Cleaning & Preventive Maintenance
In between patient uses, the AirWedge should be wiped down with a
cleaning solution used by your hospital for medical equipment disinfection.
A 10:1 bleach solution (10 parts water: one part bleach) or disinfectant
wipes can also be used. It is important to follow the cleaning solution
manufacturer’s instructions for use, including dwell time and saturation.
NOTE: Cleaning with bleach solution may discolor fabric.
Do not launder the AirWedge.
INFECTION CONTROL
Whatever the patient is lying on to keep the surface clean may be placed
on top of the AirWedge to help keep it clean. If desired, a disposable sheet
may be used to cover the AirWedge (available for separate purchase).
If the AirWedge is used on an isolation patient, the hospital should employ
the same protocols/procedures it utilizes for the bed mattress and/or for
linens in that patient room.
PREVENTIVE MAINTENANCE
Prior to use, a visual inspection should be performed on the AirWedge
to ensure that there is no visible damage that would render the
AirWedge unusable.
The AirWedge should be periodically inspected to ensure the following:
• All inflation/deflation valves are self-sealing with no evidence of leakage.
• All rails straps are attached to the AirWedge.
• All snaps are intact and operational.
• There are no punctures or tears in the AirWedge.
If any damage is found that would cause the AirWedge not to function as
intended, the AirWedge should be removed from use and returned to
HoverTech International for repair.