Function
96
WARMUP OUTPUT SET
Specifies the warm-up output.
Operation procedure
1.
Press [DATA OUTPUT SET] in the OUTPUT SET screen.
The DATA OUTPUT screen will be displayed.
2.
Press [WARMUP OUTPUT SET].
The WARMUP OUTPUT screen will be displayed.
3.
Specify the warm-up output.
Item
Setting range
Default
value
Description
Output Mode
Hold
Hold
Locks the TP/TN concentration analog output during warm-up to the
previous value.
Preset
Locks the TP/TN concentration analog output during warm-up to the
preset value.
Direct
Outputs the TP/TN concentration analog output during warm-up to
the measurement value.
Preset
0 mA to 22 mA
22 mA
Specifies the preset value for the TP/TN concentration analog output
during warm-up.
Main
MENU
SETTING
OUTPUT SET
[MENU]
[OUTPUT SET]
[SETTING]
Refer to “ To display the [MENU] item screen” (page 62).
[DATA
DATA OUTPUT
OUTPUT SET
[WARMUP
Preset
Output
Output Mode
Preset
WARMUP OUTPUT
<Input a value>
OUTPUT
SET]
OUTPUT SET]
Mode
[ ]
[ ]