Model 432A

Section II

Installation

2-1

SECTION II

INSTALLATION

2-1.

INITIAL INSPECTION.

2

2. MECHANICAL CHECK.

2–3. If damage to the shipping carton is evident, ask that

the carrier’s agent be present when the instrument is

unpacked. Inspect the instrument for mechanical

damage. Also check the cushioning material for signs of

severe stress.

2

–4.

PERFORMANCE CHECKS.

2–5. The electrical performance of the Model 432A

should be verified upon receipt. Performance checks

suitable for incoming inspection are given in Section V,

Maintenance.

2

–6.

DAMAGE CLAIMS.

2

7. If the instrument is mechanically damaged in

transit, notify the carrier and the nearest Hewlett-

Packard field office immediately. A list of field offices is

at the back of this manual. Retain the shipping carton

and padding material for the carrier’s inspection. The

field office will arrange for replacement or repair of your

instrument without waiting for claim settlements

against the carrier.

2

8. Before shipment this instrument was inspected

and found free of mechanical and electrical defects. If

there is any deficiency, or if electrical performance is not

within specifications, notify your nearest Hewlett-

Packard Sales and Service Office.

2-9.

THREE-CONDUCTOR POWER CABLE.

2

10. To protect operating personnel, the National

Electrical Manufacturers Association (NEMA)

recommends that the instrument panel and cabinet be

grounded. All Hewlett-Packard instruments are

equipped with a three-conductor power cable which,

when plugged into an appropriate receptacle, grounds

the instrument. The offset pin on the power cable three-

prong connector is the ground wire.

2

11. To preserve the protection feature when

operating the instrument from a two-connector outlet,

use a three-prong to two-prong adapter and connect the

green pigtail on the adapter to ground.

2-12.

PRIMARY POWER REQUIREMENTS.

2

13. The Model 432A operates from 115 or 230 volts ac

line voltage. Line frequency may vary from 48 to 440 Hz.

A slide switch on the rear panel is moved to the correct

position for the line voltage available. Before operating

the equipment, ensure that the fuse installed in the

instrument corresponds to the value marked on the

panel for the line voltage available (1/8 amp slo-blow).

2-14.

INTERNAL BATTERY OPERATION.

2

15. Model 432A Option 001 instruments contain an

internal battery and a battery charging assembly. By

connecting the 432A to an ac source, the battery may be

charged overnight. The battery can be maintained in the

charging state indefinitely without damage. It will

assume its full capacity, 1.25 ampere-hours, and will not

charge in excess of that. This enables the instrument to

operate for approximately 20 hours continuously without

recharging.

2

–16.

BATTERY INSTALLATION.

a. Set power switch to off and remove power plug

from rear panel.

b. Remove top and bottom, and side instrument

covers.

c.

The battery is installed with the terminals toward

the right hand side of the instrument when faced from

the front. The two terminals on the battery fit into spaces

provided on the circuit board.

d. Using the retaining nuts, fasten the battery firmly

in place. Be careful not to short the battery terminals at

any time as this may cause battery cell damage.

e.

Install assembly A7, battery charging board, in

the space provided for it just ahead of the battery.

f.

Reinstall instrument covers and adjust circuit.

Instrument is now ready for operation.

2

–17.

BATTERY STORAGE.

2

18. Store the battery at or below room temperature.

Extended storage at high temperature will reduce the

cell charge, but will not damage the battery if the storage

temperature is below 140

°

F. Install the battery in the

instrument and recharge before using Model 432A in

battery operation.

2-19.

RACK MOUNTING.

2

20. Model 432A is narrower than full-rack width. It is

what is termed a sub-modular unit. When used alone,

the instrument can be bench mounted. When used in

combination with other sub-modular units it may be

bench or rack mounted. The HP 1051A and 1052A

Combining Cases and Rack Adapter Frames are

designed specifically for this purpose.

2

–21.

COMBINING CASE.

2

22. A model 1051A Combining Case  is shown in

Figure 2

1. This case is full rack width and accepts

varying combinations of submodular instruments. The

case, purchased separately, is provided with a rack

mounting  kit. The  combining  case  will  hold  three

WARNING

If this instrument is not used as speci

fi

ed, the protection provided by

the equipment could be impaired. This instrument must be used in

a normal condition only (in which all means for protection are

intact).

