background image

TM11-6625-1576-15

Section IV
Paragraphs 4-13 to 4-25 and Figure 4-2

the emitter of A2Q3 to the collector of A2Q2. Overall
negative feedback from the emitter circuit of A2Q4 to
the source of A2Q1 results in unity gain from the im-
pedance converter.

4-13. The bias points of the transistors in the imped-
ance converter are selected to minimize instrument

induced distortion. A2Q 1, an extremely low noise,
high impedance field effect transistor, is the major

component that makes linearity of the Impedance con-
verter independent of the signal source impedance.

4-14. REJECTION AMPLIFIER CIRCUIT.

4-15. The rejection amplifier circuit (see Figures 6-3
and 6-6) consists of the preamplifier (A3Q1) thru A3Q3),
the Wien bridge resistive leg and auto control loop
(A5Q1 thru A5Q9 with associated lamp and photocell),
the reactive leg and auto control loop (A5Q10 thru
A5Q 18 with associated lamp and photocell), and the
bridge amplifier (A3Q4 thru A3Q6).

4-16. PREAMPLIFIER CIRCUIT.

4-17. The signal from the impedance converter is ap-
plied to the preamplifier, which is used during SET

LEVEL and DISTORTION measuring operations. Neg-

ative feedback from the junction of A3R1O and A3R11 is
applied to the junction of A3R2 and A3C2 to establish

the operating point for A3Q1. Negative feedback from
the emitter of A3Q3 is applied to the emitter of A3Q1
to stabilize the preamplifier. The preamplifier, like
the impedance converter, is designed for high open
loop gain and low closed loop gain to minimize instru-
ment induced distortion.

4-18. WIEN BRIDGE CIRCUIT.

4-19. In the distortion measuring operation the Wien
bridge circuit is used as a rejection filter for the

fundamental frequency of the input signal. With the

FUNCTION selector, S1, in the DISTORTION position,
the Wien bridge is connected as an interstage coupling
network between the preamplifier circuit and the bridge
amplifier circuit. The bridge is tuned to the fundamen-
tal frequency of the input signal by setting the FRE-

QUENCY RANGE selector, S4, for the applicable fre-
quency range, and tuning the capacitors C4A through

C4D. The bridge circuit is balanced by adjusting the
COARSE balance control, R4, and the FINE balance
control, R5. In the AUTOMATIC MODE fine tuning
and balancing are accomplished by photoelectric cells
which are in the resistive and reactive legs of the
Wien bridge. The error signals for driving the photo-
cells are derived by detecting the bridge output using

the input signal as a reference.

4-20. When the Wien bridge is not tuned exactly to the
frequency to be nulled, a portion of the fundamental
frequency will appear at the bridge output. The phase
of this signal depends on which leg of the bridge is not
tuned, or on the relative errors in tuning if neither is
set correctly. The magnitude of the signal is propor-
tional to the magnitude of the tuning error of either or
both legs of the bridge.

4-2

Model 333A/334A

Figure 4-2. Bridge Waveforms

4-21. Figure 4-2a is a sinusoid input to the Wien bridge.
If the resistive leg of the bridge i

slightly unbalanced,

the output of the bridge is very small, but has the
waveform shown in Figure 4-2b and is in phase with
the input. As the resistive leg is tuned, the signal

approaches zero amplitude at null and then becomes

larger, but 180° out of phase, if the null position is

passed. When the resistive leg is correctly tuned
and the reactive leg is tuned through null, a similar
waveform is produced, Figure 4-2c. The only differ-
ence is that the reactive signal is 90° out of phase with
the resistive signal.

4-22. When the bridge output is detected using the
input signal as the reference, the error signals in

phase or 180° out of phase with the reference develop
a voltage which is used to vary the resistance in the

resistive leg of the bridge, to tune it to the correct

null position, Signals of the form in Figure 4-2c do
not develop any voltage as the resistive detector is

insensitive to input differing from the reference by 90° .

4-23. In an independent, but similar control loop, the

bridge input signal is shifted 90° and used as the

reference signal for the detector. This detector

develops control voltages to null the reactive leg of
the bridge, but is insensitive to signals of the form

in Figure 4-2b which are caused by small tuning

errors of the resistive branch.

4-24. The result is that the two control loops derive

information from a common source and develop two
independent control signals for nulling the two legs

of the bridge. These control voltages are used to
vary the brilliance of lamps, which in turn causes

resistance changes in photocells which form part of

the Wien bridge.

4-25. Refer to Figure 4-3 for the phase relationship
of the bridge error signal and reference voltage at the
base of A5Q4. The shaded portions of the error sig-
nals (b and c) indicate that part of the error signal
which contributes to the dc lamp control voltage.  As

indicated in d, any error signal that is 90° out of

phase with the reference does not affect the dc lamp
control voltage.

Содержание 333A

Страница 1: ...1 6625 1576 15 DEPARTMENT OF THE ARMY TECHNICAL MANUAL ORGANIZATIONAL DS GS AND DEPOT MAINTENANCE MANUAL DISTORTION ANALYZER HEWLETT PACKARD MODELS 333A AND 334A HEADQUARTERS DEPARTMENT OF THE ARMY MAY 1967 ...

Страница 2: ...TM 11 6625 1576 15 WARNING DANGEROUS VOLTAGES EXIST IN THIS EQUIPMENT Be careful when working on the power supply and on the 115 volt ac line connections DO NOT TAKE CHANCES ...

Страница 3: ...stortion Measurement in AM Carriers 4 1 4 8 Voltmeter Operation 4 1 4 10 Schematic Theory 4 1 4 11 Impedance Converter Circuit 4 1 4 14 Rejection Amplifier Circuit 4 2 4 36 High Pass Filter 4 5 4 38 Meter Circuit 4 5 4 46 Power Supply Circuit 4 7 4 51 RF Detector Circuit 334A only l 4 7 Section Page V MAINTENANCE 5 1 5 l Introduction 5 1 5 2 Test Equipment Required 5 1 5 5 Performance Checks 5 1 S...

Страница 4: ......

Страница 5: ...TM 11 6625 1576 15 Model 333A 334A Section 1 Figure 1 1 and Table 1 1 Figure 1 1 Model 333A Distortion Analyzer 1 1 Table 1 1 Specifications ...

Страница 6: ...TM 11 6625 1576 15 Section I Model 333A 334A Table 1 1 Table 1 1 Specifications Cont d 1 2 ...

Страница 7: ...t h r o u g h p u b l i c a t i o n s s u p p l y c h a n n e l s T h e i n d e x l i s t s t h e i n d i v i d u a l p a r t s 1 0 2 0 3 5 P e t c a n d t h e l a t e s t c h a n g e s t o a n d r e v i s i o n s o f e a c h e q u i p m e n t p u b l i c a t i o n l A 3 F o r m s a n d R e c o r d s a R e p o r t s o f M a i n t e n a n c e a n d U n s a t i s f a c t o r y E q u i p m e n t U s ...

Страница 8: ......

Страница 9: ...s a fre quency range of 5 cps to 3 Mc 20 cps to 500 Kc for 300 pv range and a voltage range of 300 pv to 300 v rms full scale 1 5 The AM detector included in the Model 334A is a broadband dc restoring peak detector consisting of a semiconductor diode and filter circuit AM distortion levels as low as O 3 can be measured on a 3 v to 8 v rms carrier modulated 30 in the standard broadcast band and low...

Страница 10: ......

Страница 11: ...pin on the power cable three prong connec tor is the ground wire 2 9 To preserve the protection feature when operat ing the instrument from a two contact outlet use a three prong to two prong adapter and connect the green pigtail on the adapter to ground 2 10 INSTALLATION 2 11 The Model 333A 334A is fully transistorized therefore no special cooling is required However the instrument should not be ...

Страница 12: ...TM 11 6625 1576 15 Section III Model 333A 334A Figure 3 1 Figure 3 1 Front and Rear Panel Description 3 0 ...

Страница 13: ...ver age to effective values in a true sine wave is approxi mately O 9 to 1 When the meter is used to measure complex waves the voltage indicated may not be the rms value of the signal applied This deviation of meter indication exists because the ratios of average to effective values are usually not the same in a com plexwave as in a sine wave The amount of deviation depends on magnitude and phase ...

Страница 14: ... SET LEVEL position thus making the next range 30 etc h Adjust SENSITIVITY VENIER control for full scale deflection j Position FRIQUENCY RANGE selector and frequency dial to fundamental frequency of input signal 3 2 Model 333A 334A k Position FUNCTION selector to DISTORTION m Adjust METER RANGE selector and frequency dial vernier control for minimum meter indication n Adjust COARSE and FINE BALANC...

Страница 15: ...tion III Paragraphs 3 23 to 3 27 3 25 The 333A and 334A perform as general purpose AC Voltmeters when the FUNCTION selector is set to VOLTMETER position NOTE With the FUNCTION selector in VOLT METER position the SENSITIVITY se lector is disabled a Position METER RANGE selector to a range exceeding the value of the signal to be measured b Connect signal to INPUT terminals c Select a METER RANGE to ...

Страница 16: ...F i g u r e 4 1 B l o c k D i a g r a m T M 1 1 6 6 2 5 1 5 7 6 1 5 S e c t i o n I V F i g u r e 4 1 Model 333A 334A ...

