Section V
ModeI400F/FL
Paragraphs 5-14 to 5-15 and Tables 5-2 to 5-3
5-2
Table 5-2. Full Scale Calibration Tolerances (400F)
30
MV
RANGE
100
MY
RANGE
1 VOLT RANGE
(0.
1
mv Range Cheek)
METER
METER
METER
FREQ.
INDICATION
FREQ.
INDICATION
FREQ.
INDICATION
MIN.
MAX.
MIN.
MAX.
MIN.
MAX.
'0
28.8
31. 2
'0
96
104
20
0. 96
1. 04
40
29.4
30.6
40
98
10'
40
0.98
1. 02
400
29.7
30.3
400
99
101
400
0.99
1. 01
1000
29.7
30.3
1000
99
101
1000
0.99
1. 01
10 K
29. 7
30. 3
10 K
99
101
10 K
0.99
1. 01
100 K
29. 7
30.3
100 K
99
101
100 K
0.99
1. 01
I M
29. 7
30. 3
I M
99
101
I M
0. 99
1. 01
' M
29.4
30. 6
2 M
98
10'
2 M
0.98
1. 02
4 M
28. 8
31. 2
4 M
96
104
4 M
0. 96
1. 04
Table 5-3. Full Scale Calibration Tolerances (400FL)
30
MY
RANGE
100
MV
RANGE
1 VOLT RANGE
(0.1 mv Range Check)
METER
FREQ.
INDICATlON
FREQ.
MIN.
MAX.
20
28.5
31. 5
'0
40
29.4
30.6
40
400
29. 7
30.3
400
1000
29. 7
30.3
1000
10 K
29. 7
30. 3
10 K
100 K
29. 7
30.3
100 K
I M
29.7
30. 3
I M
2 M
29.4
30. 6
2 M
4 M
28. 5
31. 5
4 M
j. Adjust test oscillator set for a 30 mv output
at 400 Hz, using as a reference the 400F/FL
meter indication obtained in step d of this
paragraph. Set a reference on meter of test
oscillator and use amplitude control to main
tain the set reference whenever frequency of
osc11lator is varied.
k. Repeat step d of this paragraph for each fre
quency Hsted in Table 5-2 (400F) or Table
5-3 (400FL).
m. Perform steps eandfof this paragraph, main
tainingthe output amplitude of the test oscil
lator at 30 my.
n. Repeat step d of this paragraph for each fre
quency listed in Table 5-2 (400F) or Table 5-3
(400FL).
o. Set 400F/FL RANGE sw.itch tol volt, and dis
COlUlect shorts between test paints.
METER
METER
INDICATION
FREQ.
INDICATION
MIN.
95
98
99
99
99
99
99
98
95
MAX.
MIN.
MAX.
105
20
0. 95
1. 05
102
40
0.98
1. 02
101
400
0.99
1. 01
101
1000
0.99
1. 01
101
10 K
0.99
1. 01
101
lOOK
0. 99
1. 01
101
I
M
0.99
1. 01
102
2 M
0.98
I. 02
105
4 M
0. 95
1. 05
p. Adjust test oscillator for a 1 volt output at
400 Hz using as a reference the 400F/FL
meter indication obtained i n step h of this
paragraph.
q. Repeat step d of this paragraph for each fre
quency Hsted in Table 5-2 or Table 5-3.
5-14. RANGE TRACKING CHECK.
5-15. After verifying the 400F/FL full scale calibra
tion with the accuracy and frequency response tests,
check the range tracking of the instrument with the
following procedures. Use the test setup shown in
Figure 5-1 for the range tracking check.
a. Set switch SI to Position A.
b. Set 400F/FL RANGE switch to 30 mv.
c. Adjust test oscillator for a 400F/FL meter
indication of 30 mv at 400 Hz.
01795-1
Содержание 400F
Страница 2: ...Verso Filler Page ...
Страница 3: ...400F FL AC VOLTMETER OPERATING AND SERVICE MANUAL HEWLETT PACKARD ...
Страница 6: ...Verso Filler Page ...
Страница 16: ...Section IV Figure 4 1 4 0 O 1 NO 0 0 ModeI400F FL u 0 8 0 j JO W t 01795 1 ...
Страница 28: ...Verso Filler Page ...
Страница 33: ...Verso Filler Page ...
Страница 53: ......
Страница 54: ...HEWLETTitPACKARD 00467 90003 Printed in U SA ...