Содержание 432A

Страница 1: ...swesen nun zu der Firma Agilent Technology gehören Um Verwirrung zu vermeiden wurde lediglich bei Produktname und Nummer der vo laufende Firmenname geändert Produkte mit dem Namen Nummer HP XXXX lauten nun mehr Agilent XXXX Z B das Modell HP 8648 heißt nun Agilent 8648 Questo manuale potrebbe contenere riferimenti ad HP o Hewlett Packard Si noti che le attività precedentemente gestite da Hewlett P...

Страница 2: ......

Страница 3: ...quipment 5 1 5 12 Service Information 5 1 Section Page V MAINTENANCE Cont 5 14 432A Performance Tests with 8477A Calibrator 5 1 5 15 Initial Setup 5 1 5 16 Meter Accuracy Test 5 1 5 17 Calibration Factor Test 5 2 5 18 Meter Linearity Check 5 2 5 19 Zero Carryover Test 5 2 5 20 Fine Zero Range Check 5 3 5 21 432A Calibration Without 8477A Calibrator 5 3 5 24 Calibration Procedure 1 5 3 5 26 Calibra...

Страница 4: ...ter Frame 2 2 2 2 HP Model 1051A Combining Case Instrument Installation 2 2 3 1 Precision Power Measurements 3 3 3 2 Front Panel Controls Connectors and Indicators 3 4 3 3 Rear Panel Controls and Connectors 3 6 3 4 Turn On and Zeroing Procedure 3 8 4 1 Model 432A Simplified Block Diagram 4 0 4 2 Model 432A Block Diagram 4 3 4 3 Model 432A RF Bridge Talking Schematic 4 4 4 4 Model 432A Meter Logic ...

Страница 5: ...magnetic compatibility Against EMC test specifications EN 55011 1991 Group 1 Class A and EN 50082 1 1992 As Detailed in Electromagnetic Compatibility EMC Technical Construction File TCF No A 5951 9852 02 Assessed by DTI Appointed Competent Body EMC Test Centre GEC Marconi Avionics Ltd Maxwell Building Donibristle Industrial Park KY11 5LB Scotland United Kingdom Technical Report Number 6893 2200 CB...

Страница 6: ...Operation outside of the environmental and electrical specifications for the product 5 Improper site preparation and maintenance or 6 Customer induced contamination or leaks THE WARANTY SET FORTH IS EXCLUSIVE AND NO OTHER WARRANTY WHETHER WRITTEN OR ORAL IS EXPRESSED OR IMPLIED HP SPECIFICALLY DISCLAIMS THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE Limitation of Re...

Страница 7: ...oduced without the prior written consent of the manufacturer Hewlett Packard Ltd Restricted Rights Legend Use duplication or disclosure by the government is subject to restrictions as set forth in subdivision b 3 ii of the Rights in Technical Data and Computer Software clause at 52 227 7013 Hewlett Packard Company 3000 Hanover Street Palo Alto California 94304 Statement of Compliance Electromagnet...

Страница 8: ... NOT perform procedures involving cover or shield removal unless you are qualified to do so Operating personnel must not remove equipment covers or shields Procedures involving the removal of covers and shields are for use by service trained personnel only DO NOT service or adjust alone Under certain conditions dangerous voltages may exist even with the equipment switched off To avoid dangerous el...

Страница 9: ...Indicates hazardous voltages WARNING Warning denotes a hazard It calls attention to a proce dure which if not correctly performed or adhered to could result in injury or loss of life Do not proceed beyond a warning note until the indicated conditions are fully understood and met CAUTION Caution denotes a hazard It calls attention to a proce dure which if not correctly performed or adhered to could...

Страница 10: ...ix Model 432A Regulatory Information Noise Declaration LpA 70dB am Arbeitsplatz operator position normaler Betrieb normal position nach DIN 45635 pt 19 per ISO 7779 ...

Страница 11: ......

Страница 12: ...ements or in an Appendix in the Manual If the change information is missing contact your HP Sales Office Sales Offices are listed at the back of the Manual SER 1234A98765 COMPANY HEWLETT PACKARD MADE IN U S A SERIAL PREFIX Figure 1 2 Instrument Identification Table 1 1 Specifications Instrument Type Automatic self balancing power meter for use with temperature compensated thermistor mount Power Ra...

