Chapter 3 Programming Your Counter for Remote Operation
Status Reporting
Programming Guide
3-35
3
Transition filters are unaffected by *CLS or queries. Transition filters are
set to default values by :STATus:PRESet and power-on.
To write to the Operation Status transition filter registers use:
:STATus:OPERation:PTRansition
:STATus:OPERation:NTRansition
To read these registers use:
:STATus:OPERation:PTRansition?
:STATus:OPERation:NTRansition?
Event Register
An event register captures changes in conditions.
An event register bit (event bit) is set TRUE when an associated event
occurs. These bits, once set, are “sticky.” That is, they cannot be cleared
even if they do not reflect the current status of a related condition, until
they are read.
Table 3-6. Transition Filter Definition
Positive
Transition
Filter Bit
Negative
Transition
Filter Bit
Transition Which Causes the Event-Bit to be set TRUE
TRUE
FALSE
positive transition
FALSE
TRUE
negative transition
TRUE
TRUE
either a positive or negative transition
FALSE
FALSE
neither transition (event reporting is disabled)
Содержание 53150A Series
Страница 1: ... Programming Guide HP 53150A 151A 152A Microwave Frequency Counter ...
Страница 2: ......
Страница 10: ...Contents x Programming Guide ...
Страница 11: ...1 Before You Start ...
Страница 20: ...Chapter 1 Before You Start Related Documentation 1 10 Programming Guide 1 ...
Страница 21: ...2 Command Summary A Quick Reference ...
Страница 41: ...3 Programming Your Counter for Remote Operation ...
Страница 100: ...Chapter 3 Programming Your Counter for Remote Operation Programming Examples 3 60 Programming Guide 3 ...
Страница 101: ...4 Command Reference A Dictionary ...
Страница 162: ...Chapter 4 Command Reference Common Commands 4 62 Programming Guide 4 ...
Страница 163: ...5 Errors ...
Страница 174: ...Chapter 5 Errors Error Types 5 12 Programming Guide 5 ...
Страница 184: ......
Страница 186: ... Manual Part Number 53150 90002 Printed in U S A NOVEMBER 1997 ...