3-5
Quality
Electromagnetic Compatibility Programs
Table 3-3 summarizes the magnetic emissions tests while the equipment is not operating.
Table 3-4 summarizes the magnetic emissions tests during operation.
Electromagnetic Field Immunity/Susceptibility Test Suite
This section summarizes the Field Immunity/Susceptibility Test Suite.
NOTE:
All tests in this suite were performed while the product was operating.
Table 3-5 summarizes the electro-static discharge field immunity tests.
Table 3-6 summarizes the radiated field immunity test.
FCC CFR 47 parts 2 and 15
FCC Class A
VCCI class A
CISPR 22 class A
Table 3-3
Magnetic Emissions Tests While Non-Operating
Standard
Level
HP Standard 765.006 complies with CFR49
IATA Dangerous Goods Regulations, 30ed
<2 milligauss at 2.1 m
Table 3-4
Magnetic Emissions Tests During Operation
Standard
Level
HP Standard 765.007
< 5 gauss p-p
Table 3-2
Conducted Emissions Tests During Operation
Standard
Level
Table 3-5
Electro-Static Discharge Field Immunity
Standard
Level
HP Standard 765.002
15 KV A.D. (operating)
IEC 801-2 level 2
4 KV C.D., 8 KV A.D.
prEN50082-2
(ISM) 4 KV C.D., 8 KV A.D.
prEN55024-2
(ITE) 3 KV C.D., 8 KV A.D.
Table 3-6
Radiated Field Immunity
Standard
Level
prEN55024-3 (ITE/ISM
3 V/m (IEC 801-3 level2)
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com