background image

20

8

Appendix

8.1

Frequently asked questions

Question

Answer

How long is the 
sensor's service 
life?

It depends on measurement samples and conditions.
It should be approx. 1500 measurements for typical samples.
Note that sensor deterioration and failure are not included in 
the warranty.

How can I check the 
sensor's condition?

Measure the provided standard solution to check how much 
the measured conductivity value drifts from the specified 
conductivity. If the measured value drifts, perform calibration.

What shall I do if 
calibration is failed?

Dirt in the measurement cell and electrode is the main cause of 
calibration failure. Perform electrode surface treatment (refer 
to page 8). If calibration is still failed after this, replace the sen-
sor.

Are there any helpful 
tips or precautions 
to be aware for 
measurement?

Ambient air may cause the measured values to fluctuate. 
To reduce environment interference, close the protection 
cover.

When the sample amount is enough, washing the sensor twice 
or so with the sample allows more accurate measurement.

Residue between the protection cover and flat sensor prevents 
accurate measurement. Before measurement of the next 
sample, clean the sensor with tap water and remove moisture.

Can I measure high- 
or low-temperature 
samples?

This product can not measure a sample with a temperature 
outside the meter's operating temperature range (5

q

C to 40

q

C).

The difference between the sample temperature and ambient 
temperature increases the measurement error. Perform mea-
surement after the sample reaches the ambient temperature. 

Can I prepare the 
moistening solution 
myself?

You can prepare the moistening solution by mixing neutral 
detergent with 20 times to 100 times as much water.

The measured value 
does not change 
after changing the 
sample.

If 

 lights steadily, the measured value is locked.

Press the MEAS switch to unlock the value. 
If the value does not change after unlocking, the sensor may 
be damaged. Replace the sensor.

The temperature 
alarm icon blinks 
during 
measurement.

The measuring environment temperature may not meet the 
specified operating temperature (5

q

C to 40

q

C).

When the environment temperature is within the specified 
range and the alarm icon blinks, replace the sensor.

Содержание LaquaTwin B-771

Страница 1: ...CODE GZ0000297065...

Страница 2: ......

Страница 3: ...Refer to page 6 to page 19 for detailed instructions...

Страница 4: ...with a written agreement In any one of the following cases none of the warranties set forth herein shall be extended z Any malfunction or damage attributable to improper operation z Any malfunction at...

Страница 5: ...J Trademarks Generally company names and brand names are either registered trademarks or trademarks of the respective companies R TM symbols may be omitted in this manual...

Страница 6: ...tive 2006 66 EC in the European Union The symbol might be put with one of the chemical symbols below In this case it satisfies the requirements of the Directive 2006 66 EC for the object chemical This...

Страница 7: ...ble protection against harmful interference when the equipment is operated in a commercial environment This equipment generates uses and can radiate radio frequency energy and if not installed and use...

Страница 8: ......

Страница 9: ...e 8 4 5 Electrode surface treatment 8 5 Calibration 9 5 1 About calibration 9 5 2 One point calibration 10 5 3 Two point calibration 11 6 Measurement 12 6 1 Sample setting 12 6 2 Measurement operation...

Страница 10: ......

Страница 11: ...e sensor for the first time perform electrode surface treatment as instructed on page 8 1 2 Consumable parts sold separately Meter model B 771 Sensor S070 1 Meter B 771 1 Storage case 1 Batteries CR20...

Страница 12: ...u2 4 MEAS switch Switches the calibration mode to the measurement mode activates deactivates the reading locking function in the measurement mode and starts applies settings in the special setting mod...

Страница 13: ...Blinks when the measuring environment temperature does not meet the specified operating temperature 5qC to 40qC 13 Stability icon Lights up when the measured value is stabilized 14 Measured value disp...

Страница 14: ...ng gasket is clean and undamaged The waterproofing gasket is seated properly in the groove with no twisting or warping The meter and sensor are not deformed z Neither the meter nor sensor is waterproo...

Страница 15: ...ed value is unstable J Battery z Keep batteries out of reach of children If someone accidentally swallows a battery call a doctor immediately z Do not throw batteries in fire z Do not attempt to recha...

Страница 16: ...ies at once Do not use old and new batteries together J Inserting the batteries 1 Slide both batteries into the battery case as shown Be sure to use two CR2032 batteries and put them with the plus sid...

Страница 17: ...z When removing the sensor do not let any water get inside the meter If some moisture remains on the waterproof gasket wipe it off carefully J Attaching the sensor 1 Confirm that the waterproofing gas...

Страница 18: ...power is turned ON and the meter model number is displayed on the LCD J Power OFF 1 Press and hold the ON OFF switch for over 2 seconds The power is turned OFF Note Before using the sensor for the fir...

Страница 19: ...of the standard solution or a sample into the measurement cell without bubbles inside If not the measurement may be inaccurate z If remains blinking and Err error display appears the calibration is f...

Страница 20: ...solution beforehand may provide more accurate calibration 3 Close the protection cover and press the CAL switch for over 2 seconds and blink and the calibration value is displayed After the calibratio...

Страница 21: ...layed After the calibration is completed and stop blinking and light steadily 4 After the calibration for 1 41 mS cm is completed open the protection cover and remove the 1 41 mS cm standard solution...

Страница 22: ...mmersing it completely If the product is accidentally dropped into water take it out of water and remove the moisture on it z Put an appropriate amount of the standard solution or a sample into the me...

Страница 23: ...gently 2 or 3 times J Scooping 1 Open the protection cover 2 Immerse the sensor into the sample and stir gently 2 or 3 times and then scoop up some of the sample with the sensor 3 Place the meter flat...

Страница 24: ...at the meter is in the measurement mode and set a sample on the sensor 2 Read the displayed value when appears J With using the reading locking function 1 Confirm that the meter is in the measurement...

Страница 25: ...in turn 7 1 Measurement unit setting page 16 Used to select measurement unit 7 2 Temperature display mode page 17 Displays the ambient temperature measured using an internal temperature sensor 7 3 Ca...

Страница 26: ...asurement mode is returned Used to select measurement unit 1 Press and hold the MEAS switch for over 3 seconds in the measurement mode to enter the special setting mode All items appear on the LCD and...

Страница 27: ...d hold the MEAS switch for over 3 seconds in the measurement mode to enter the special setting mode All items appear on the LCD and then the dis play changes as shown right 2 Press the CAL switch unti...

Страница 28: ...mode is returned 1 Press and hold the MEAS switch for over 3 seconds in the measurement mode to enter the special setting mode All items appear on the LCD and then the dis play changes as shown right...

Страница 29: ...wn right 2 Press the CAL switch until Init appears 3 Press the MEAS switch for 0 5 seconds appears 4 Press the CAL switch for over 2 seconds All settings and all calibration data are reset to the fact...

Страница 30: ...ple allows more accurate measurement Residue between the protection cover and flat sensor prevents accurate measurement Before measurement of the next sample clean the sensor with tap water and remove...

Страница 31: ...n OFF and ON again refer to page 19 If Er1 is still displayed after the initialization the internal IC in the meter defects Replace the product with new one the meter can not be repaired Er2 is displa...

Страница 32: ...y life Approx 400 hours in continuous use Main materials ABS epoxy Outer dimensions mass 164 mm 29 mm 20 mm excluding projections Approx 45 g meter only without batteries Main functions Temperature co...

Страница 33: ......

Страница 34: ...2 Miyanohigashi Kisshoin Minami ku Kyoto 601 8510 Japan http www horiba com...

Отзывы: