Adjust oscillator output so that null meter in
dication is half way between indications re
corded in steps d and e. This sets oscillator
output to 1 volt rms.
g. Record error of instrument under test.
DO NOT CHANGE OSCILLATOR FREQUENCY
RANGE WITHOUT FIRST LOWERING OSCIL
LATOR OUTPUT. SWITCHING TRANSIENTS
MAY DESTROY THERMOELEMENT.
h. Adjust frequency to next desired test frequency.
Then readjust oscillator output for same in
dication as in step f. Record error of instru
ment under test.
i. Repeat steps g through i for each test fre
quency deSired.
22.
MAINTENAN CE.
23. THERMAL CONVERTER CALIBRATION.
24. Figure 6 shows the setup for calibrating a thermal
converter to an NBS calibrated reference thermal con
verter. With this setup, the thermal converter can
be calibrated to"\vithin
±O.
005% of the reference thermal
converter. The following steps describe the calibration
procedure.
a. Sct test oscillator and both reference supply
outputs to minimum.
NBS REFERENCE
THERMAL CO NVERTER
�,,',
Page 6
NULLMETER
hp
419A
NU LLMETER
hp
419A
I�i@
�'"""
'"",'"'�
U NDER TEST
Figure 6. Thermal Converter Calibration
b. Connect both thermal converters, both ref
erence supplies, both null meters, and the
test oscillator as shown in Figure 6. Keep
connections to thermal converters as short as
possible.