TB 9-6625-2049-35
21
(a)
ENTRY FREQ
.
(b)
100
kHz
.
(c)
ENTRY AMPTD
.
(d)
10 V RMS
.
(e)
AMPTD CAL
.
(5) Multimeter will indicate between 9 and 11 V ac.
(6) Press keys and enter values using
DATA
keys as listed in (a) through (f) below:
(a)
FUNCTION
square wave.
(b)
ENTRY FREQ
.
(c)
2
kHz
.
(d)
ENTRY AMPTD
.
(e)
20 V RMS
.
(f)
AMPTD CAL
.
(7) Multimeter will indicate between 19.6 and 20.4 V ac.
(8) Press
FUNCTION
triangle key and enter
11.5 V RMS
using
DATA
keys.
(9) Press
AMPTD CAL
key. Multimeter will indicate between 11.27 and 11.73 V ac.
(10) Press
FUNCTION
positive ramp
key and repeat (9) above.
b. Adjustments
. No adjustments can be made.
20. Power Supply
NOTE
Do not perform power supply check if all other parameters are
within tolerance.
a. Performance Check
(1) Connect multimeter
INPUT HI
to -15 V TP (fig. 5) and
LO
to ground side of C9
(fig. 5). Adjust R22 (fig. 5) for a multimeter indication of -14.97 to -15.03 V dc (R).
(2)
Connect multimeter
INPUT HI
to +15 V TP (fig. 5). Readjust R22 (fig. 5) if
multimeter does not indicate between 14.9 and 15.1 V dc.
(3)
Connect multimeter
INPUT HI
to +5 V TP (fig. 5). Readjust R22 (fig. 5) if
multimeter does not indicate between 5.01 and 5.05 V dc.
(4) Repeat (1)
through (3) above for best in-tolerance condition on all test points.