Страница 17: ...ost attenuator is used to limit the input signal level applied to the metering circuit to 1 mv for full scale deflectlon The metering circuit sensitivity is increased to 300 uV for full scale deflection on the 300 pv range The metering circuit provides a visual indication of the OPERATION distortion level of the input signal In addition to the visual indication provided by the meter the OUTPUT ter...

Страница 18: ...g the bridge output using the input signal as a reference 4 20 When the Wien bridge is not tuned exactly to the frequency to be nulled a portion of the fundamental frequency will appear at the bridge output The phase of this signal depends on which leg of the bridge is not tuned or on the relative errors in tuning if neither is set correctly The magnitude of the signal is propor tional to the magn...

Страница 19: ...of the FREQUENCY RANGE switch S4 switches A5R56 in parallel with A5R55 on the top three frequency ranges A6DS2 will become brighter and lower the resistance of A6V2 through A6V5 making variation in resistance less than on the two lower ranges However less variation in resistance is needed to tune the leg because the impedance in the reactive leg becomes progressively less as the higher frequency r...

Страница 20: ...tector A5Q4 4 30 Refer to Figure 4 5 simplified partial schematic for detector operation The discussion is applicable to both resistive and reactive detector circuits 4 31 The signals from the error amplifier A5Q2 and A5Q3 will be equal and of opposite phase and will cancel out each other when the detector A5Q4 is off However when the positive half of the reference square wave gates A5Q4 on the si...

Страница 21: ...tivity is increased to 300 uv for full scale deflection on the 300 uV range by switching resistors A2R29 and A2R30 into the calibration network Resis tor A2R41 and capacitor A2C29 are also switched into the calibration network on the 300 Pv range to extend the passband of the amplifier 4 42 METER AMPLIFIER CIRCUIT 4 43 The meter amplifier circuit consists of a five stage amplifier circuit A2Q5 thr...

Страница 22: ...TM 11 6625 1576 15 Section IV Figure 4 7 Model 333A 334A Figure 4 7 Bandwidth Versus Null Depth 4 6 ...

Страница 23: ...nal series regulator type and operates the same as the 25 volt regulated supply 4 50 Diodes A1CR5 and A1CR6 are coupling and pro tection diodes for external battery supplies The diodes protect the series regulator circuits from application of incorrect polarity at the battery input terminals The diodes also protect external batteries from being charged when the ac power is being used with batterie...

Страница 24: ...TM 11 6625 1576 15 Section V Table 5 1 Model 333A 334A Table 5 1 Test Equipment Required 5 0 ...

Страница 25: ...as follows MIN Position 1 next step Position 2 etc to Max Position 6 5 9 FUNDAMENTAL REJECTION CHECK a b S e t Connect 33lA 332A as shown in Figure 5 1 Set Distortion Analyzer controls as follows FUNCTION Selector VOLTMETER METER RANGE Selector 1 VOLT FREQUENCY RANGE Selector X1OO Frequency Dial 50 filter White Instr Lab Model 2640 5Kc on Distortion Analyzer meter d Switch Distortion Analyzer FUNC...

Страница 26: ...control for a O db indication on the Distortion Analyzer meter h Switch Distortion Analyzer FUNCTION selector to DISTORTION The meter reading shall not change more than 0 6 db Model 333A 334A j Repeat steps a through h at the frequency settings indicated in Table 5 2 The meter readings shall change within the limits specified Table 5 2 Second Harmonic Accuracy Check 5 11 a b c DISTORTION INTRODUCE...

Страница 27: ...t Check 5 12 FREQUENCY CALIBRATION ACCURACY CHECK b c TM 11 6625 1576 15 Section V Paragraph 5 12 Table 5 3 and Figure 5 3 NOTE From 5 cps to 10 cps the FREQUENCY dial may be as much as 3 low In this test the dial is held constant and the in put frequency is varied and monitored If the dial is low the input frequency at null will be high If the period of the in put frequency is measured it will be...

Страница 28: ...eter shall indicate less than 30 pf d Switch the Distortion Analyzer on the 0 3 range and measure capacitance Meter shall indicate less than 60 pf 5 4 e Set Distortion Analyzer controls as follows FUNCTION Selector DISTORTION SENSITIVITY Selector MIN METER RANGE Selector VOLTS f Measure Capacitance at each SENSITIVITY selector setting of the Distortion Analyzer The L C meter shall indicate less th...

Страница 29: ...ut c Set Distortion Analyzer FUNCTION Selector to VOLTMETER d Set voltmeter calibrator for 400 cps output e Check the Distortion Analyzer voltmeter full scale readings on all ranges against the appropriate rms input voltages from the voltmeter calibrator The voltmeter accuracy shall be within 2 f Set the Distortion Analyzer METER RANGE selector to 1 VOLT range Transformer set to 105 v 125 v 5 18 H...

