Model 1l049A
b. Using voltmeter as a monitor, slowly adjust
power supply output to 0.9 volts lor 1l050A
or 2. 85 volts for 1l049A.
c. Conneat voltmeter to thermoelement. Output
should be 6.5 mV ±O. 3 mV.
If
output is high,
increase R2; if output is low, decrease R2.
30. An alternative but not as accurate method of
selecting R2 is to assume that the thermoelement
resistunce is 90
(l
and select R2 as 510
(l
for the 11049A
and 110
[l
for the 1l050A. If this method is used, the
nominal 7. 5 mV output may be high or low by as much
as 10%.
POWER SUPPLY
hp72IA
DC
VOLTMETER
hp 419A
+ -
0
,:)
'NPul
OuTPuT
Figure
7.
Selection of R2
31. SELECTION OF R1.
32. RI shunts the series combination of the thermo
element and R2 to adjust the input resistance to 50
Q
±O. 15
Q
and is made up of two approximately equal
resistors connected in parallel. In the 11049A Rl is
about 55
[2,
in the Il050A Rl is about 65
Q,
and in the
POWER SUPPLY
hp"i2IA
5Qn�Q.I%
0698-5418
Page 8
200ntO.I'/o
0811-2162
Figure 8. Selection of Rl
lhERMO'
El.EMENT
1105lA Rl is about 110
Q,
The resistors used for R1
should be 1/2 watt, ±5%, metal film resistors.
33. Figure 8 shows a bridge for selecting Rl. R1 is
adjusted for null with the drive voltage set so that the
thermoelement operates at near rated full scale input.
Proceed as follows.
a. With the power supply off, construct the
bridge shown in Figure 8.