background image

Operation

This section is organized into the following two sections:

Attaching a socket module onto HP 16442A.

Mounting and wiring DUT on the socket module.

Select the proper environment to execute the measurements. Place HP 16442A on a flat
surface. The HP 16442A shields from electromagnetic interference, but Hewlett-Packard
recommends that the measurement is performed in as noiseless environment as possible for
more stable and accurate measurement results.

You place the DUT into the provided socket modules in HP 16442A, then measure the dc
characteristics by using the connected measurement units.

Warning

To prevent electric shock, be sure to connect the cable between the interlock
(Intlk) terminals of measurement instrument and HP 16442A before performing
measurements.

Warning

Do

not

operate HP 16442A while the High Voltage indicator is lit. Dangerous

voltages of up to 200 V may be present on the connection between
measurement units and DUT.

Warning

Do

not

get connection cables caught in the lid of HP 16442A during

measurements. Dangerous voltage of up to 200 V may be present on the
connection cables.

Note

After setting up HP 16442A and DUT, confirm that the connections are
correct. Shut lid of HP 16442A to close switches of the interlock (Intlk)
terminal.

Caution

To prevent damage to instrument or DUT,

do not

remove or attach DUT in

HP 16442A when voltage or current is present.

Caution

Do not

touch any metal surface that is in the circuit path. Oil, perspiration,

and dirt prevent good electric contact and degrade measurement accuracy.

4

HP 16442A User’s Gude

Содержание 16442A

Страница 1: ...Test Fixture User s Guide A B C D E HP Part No 16442 90000 Printed in Japan August 1998 Edition 2 E0897 ...

Страница 2: ...the purchase price upon prompt return of the product The Hewlett Packard products may contain remanufactured parts equivalent to new in performance or may have been subject to incidental use The warranty period begins on the date of delivery or on the date of installation if installed by Hewlett Packard If customer schedules or delays Hewlett Packard installation more than 30 days after delivery w...

Страница 3: ...t this product met its published specifications at the time of shipment from the factory Hewlett Packard further certifies that its calibration measurements are traceable to the National Institute of Standards and Technology NIST to the extent allowed by the Institute s calibration facility and to the calibration facilities of other International Standards Organization members ...

Страница 4: ...n personnel must not remove instrument covers Component replacement and internal adjustments must be made by qualified maintenance personnel Do not replace components with power cable connected Under certain conditions dangerous voltages may exist even with the power cable removed To avoid injuries always disconnect power and discharge circuits before touching them DO NOT SERVICE OR ADJUST ALONE D...

Страница 5: ...A connection to the frame chassis of the equipment which normally includes all exposed metal structures Indicates earth ground terminal A Alternating current F Direct current ON Supply OFF Supply STANDBY Supply WARNING The warning sign denotes a hazard It calls attention to a procedure practice condition or the like which if not correctly performed or adhered to could result in injury or death to ...

Страница 6: ......

Страница 7: ...ices DUTs to measurement units HP 16442A has thirteen channels 6 source monitor unit SMU channels Up to 3 channels can be connected to SMUs by Kelvin connections 2 voltage source unit VSU channels 2 voltage monitor unit VMU channels 2 pulse generator unit PGU channels 1 ground unit GNDU channel This section provides the following procedures which you must perform before using the HP 16442A 1 Inspe...

Страница 8: ...ble black 16442 61603 3 8 pin plug pin plug connection cable red 16442 61604 3 9 pin plug pin plug connection cable blue 16442 61605 3 10 miniature banana miniature clip connection cable black 16442 61606 3 11 miniature banana miniature clip connection cable red 16442 61607 3 12 miniature banana miniature clip connection cable blue 16442 61608 3 13 miniature banana miniature banana connection cabl...

Страница 9: ...a flathead screw into hole of each stabilizer After installing stabilizers connect terminals on the rear panel to the desired measurement units Note When you attach HP 16442A to HP 16440A SMU Pulse Generator Selector or HP 16441A R Box you do not need to install stabilizers For attaching to selector or R Box procedure see the User s Guide for selector or R Box HP 16442A User s Gude 3 ...

Страница 10: ...ck Intlk terminals of measurement instrument and HP 16442A before performing measurements Warning Do not operate HP 16442A while the High Voltage indicator is lit Dangerous voltages of up to 200 V may be present on the connection between measurement units and DUT Warning Do not get connection cables caught in the lid of HP 16442A during measurements Dangerous voltage of up to 200 V may be present ...

Страница 11: ...uring operation If more than 40 V will be forced to HP 16442A be sure to connect Intlk terminal of HP 16442A to Interlock terminal of measurement instrument using an Interlock LED cable HP 16493J option 001 for 1 5 m HP 16493J option 002 for 3 0 m When the lid of HP 16442A is open two switches of the Intlk terminal are open The instrument recognizes that the lid of HP 16442A is not closed and does...

Страница 12: ...The following figure shows pin assignment of the Intlk terminal 6 HP 16442A User s Gude ...

Страница 13: ...d of HP 16442A 3 Check that the black plastic fasteners of the socket module are unlocked pulled up 4 Push the socket module onto HP 16442A 5 Press the black fasteners down to lock the socket module into place Use the following procedure to remove a socket module from HP 16442A 1 Confirm that voltage or current is not forced to HP 16442A 2 Open the lid of HP 16442A 3 Disconnect the connection cabl...

Страница 14: ...es to connect the FORCE terminals on HP 16442A to the FORCE terminals on the socket module 2 Kelvin Connections Socket Modules for Power Devices The three socket modules for power devices have separate FORCE and SENSE terminals for Kelvin connections As shown in the example in Figure 1 2 use miniature banana miniature banana connection cables to connect the FORCE and SENSE terminals of HP 16442A t...

Страница 15: ...s of HP 16442A to the FORCE terminal on the socket module as follows insert one pin plug into the FORCE terminal of the socket module then insert the other pin plug into the tail hole of the inserted pin plug Figure 1 3 Mounting DUT on Socket Module for Low Power Devices To connect sense terminals as near as possible to the DUT connect the SENSE terminals of HP 16442A directly to the DUT with the ...

Страница 16: ...e the DUT on the blank Teflon board Universal Socket Module The following figure shows an example of a universal socket module You solder the connection pins into the universal socket then insert the pin plug of the connection cable into the connection pin Note that hydrochloric acid residue deposited by solder may cause electrical problems Hewlett Packard recommends that you solder the wiring on ...

Страница 17: ...dure Chassis lid inner plate and socket modules 1 Make sure that voltage or current is not forced to HP 16442A 2 Disconnect all cables from HP 16442A 3 Using lint free paper gently wipe the chassis lid inner plate and socket module For any area that will not come clean dip the lint free paper into alcohol and wipe the area gently Connection cables 1 Make sure that voltage or current is not forced ...

Страница 18: ...k circuit is shorted LED turns off within 1 sec from when interlock circuit is open 8 To stop the interlock test select NNNNNNNNNNNNNN STOP secondary softkey Servicing the Test Fixture This section provides information for trained service personnel to repair the test fixture Replaceable Parts provides the HP part numbers of replaceable parts When a part needs to be replaced order the parts from th...

Страница 19: ...ering make sure that there are no lint bridges so that leakage current is minimal Keep the following items in mind when repairing socket modules Place a ferrite bead at the 8 mm point of the terminal on the socket side and fix the bead using epoxy adhesive Never allow the ferrite beads to make contact with each other For the three socket modules used for power devices wrap the contact of the wire ...

Страница 20: ...Replaceable Parts Figure 1 5 View of Cable Connections in the Test Fixture 14 HP 16442A User s Gude ...

Страница 21: ...ut 6 8150 4684 32cm 1 Wire 24 Orange 7 8150 4682 32cm 1 Wire 24 Brown 8 8150 4683 32cm 1 Wire 24 Red 9 8150 4685 32cm 1 Wire 24 Yellow 10 8120 0102 22cm 2 15cm 6 Coaxial Cable 50Ω 11 8150 0462 40cm 1 Wire 24 White Green 12 8150 0458 30cm 8 Wire 24 White Brown 13 8150 0462 32cm 1 Wire 24 White Yellow 14 8120 0122 16cm 1 Coaxial Cable 15 8120 0122 16cm 1 Coaxial Cable 16 8150 0456 5cm 4 Wire 24 Whit...

Страница 22: ...ectors Reference Designation HP Part Number Quantity Description 1 1250 1906 7 Triaxial Connector 2 1250 0083 6 BNC Connector 3 2190 0016 6 Lock Washer 4 0360 1190 2 Terminal Washer 5 2950 0001 6 Nut 6 5000 4218 1 Terminal Washer 7 1252 1419 1 6 pin Connector 16 HP 16442A User s Gude ...

Страница 23: ... Parts List for Connections between Lid and Body Reference Designation HP Part Number Quantity Description 1 16442 24001 1 Sleeve 2 0515 1973 1 Bolt M4 3 16442 61624 1 Cable Assy 4 0515 1550 1 Screw M3 L 8 P H 5 16442 23001 1 Pin HP 16442A User s Gude 17 ...

Страница 24: ...Figure 1 8 TO 3 or TO 66 Package Socket Module Figure 1 9 In Line Package Socket Module Figure 1 10 Axial Lead Package Socket Module 18 HP 16442A User s Gude ...

Страница 25: ...Figure 1 11 TO 4 pin Package Socket Module Figure 1 12 Dual In Line 28 pin Package Socket Module Figure 1 13 Dual In Line 18 pin Package Socket Module HP 16442A User s Gude 19 ...

Страница 26: ...Figure 1 14 TO 8 10 and 12 pin Package Socket Module Figure 1 15 Universal Socket Module 20 HP 16442A User s Gude ...

Страница 27: ... 5 C and 50 relative humidity Functions HP 16442A lets you test various shaped DUTs HP 16442A has 6 SMU channels when using non Kelvin connections 3 SMU channels when using Kelvin connections 2 VSU channels 2 VMU channels 2 PGU channels 1 GNDU channel interlock Intlk 6 pin connector Voltage and current range Channel Maximum Voltage Maximum Current SMU 200 V 1 A VSU 40 V 100 mA VMU 40 V PGU 40 V 20...

Страница 28: ...re clip connection cable black 16442 61607 3 miniature banana miniature clip connection cable red 16442 61608 3 miniature banana miniature clip connection cable blue 16442 61609 3 miniature banana miniature banana connection cable black 16442 61610 3 miniature banana miniature banana connection cable red 16442 61611 3 miniature banana miniature banana connection cable blue 16058 60004 1 blank Tefl...

Страница 29: ...Pin Plug Pin Plug Miniature Banana Miniature Clip Miniature Banana Miniature Banana HP 16442A User s Gude 23 ...

Страница 30: ...Socket modules furnished Blank Teflon Board Dual In Line Packages 28 pin 24 HP 16442A User s Gude ...

Страница 31: ...Axial lead package TO Package 12 pin HP 16442A User s Gude 25 ...

Страница 32: ...6442 61602 812 adds 6 pin plug pin plug connection cables black 2 16442 61603 red 2 16442 61604 blue 2 16442 61605 813 adds 6 miniature banana miniature clip connection cable black 2 16442 61606 red 2 16442 61607 blue 2 16442 61608 814 adds 6 miniature banana miniature banana connection cables black 2 16442 61609 red 2 16442 61610 blue 2 16442 61611 821 adds TO packages socket module 4 pin 16088 6...

Страница 33: ...Connection pins optional Socket modules optional TO package TO 3 or TO 66 In Line package 3 pin HP 16442A User s Gude 27 ...

Страница 34: ...Dual In Line package 18 pin TO package 4 pin 28 HP 16442A User s Gude ...

Страница 35: ...TO package 8 pin TO package 10 pin HP 16442A User s Gude 29 ...

Страница 36: ...Universal socket module 30 HP 16442A User s Gude ...

Страница 37: ...e 40 C to 70 C Operating Humidity 5 to 80 relative humidity at no condensation Storage Humidity 5 to 90 relative humidity at 65 C Weight Approximately 2 5 kg Dimensions W H D 260 mm 140 mm 260 mm not including stabilizers 260 mm 140 mm 320 mm including stabilizers HP 16442A User s Gude 31 ...

Страница 38: ...Residual Resistance 60 mΩ typical PGU channel Characteristic Impedance 50 Ω typical GNDU channel Residual Resistance force 40 mΩ typical Residual Resistance sense 40 mΩ typical Socket Board Parts number Description Max Voltage Residual Resistance 16088 60001 2 Tr 200 V 16088 60002 28pin DIP 230 V 16088 60003 18pin DIP 230 V 16088 60004 8pin TO 230 V 16088 60005 10pin TO 200 V 16088 60006 12pin DIP...