Calibrating temperature sensor
41
■
Temperature settings
●
Calibrating temperature sensor
The temperature sensor or temperature compensation electrode in the combination
electrode has ±1°C accuracy without calibration. You can use a known temperature
solution to calibrate the temperature sensor.
1. Insert the temperature connector into the temperature connector (T) on the
instrument.
2. Immerse the electrode into the solution until the temperature sensor is
immersed.
3. Press the
key to enter the setting mode.
4. Press the
keys to select "TEMP" (temperature calibration setting) and
then press the
key.
5. Make sure that "ATC" is displayed, and press the
key.
The temperature setting screen is displayed.
6. Enter the set temperature and press the
key to confirm.
The temperature sensor is calibrated.
key
key
key
Enter a
number
key