Function
92
DATA OUTPUT SET
Specifies the data output.
CAL. OUTPUT SET
Specifies the TP/TN/COD concentration analog output during calibration.
Operation procedure
1.
Press [DATA OUTPUT SET] in the OUTPUT SET screen.
The DATA OUTPUT screen will be displayed.
2.
Press [CAL. OUTPUT SET].
The CAL. OUTPUT screen will be displayed.
3.
Specify the calibration output.
Item
Description
Reference
Specifies the TP/TN concentration analog output during calibration. Page 92
Specifies the TP/TN concentration analog output during alarm
activation.
Specifies the TP/TN concentration analog output during
maintenance.
WARMUP OUTPUT SET
Specifies the TP/TN concentration analog output during warm-up.
Operation mode
Setting range
Default value
Description
Output Mode
Hold
Hold
Locks TP/TN concentration analog output to the
previous value.
Preset
Locks TP/TN concentration analog output to a
preset value.
Direct
Locks TP/TN concentration analog output to the
previous value.
Preset
0 mA to 22 mA
22 mA
Specifies the preset value for the TP/TN
concentration analog output.
Main
MENU
SETTING
OUTPUT SET
[MENU]
[OUTPUT SET]
[SETTING]
Refer to “ To display the [MENU] item screen” (page 62).