Power Measurement
How to Measure Power
principally for polarization dependent measurements, but can be
used for other types of measurement.
You can choose three modes of operation from the MinMax mode
menu:
•
<C
ONTINUOUS
> mode, which compares each new measured
value with the maximum and minimum values so far, and
replaces them as necessary.
This mode is useful for measuring the Polarization Dependent
Loss (PDL) of a component. Run the application while
sweeping the polarization of the source applied to the
component.
•
<W
INDOW
> mode, which compares each new measured value
with the maximum and minimum values of each of the
previous N samples. When a new measurement is taken it is
added to a buffer containing the previous N samples and the
oldest measurement is deleted.
•
<R
EFRESH
> mode, which adds each new measurement to a
buffer. The minimum and maximum values, in this buffer, are
displayed. After N samples are added to the buffer, the buffer
resets and a new buffer is created.
You can use Window and Refresh modes, for example, when
you are searching for or setting the position of minimum PDL.
Содержание 8163A Series
Страница 1: ...HP 8163A Lightwave Multimeter HP 8164A Lightwave Measurement System User s Guide ...
Страница 3: ...HP 8163A Lightwave Multimeter HP 8164A Lightwave Measurement System User s Guide ...
Страница 13: ...13 Safety Summary ...
Страница 25: ...1 1 Getting Started ...
Страница 40: ...40 Getting Started A Description of the User Interface Figure 1 13 The HP 8163A System Configuration Menu ...
Страница 49: ...2 2 Additional Features ...
Страница 61: ...3 3 Power Measurement ...
Страница 76: ...76 Power Measurement How to Measure Power Figure 3 8 Manual Range Mode Within Range ...
Страница 87: ...4 4 Laser Sources ...
Страница 94: ...94 Using Laser Source Modules How to Use Laser Source Modules ...
Страница 95: ...5 5 Tunable Lasers ...
Страница 99: ...99 Tunable Laser Sources How to Set the Power Figure 5 1 Setting High Power parameters ...
Страница 103: ...103 Tunable Laser Sources How to Set the Power Figure 5 3 Setting Attenuation ...
Страница 113: ...113 Tunable Laser Sources How to Perform a Wavelength Sweep Figure 5 10 Executing a Stepped Sweep ...
Страница 130: ...130 Tunable Laser Sources How to Use Auxiliary Functions ...
Страница 131: ...A A Installation and Maintenance ...
Страница 159: ...B B Accessories ...
Страница 167: ...C C Specifications ...
Страница 175: ...D D Performance Tests ...
Страница 188: ...188 Performance Tests Test Record ...
Страница 192: ...192 Performance Tests Test Record ...
Страница 193: ...E E Cleaning Procedures ...
Страница 199: ...F F Firmware Updates ...
Страница 204: ...204 F Firmware Updates Firmware Update Process ...