TB 9-6625-2049-35
11
(4) Set oscilloscope controls to measure ramp retrace time from the 90 to 10 percent
points. Ramp retrace time will be
≤
3
µ
s.
(5) Press
FUNCTION
negative ramp key and repeat (4) above.
b. Adjustments
. No adjustments can be made.
12. Frequency Accuracy
a. Performance Check
(1) Connect TI
SIGNAL
to frequency counter
CHANNEL A
.
(2) Press keys and enter values using
DATA
keys as listed in (a) through (f) below:
(a)
SIGNAL
off (option 002 not lit).
(b)
FUNCTION
sine wave.
(c)
ENTRY FREQ
.
(d)
20 MHz
.
(e)
ENTRY AMPTD
.
(f) .
99 VOLT
.
(3) Set frequency counter
50
Ω
Ω
/1M
Ω
Ω
switch to
50
Ω
Ω
.
(4) Set frequency counter controls to measure frequency. If frequency counter does
not indicate between 19.99990 and 20.00010 MHz, perform
b
below.
(5) Press
FUNCTION
square wave key. Frequency counter will indicate between
9.999950 and 10.000050 MHz.
(6) Disconnect frequency counter from TI
SIGNAL.
(7) Connect frequency counter
CHANNEL A
to TI
SYNC OUT.
(8) Press TI
FUNCTION
triangle key
.
(9) Set frequency counter controls to measure period. Frequency counter will
indicate between 99.99950 and 100.00050
µ
s.
(10) Press
FUNCTION
positive ramp key. Frequency counter will indicate between
99.99950 and 100.00050
µ
s.
(11) Press
FUNCTION
negative ramp key. Frequency counter will indicate between
99.99950 and 100.00050
µ
s.
(12) Disconnect frequency counter from TI
SYNC OUT
.
(13) Connect frequency counter to TI
AUX 21-60 MHz
(rear panel).
(14) Press keys and enter values using
DATA
keys as listed in (a) through (c) below:
(a)
FUNCTION
sine wave.
(b)
ENTRY
FREQ
.
(c)
60
MHz
.