Страница 30: ...l indicate 0 9 the toler ances indicated k Switch the Frequency Response Test Set RANGE SELECTOR to the 1 3 Mc position Set the FREQ TUNING dial to the frequencies listed in Table 5 5 Adjust the Frequency Response Test Set AMPLITUDE control until the meter indicates SET LEVEL after each frequency setting The Distortlon Analyzer meter shall indicate 0 9 the tolerances indicated 5 6 n Repeat steps b...

Страница 31: ... 33 Adjustment locations are shown in Figure 5 8 5 24 The Adjustment and Calibration Procedure is performed with the ac power cord connected to nomi nal line voltage 115 v 230 v 50 to 1000 cps MODE switch to MANUAL HIGH PASS FILTER to OUT and NORM RF DET switch to NORM unless otherwise specified 5 25 METER MECHANICAL ZERO SET 5 26 The meter is properly zero set when the pointer rests over the zero...

Страница 32: ...TM11 6625 1576 15 Section V Model 333A 334A Figure 5 8 Figure 5 8 Component and Adjustment Location 5 8 ...

Страница 33: ... Figure 5 9 g Repeat steps b through f adjusting A3R30 in stead of A3R16 5 29 BRIDGE BALANCE ADJUSTMENT C3 MAINTAIN THE LEAD DRESS TO C3 AND TO THE TUNING CAPACITOR C4 ANY CHANGE IN LEAD DRESS WILL CAUSE A CHANGE IN CAPACITANCE a Connect the Test Oscillator to the Distortion Analyzer b Set the Distortion Analyzer controls as follows FUNCTION Selector SET LEVEL METER RANGE Switch SET LEVEL Range FR...

Страница 34: ...er meter indication of 0 9 on the 1 0 scale 5 10 d Adjust the frequency response test set meter control for a set level indication e Change frequency response test set frequency t 20 cps f Readjust the amplitude until the frequency re sponse test set meter indicates set level g The Distortion Analyzer meter shall indicate 0 9 5 NOTE If reading exceeds these limits change A2C25 Typical value for th...

Страница 35: ...F THE INSTRUMENT AVOID TOUCHING ANY OF THESE CIR CUITS WITH THE BARE FINGER AS SKIN OILS ARE EXTREMELY CONTAMI NATING IF HANDLING IS NECESSARY WEAR CLEAN COTTON OR RUBBER GLOVES DO NOT USE A PENCIL TO TRACE CIRCUITS IN THE INSTRUMENT GRAPHITE PENCIL LEAD IS AN EX TREMELY GOOD CONDUCTOR AND MN ACCIDENTLY INTRODUCED PATH OF THIS TYPE IS SOMETIMES DIFFICULT TO LOCATE TO AVOID SURFACE CON TAMINATION O...

Страница 36: ...TM 11 6625 1576 15 Section V Table 5 8 Model 333A 334A 5 12 ...

Страница 37: ...nent lead hole by heating the hole with the iron and inserting a wooden toothpick Re move the toothpick after the solder has cooled and insert the new component lead d To replace components shape new leads and insert them in lead holes Reheat with soldering iron and add a small amount of new solder as required to insure a good electrical connection e Clean excessive flux from the connection and ad...

Страница 38: ...TM 11 6625 1576 15 S e c t i o n V T a b l e 5 9 Model 333A 334A 5 14 ...

Страница 39: ...Figure 5 10 Troubleshooting Logic 5 15 S e c t i o n V Figure 5 10 T M 1 1 6 6 2 5 1 5 7 6 1 5 ...

Страница 40: ...Figure 5 11 Frequency Tuning Assembly 5 16 Model 33A 334A Section V Table 5 10 TM 11 6625 1576 15 ...

Страница 41: ...ic diagram An A5 board component location diagram is included wiring data and A5 board component location diagram because the A5 components are not identified on the for the 33A and 334A Distortion Analyzers The sche prlnted circuit boards A partial interconnecting matic diagram illustrate the circuits contained within diagram is also included to assist in reconnecting each assembly as well as the...

Страница 42: ...6 2 Figure 6 1 P O Internal Wiring Data Section VI Figure 6 1 Model 333A 334A TM 11 6625 1576 15 ...

Страница 43: ...tant General HAROLD K JOHNSON General United States Army Chief of Staff Distribution Active Army USAMB 1 USACDCEC 1 USACDCCEA 1 TOAD 6 USACDCCEA LEAD 8 Ft Huachuca 1 NG None USAR None For explanation of abbreviations used see AR 320 50 U S GOVERNMENT PRINTING OFFICE 1990 0 262 912 30346 SAAD 5 Eighth USA 5 ...

Страница 44: ......

Страница 45: ......

Страница 46: ......

Страница 47: ......

Страница 48: ......

Страница 49: ......

Страница 50: ......

Страница 51: ......

Страница 52: ......

Страница 53: ...PIN 016265 000 ...

Отзывы: