iv
Measurement Accuracy
Measurement Accuracy
To ensure that you obtain the specified accuracy, you must perform a plug-in
module vertical calibration. The calibration must also be performed when you
move a plug-in module from one slot to another, or from one mainframe to
another. Refer to
Chapter 3, “Calibration Overview”
for information on per-
forming a plug-in module vertical calibration.
C A U T I O N
The HP 83487A optical/electrical plug-in module input circuitry can be
damaged when the
total
input power levels 10 dBm (10 mW) on the
optical channel or ±2 V + peak ac (+16 dBm) on the electrical channel. To
prevent input damage, this specified level must not be exceeded.
Measurement accuracy—it’s up to you!
Fiber-optic connectors are easily damaged when connected to dirty or damaged cables
and accessories. The HP 83487A optical/electrical plug-in module’s front-panel INPUT
connector is no exception. When you use improper cleaning and handling techniques,
you risk expensive instrument repairs, damaged cables, and compromised measure-
ments.
Before you connect any fiber-optic cable to the HP 83487A optical/electrical plug-in
module, refer to
“Cleaning Connections for Accurate Measurements” on page 5-14
.
Содержание 83487A
Страница 1: ...User s Guide HP 83487A Optical Electrical Plug In Module ...
Страница 6: ...vi General Safety Considerations ...
Страница 8: ...Contents 2 Contents ...
Страница 23: ...2 Channel Setup Menu 2 2 Displaying the Channel Setup Menus 2 5 Channel Setup Menu ...
Страница 25: ...2 3 Channel Setup Menu Channel Setup Menu Figure 2 1 Optical Channel Setup menu ...
Страница 26: ...2 4 Channel Setup Menu Channel Setup Menu Figure 2 2 Electrical Channel Setup menu ...
Страница 58: ...3 22 Calibration Overview Complete Calibration Summary ...
Страница 67: ...4 9 Specifications and Regulatory Information Declaration of Conformity Declaration of Conformity ...
Страница 68: ...4 10 Specifications and Regulatory Information Declaration of Conformity ...
Страница 98: ...Index 4 Index ...