Страница 13: ......

Страница 14: ...clean the module instrument Use a soft clean damp cloth to clean the front panel and side covers CAUTION Mains supply voltage fluctuations should not exceed 10 of the nominal selected line voltage CAUTION Before switching on this instrument make sure that the line voltage slide switch is set to the voltage of the power supply and the correct fuse is installed see Figure 3 3 Ensure the power supply...

Страница 15: ......

Страница 16: ... panel for the line voltage available 1 8 amp slo blow 2 14 INTERNAL BATTERY OPERATION 2 15 Model 432A Option 001 instruments contain an internal battery and a battery charging assembly By connecting the 432A to an ac source the battery may be charged overnight The battery can be maintained in the charging state indefinitely without damage It will assume its full capacity 1 25 ampere hours and wil...

Страница 17: ......

Страница 18: ......

Страница 19: ......

Страница 20: ......

Страница 21: ......

Страница 22: ...cale note the value in dBM of the range in use and subtract from it the reading on the meter dBm scale 6 Mechanical Meter Zero Sets meter suspension so that meter indicates zero To adjust the zero a Turn POWER switch off b Turn the adjustment screw clockwise until the indicator falls below zero and comes back up to zero again c Turn the adjustment very slightly counter clockwise to free up the mec...

Страница 23: ......

Страница 24: ...red in parallel with front panel connector Only one mount at a time may be used with the power meter 5 VRF Input Connected directly to RF bridge Used for calibrating power meter with HP 8477A Power Meter Calibrator Also used for precision power measurements 6 VCOMP Input Connected directly to compensation bridge Used for calibrating power meter with HP 8477A Power Meter Calibrator Also used for pr...

Страница 25: ......

Страница 26: ...ount used 4 Turn the 432A POWER switch ON For battery operation the AC LINE indicator does not turn on 5 Set RANGE selector to COURSE ZERO and then zero the meter with the COURSE ZERO screwdriver adjustment Note The power meter should be zeroed with the RF power source turned off or the mount disconnected from the source 6 Set the range selector to the 0 01 mW range then depress the FINE ZERO swit...

Страница 27: ......

Страница 28: ......

Страница 29: ......

Страница 30: ......

Страница 31: ......

Страница 32: ......

Страница 33: ......

Страница 34: ......

Страница 35: ......

Страница 36: ......

Страница 37: ......

Страница 38: ......

Страница 39: ......

Страница 40: ......

Страница 41: ......

Страница 42: ......

Страница 43: ......

Страница 44: ......

Страница 45: ......

Страница 46: ......

Страница 47: ......

Страница 48: ......

Страница 49: ......

Страница 50: ......

Страница 51: ......

Страница 52: ......

Страница 53: ......

Страница 54: ......

Страница 55: ......

Страница 56: ......

Страница 57: ......

Страница 58: ......

Страница 59: ......

Страница 60: ......

Страница 61: ......

Страница 62: ......

Страница 63: ......

Страница 64: ......

Страница 65: ......

Страница 66: ......

Страница 67: ......

Страница 68: ......

Страница 69: ......

Страница 70: ......

Страница 71: ......

Страница 72: ......

Страница 73: ......

Страница 74: ...solete Model 432A 100 Model 432A 101 was obsolete in November 2005 due to parts depletion Replacement product for Model 432A 100 The Model 432A 100 has been replaced with Model 432 101 as this product supports 100 VAC line voltage input Ordering Information Customers should place order for Model 432A 101 instead of Model 432A 100 ...

Страница 75: ......

Страница 76: ......

Страница 77: ......

Страница 78: ...pectrum Way Mississauga Ontario L4W 5G1 905 206 4725 Europe Africa and Middle East Austria 0 1 25000 0 Belgium and Luxembourg 02 778 3417 Baltic Countries 358 08872 2100 Czech Republic 420 2 4743111 Denmark 45 99 10 00 Finland 90 88 721 France 0 1 69 82 60 60 Germany 0180 532 62 33 Greece 30 1 7264045 Hungary 36 1 4618219 Ireland 01 284 4633 Israel 972 3 5380333 Italy 02 92 122 241 Netherlands 020...

Отзывы: