![horiba LAQUAact-EC110 Скачать руководство пользователя страница 91](http://html1.mh-extra.com/html/horiba/laquaact-ec110/laquaact-ec110_instruction-manual_2162260091.webp)
When an error message appears
75
How to resolve errors or troubles
This section describes the causes of typical problems and the actions to be taken,
including questions frequently asked by customers. Check these before contacting us.
■
When an error message appears
If "ERROR No.00XX" is displayed while you are using the instrument, check the error in
the error list below to check the cause and action to be taken.
●
ERROR No.0001
Memory error
Data cannot be read from or written to the internal memory.
●
ERROR No.0006
Maximum calibration points exceeded
Attempted to perform 6th point calibration during conductivity calibration.
ERROR No.
Description
Definition of error
0001
Memory error
Data cannot be read from or written to the internal
memory.
0006
Maximum calibration
points exceeded
6th point calibration is attempted.
0009
Printer error
There is a problem with the printer.
0010
Memory full
The number of the data saved has exceeded the
limit of the internal memory.
0011
Cell constant is out of
range
Cell constant is out of automatic calculation range.
Cause
How to solve problem
The instrument does not start
properly due to noise or other at
power ON.
Disconnect the AC adapter, and then press the
key.
The defect of the internal IC
Contact your dealer for repair.
Cause
How to solve problem
6th point calibration is
attempted.
Up to five points can be calibrated.
Содержание LAQUAact-EC110
Страница 1: ...Handheld Conductivity Meter LAQUAact EC110 LAQUAact EC120 Instruction Manual CODE GZ0000427797...
Страница 2: ......
Страница 12: ...X M E M O...
Страница 16: ...XIV M E M O...
Страница 60: ...M E M O 44...
Страница 88: ...M E M O 72...
Страница 96: ...M E M O 80...
Страница 104: ...M E M O 88...