Hovertech AirWedge Скачать руководство пользователя страница 5

 

AWManual, Rev. B

 

www.HoverMatt.com  |  5

User Manual

Cleaning & Preventive Maintenance

In between patient uses, the AirWedge should be wiped down with a  

cleaning solution used by your hospital for medical equipment disinfection. 

A 10:1 bleach solution (10 parts water: one part bleach) or disinfectant  

wipes can also be used. It is important to follow the cleaning solution  

manufacturer’s instructions for use, including dwell time and saturation. 

NOTE: Cleaning with bleach solution may discolor fabric.

Do not launder the AirWedge.

INFECTION CONTROL

Whatever the patient is lying on to keep the surface clean may be placed 

on top of the AirWedge to help keep it clean. If desired, a disposable sheet 

may be used to cover the AirWedge (available for separate purchase).

If the AirWedge is used on an isolation patient, the hospital should employ 

the same protocols/procedures it utilizes for the bed mattress and/or for 

linens in that patient room.

PREVENTIVE MAINTENANCE

Prior to use, a visual inspection should be performed on the AirWedge  

to ensure that there is no visible damage that would render the  

AirWedge unusable.

The AirWedge should be periodically inspected to ensure the following:

•  All inflation/deflation valves are self-sealing with no evidence of leakage.
•  All rails straps are attached to the AirWedge.
•  All snaps are intact and operational.
•  There are no punctures or tears in the AirWedge.

If any damage is found that would cause the AirWedge not to function as 

intended, the AirWedge should be removed from use and returned to  

HoverTech International for repair.

Product Specifications/Required Accessories

Содержание AirWedge

Страница 1: ...ioning Device 3 Part Identification HT Air Air Supply 3 HT Air Keypad Functions 3 Instructions for Use 4 Product Specifications Required Accessories 4 Cleaning Preventive Maintenance 5 Warranty Statem...

Страница 2: ...ssional before positioning patients using the AirWedge Make sure patient is centered on the AirWedge before inflating Caregiver s must ensure the patient is attended and stabilized while on the AirWed...

Страница 3: ...nflation speed by the number of flashes i e two flashes equals the second inflation speed STANDBY Used to stop inflation air flow Amber LED indicates STANDBY mode Part Identification AirWedge Adjustab...

Страница 4: ...low Move hose to chest inflation valve and inflate to desired height Then finish inflating head pillow as needed 4 The firmness can be reduced by pressing the diaphragm in each inflation valve to slow...

Страница 5: ...e sheet may be used to cover the AirWedge available for separate purchase If the AirWedge is used on an isolation patient the hospital should employ the same protocols procedures it utilizes for the b...

Страница 6: ...nty This warranty does not cover normal wear and tear Component parts particularly any optional equipment valve caps their attachments and cords will show wear with use over time and eventually may ne...

Страница 7: ...he repair time HTI s warranty covers manufacturer s defects in material and workmanship If a repair is not covered under warranty a minimum repair fee of 100 per item will be assessed plus return ship...

Страница 8: ...User Manual 8 HoverTech International AWManual Rev B 4482 Innovation Way Allentown PA 18109 800 471 2776 Fax 610 694 9601 www HoverMatt com Info HoverMatt com...

Отзывы: