Section V
Moael
OJ LA.
Table 5-1. Required Test Equipment.
Instrument
Required
Recommended
Type
Characteristics
Model
Oscilloscope
Pass band: 10 Hz to 10 MHz
-hp- Mode! laOA Oscilloscope
Sensitivity; 50 mV /em
with 1801 A and 1820A plug-ins
Input Impedance: 1 megohm
Electronic Counter
Range: 10 Hz to 10 MHz
-hp- Model 5245L Electronic
Accuracy:
±
5 counts
counter
AC Voltmeter
Frequency Range: 10 Hz to 10 MHz
-hp- Model 3400A RMS
Voltage Range: lmV tol0V
Voltmeter (with known
Accuracy:
±
l%
accuracy)
DC output
Distortion Analyzer
Distortion Sensitivity: greater than 42 dB
·hp- Model 331 A (332(333(334
Wave Analyzer
Frequency Range: 600 kHz to 22 MHz
-hp- Model HD5- 312A
DC Null Voltmeter
Range: 10 microvolts to 30 V
-hp- Model 419A
Accuracy:
±
2% of full scale
t
AC Differential
Range: lV to 10V
-hp- Model 741
B AC-DC
Voltmeter
Accuracy:
±
0.1%
Differentia! Voltmeter/
Stability:
±
0.1% Per Mo.
DC Standard
Attenuator
Attenuation Range: 90 dB in 10 dB steps
-hp* Model 3550 VHF
Frequency Range: 10 Hz to 10 MHz
Attenuator (with known
accuracy)
Amplifier
Gain: 40 dB
-hp· Model 461 A Amplifier
Frequency Range: 10 Hz to 10 MHz
Thermal Converter
Input: 3 V RMS
·hp- Model 11 049A
Output: 7 mV dc
Thermal Converter
Accuracy:
±
0.2%
Frequency Range: 10 Hz to 10 MHz
Input Impedance: 50 ohms
Oto 10 mV
See Figure 5-3 for schematic
Reference Supply
a. Resistor: fxd, 6500
n ±
1%
a. -hp- Part No. 0811-0392
b. Resistor: var, 500
n ±
5%
b. ·hp- Part No. 21"00-0324
c. Resistor: var, 50
n ±
5%
c. -hp- Part No. 2100- 1481
d. BatterYl 1.34 V
d. Mallory RM42R
Terminating
a. Feedthru, 50
n
a
.
•
hp- Model 11048C
Resistance
b. Feedthru, 600
n
b. -hp- Mode! 11 095A
Adapter
BNC to Binding Post
whp_ Modell 0111 A Adapter
t
Recorder
Chart Speed: 1 "/hr or less
·hp- Model 680 6" Strip
Chart Recorder
DC Digital
Range: 1 V
-hp- Model 34740A/ 34702A
Voltmeter
Digits: 4
Measurement SYstem
t
Required only if Amplitude StabilitY Test {Paragraph 5·19) is performed.
5-0