Function
94
ALARM OUTPUT SET
Specifies the alarm output.
Operation procedure
1.
Press [DATA OUTPUT SET] in the OUTPUT SET screen.
The DATA OUTPUT screen will be displayed.
2.
Press [ALARM OUTPUT SET].
The ALARM OUTPUT screen will be displayed.
3.
Specify the alarm output.
Item
Setting range
Default value
Description
Output Mode
Hold
Hold
Locks TP/TN concentration analog output during alarm
activation to the previous value.
Preset
Locks TP/TN concentration analog output during alarm
activation to a preset value.
Direct
Outputs the measurement value to the TP/TN
concentration analog output during alarm activation.
Preset
0 mA to 22 mA
22 mA
Specifies the preset value for the TP/TN concentration
analog output during alarm activation.
[MENU]
Main
MENU
SETTING
OUTPUT SET
[OUTPUT SET]
[SETTING]
Refer to “ To display the [MENU] item screen” (page 62).
DATA OUTPUT
[DATA
OUTPUT SET
[ALARM
OUTPUT
SET]
OUTPUT
SET]
Output Mode
ALARM OUTPUT
Preset
Output
Preset <Input a value>
Mode
[ ]
[ ]