A mm Wave Controller Replaceable Parts Figure 6 32 HP 85105A Front Panel ...
Страница 222: ...6 58 HP 85106D mm Wave System HP 85105A mm Wave Controller Replaceable Parts Figure 6 33 Rear Panel Standard Option 050 ...
Страница 228: ...6 64 HP 85106D mm Wave System HP 85105A mm Wave Controller Replaceable Parts Figure 6 37 HP 85105A Accessories Standard ...
Страница 238: ...B 2 HP 85106D mm Wave System Upgrade Paths ...
Страница 243: ...HP 85106D mm Wave System v work surface installation 1 12 ...
Страница 244: ... vi HP 85106D mm Wave System ...