TG
Frequency
Response
(HP
70301A)
TG
Frequency
Response
(HP
70301A)
T
ested
Specification
AMPLITUDE
A
CCURA
CY:
Amplitude
Flatness
rel.
to
2.7
GHz
Equipment
P
o
w
er
Meter
MW
P
o
w
er
Sensor
Equipment
Setup
With
the
MW
p o
w
er
sensor
connected
to
the
p o
w
er
meter,
connect
the
MW
p o
w
er
sensor
to
the
RF
OUTPUT
of
the
tracking
generator.
Description
This
test
measures
the
RF
output
amplitude
v
ariations
of
the
tracking
generator
o
v
er
its
complete
frequency
range.
The
tracking
generator
is
set
to
a
frequency
of
2.7
GHz,
then
a
p o
w
er
meter
is
used
to
set
the
tracking
generator
output
amplitude
to
02
dBm
60.15
dB.
This
establishes
the
reference
amplitude.
The
tracking
generator
is
then
incremen
ted
in
frequency
,
the
output
p o
w
er
is
read
b
y
the
p o
w
er
meter,
and
the
v
ariation
from
the
reference
amplitude
is
stored.
F
requency
resp onse
is
measured
at
a
total
of
201
p oints.
In
Case
of
F
ailure
If
this
test
fails,
the
following
mo
dules
may
need
repair
or
adjustmen
t:
T
rac
king
Generator.
V
erification
4-5
Содержание 70301A
Страница 6: ......
Страница 10: ......
Страница 16: ...Figure 1 2 Front Panel and Rear Panel Features 1 6 General Information ...
Страница 20: ...Figure 1 3 Packaging Materials for HP 70001A Mainframe 1 10 General Information ...
Страница 21: ...Figure 1 4 Packaging Materials for Modules General Information 1 11 ...
Страница 26: ......
Страница 32: ...Figure 2 3 HP 71210C with HP 70300A and HP 70301A Rear Panel Cabling 2 6 Installation ...
Страница 50: ......
Страница 58: ......