3
5.2
Temperature display check
1.
Select Temp Set with MODE key on the Simulator.
2.
Start outputting simulation input value with OUTPUT
key.
3.
Input temperature simulation value in the order of
0.0°C, 25.0°C, 30.0°C, 60.0°C and 100.0°C to the
Simulator, then check the value on the instrument.
6
Specifications
Note
This Simulator supplies simulation input specific to HORIBA
conductivity meter. Therefore, the simulator output cannot be
measured by general-purpose digital multi-meter.
Output
Output
parameter
Output value
Accuracy (Varies depending on
temperature used.)
±5°C of the
temperature in X-
52 delivery
inspection(*)
5°C to 40°C
COND
Span
19.00 S/m
±0.1 S/m
±0.8 S/m ±1digit
1.900 S/m to
1.900 mS/m
±0.1% ± 1digit
±0.2% ±1digit
COND
Linear
10.00 S/m
±0.1 S/m
±0.4 S/m ±1digit
1.000 S/m to
10.00
µ
S/m
±0.1% ± 1digit
±0.2% ± 1digit
0.000 mS/m
-
-
Tempera-
ture
0.0/ 25.0/
30.0/ 60.0/
100.0
°
C
±0.1
°
C ±0.2
°
C
Ambient
temp.
5°C to 40°C
Ambient
humidity
0% to 80% (No dew condensation)
Power
Two AA alkaline dry-cell batteries (Approx. 20 hours with
continuous operation)
Dimensi
ons
170(H)
×
89(W)
×
40(D) mm
(*):
Temperature in X-52 delivery inspection is described
in the inspection report
.
1. Select Temp Set.
2. Output simulation
3. Toggle simulation
input value.
input value.