Using the automatic data save

34

4.  Enter the period of saving in minutes  and press the 

 key.

A period from 0 min to 3600 min can be set.
"0" indicates no period.

5.  Press the 

 key to enter the measurement mode.

Enter a 

number

 key

Enter a 

number

 key

 key

Содержание LAQUAact-DO110

Страница 1: ...Handheld Dissolved Oxygen Meter LAQUAact DO110 LAQUAact DO120 Instruction Manual CODE GZ0000427798...

Страница 2: ......

Страница 3: ...ble to improper operation Any malfunction attributable to repair or modification by any person not authorized by HORIBA Ltd Any malfunction or damage attributable to the use in an environment not spec...

Страница 4: ...case it satisfies the requirements of the Directive 2006 66 EC for the object chemical This product should not be disposed of as unsorted household waste Your correct disposal of WEEE waste batteries...

Страница 5: ...o provide reasonable protection against harmful interference when the equipment is operated in a commercial environment This equipment generates uses and can radiate radio frequency energy and if not...

Страница 6: ...result in death or serious injury This is to be limited to the most extreme situations This indicates a potentially hazardous situation which if not avoided could result in death or serious injury Thi...

Страница 7: ...therwise it may heat up or be ignited resulting in a fire or an accident Do not disassemble or modify the instrument Otherwise it may heat up or be ignited resulting in a fire or an accident CAUTION H...

Страница 8: ...to the eyes do not rub the eyes rinse with clean water immediately and then consult a doctor Contact with alkaline fluid could cause blindness Do not put batteries in a fire expose to heat disassemble...

Страница 9: ...the instrument provides dust proof and waterproof performance When replacing the batteries with an AC adapter or a serial cable connected the instrument does not have the dust proof and waterproof per...

Страница 10: ...ogether with a used battery Do not use a fully charged nickel metal hydride battery together with a partially charged battery Do not attempt to charge a non rechargeable battery Environmental conditio...

Страница 11: ...necessary points for correct operation and notifies the important points for handling the product This indicates the part where to refer for information This indicates reference information Descriptio...

Страница 12: ...X M E M O...

Страница 13: ...ion 7 Function layer 7 Changing the operation mode 9 Changing the measurement parameter 10 Using the backlight 11 Entering numeric values 11 Saving measurement data in the internal memory 12 Measureme...

Страница 14: ...ng the automatic power off setting default 30 min 45 Resetting to factory default settings 47 Setting the date and time 49 Performing test printing of the printer unit 51 Other settings 53 Printing me...

Страница 15: ...ponse is slow 63 The indicated value does not change No response 63 The measured value is out of the measurement range 64 Repeatability of the measured value is poor 64 Nothing appears when the power...

Страница 16: ...XIV M E M O...

Страница 17: ...and basic operation method of each part Names of each part 2 Instrument 2 Display 4 Operation key 6 Basic operation 7 Function layer 7 Changing the operation mode 9 Changing the measurement parameter...

Страница 18: ...nnects the BNC connector of the electrode 4 Temperature connector T Connects the temperature connector T of the electrode 5 Battery cover Set batteries inside 6 Electrode hook attachment section Attac...

Страница 19: ...he instrument Third number from the left in the MFG No indicates manufacturing year Forth alphabet from the left in the MFG No indicates manufacturing month The alphabet is assigned to month according...

Страница 20: ...the currently set measurement parameter 7 HOLD icon Lights when the measured value display is fixed 8 SET icon Lights when numerical values are entered 9 Temperature display area Displays the measured...

Страница 21: ...isplay Battery level is high Battery level is a little lower Battery level is low The backlight may become unavailable Battery has run out Replace the batteries or use AC adapter option ERR No 0002 is...

Страница 22: ...ET key Changes from the measurement mode to the setting mode Starts repeatability inspection after calibration is complete POWER key Turns ON OFF the power of instrument MODE key In the measurement mo...

Страница 23: ...mode and setting mode is shown as below dX and PXX indicates the program number which is shown in the screen of the instrument Data mode Screen Layer Description d1 DATA OUT Saved data display d2 AUT...

Страница 24: ...PRES KPA Air pressure correction setting P13 CAL DATA Calibration data display P14 CAL CLR Deletion of calibration data P2 TEMP Temperature setting Selection of temperature conversion or display ATC M...

Страница 25: ...tion mode data mode or setting mode is possible only from the measurement mode When changing to a different mode first change to the measurement mode and then change to the desired mode Icon Name Func...

Страница 26: ...multiple parameters For measurement an electrode corresponding to the measurement parameter is required In the measurement mode the measurement parameter can be changed by pressing the key This opera...

Страница 27: ...y press the key again while the backlight is on This function is available for DO120 Note Turning on the backlight consumes energy and shortens battery life The backlight becomes unavailable when the...

Страница 28: ...instrument is in the measurement mode 1 Press the key while the data to save is displayed The saved data is displayed for two seconds and then the display returns to the previous screen automatically...

Страница 29: ...ch measurement parameters Preparation 14 Confirmation before starting measurement 14 Turning ON the instrument 15 Connecting an electrode 18 Dissolved oxygen measurement 19 Setting the instrument 19 P...

Страница 30: ...solution for the measurement parameter If not prepare the standard solution according to your applications Are there any items that should not be wet or stained around the instrument Depending on the...

Страница 31: ...n the battery cover on the back of the instrument clockwise to lock the battery cover Note Do not replace the batteries in a dusty place or with wet hands Dust or moisture could get inside the instrum...

Страница 32: ...strument 2 Insert the AC adapter cable by fitting with the AC adapter jack of in the instrument 3 Insert AC adapter into the electrical socket Note Do not insert the cable with force when the connecto...

Страница 33: ...one second and the screen displays the version number of software and the model and then displays the measurement mode Note Do not hold down the key after the power of the instrument is turned ON Do n...

Страница 34: ...connector pin of the instrument 2 Turn the electrode connector clockwise by following the groove 3 Put the connector cover on the connector 4 When using a combination electrode equipped with a temper...

Страница 35: ...ure of the air when air calibration You can correct the sensitivity of the calibration data according to the sample temperature with this function If you do not use the automatic temperature compensat...

Страница 36: ...calibrate the temperature sensor refer to Calibrating temperature sensor page 41 3 In the case of MTC enter the temperature to be compensated for and press the key to confirm To return to the measure...

Страница 37: ...ue or sea concentration 35 0 PPT If the measured target does not contain saline matter set to 0 0 PPT 1 Press the key to change the setting mode 2 Press the keys to select DO DO correction setting and...

Страница 38: ...ct it at measurement At sea level you can use the instrument with the default setting If you are using the instrument in a high altitude area you must set the air pressure correction 1 Press the key t...

Страница 39: ...Dissolved oxygen measurement 23 4 Enter the air pressure value of the measurement location and then press the key To return to the measurement mode press the key Enter a number key key...

Страница 40: ...o wind and rain or near a heating appliance an error may occur or the value may not be stabilized Do not hold the DO electrode with hand during calibration DO electrode may be affected by temperature...

Страница 41: ...abilized the sat ratio icon changes from the blinking state to the lit state When you press the key or the key the calibrated value is displayed The calibrated value is displayed and then the display...

Страница 42: ...on solution In two point calibration immerse it in the high concentration solution first In order to immerse the temperature sensor in the standard solution make sure to immerse the electrode surely D...

Страница 43: ...n electrode 9551 20D or 9551 100D immerse at least 9 cm from the tip 9 Immerse the DO electrode in the standard solution As with the step 4 immerse at least 6 cm of the tip of the DO electrode Stabili...

Страница 44: ...e measurement mode Calibration using the standard solution is complete Note The zero standard solution and the span standard solution are determined automatically by the instrument If calibration of a...

Страница 45: ...DO electrode in the sample solution In order to immerse the temperature sensor in the standard solution make sure to immerse the electrode surely During measurement use a stirrer to stir the sample s...

Страница 46: ...han 2 0 C Saturated oxygen concentration Display value change for 10 seconds is less than 0 3 and temperature change is less than 2 0 C Ur Under lights when the measured value is below the display ran...

Страница 47: ...calibration data 39 Temperature settings 41 Calibrating temperature sensor 41 General settings 43 Setting the auto stability and auto hold function 43 Changing the automatic power off setting default...

Страница 48: ...ess the key to enter the data mode 2 Press the keys to select DATA OUT display saved data and then press the key Press the keys to change the measurement data and press the keys to change the display...

Страница 49: ...isabled If the batteries run out while using the automatic data save function the data saved until just before the batteries run out Replace the batteries and check the data This function is available...

Страница 50: ...ta save 34 4 Enter the period of saving in minutes and press the key A period from 0 min to 3600 min can be set 0 indicates no period 5 Press the key to enter the measurement mode Enter a number key E...

Страница 51: ...ata saving process During automatic data saving measurement data is displayed for one second each time a measurement takes place When more than 1000 data items are saved ERR No 0010 is displayed and d...

Страница 52: ...te all data saved in the internal memory Data cannot be deleted selectively Copy or transfer the data to a PC for storage if necessary 1 Press the key to enter the data mode 2 Press the keys to select...

Страница 53: ...to delete the saved data or select NO to cancel deleting it And then press the key to confirm When YES is selected END appears after deletion is complete To return to the measurement mode press the k...

Страница 54: ...to enter the setting mode 2 Press the keys to select DO DO correcting setting and then press the key 3 Press the keys to select CAL DATA display calibration data and then press the key The electrode...

Страница 55: ...e the calibration data set in the instrument 1 Press the key to enter the setting mode 2 Press the keys to select DO DO correcting setting and then press the key 3 Press the keys to select CAL CLR del...

Страница 56: ...he key to confirm selection When YES is selected and is pressed END appears after deletion is complete To return to the measurement mode press the key Tip When the key is pressed for more than tree se...

Страница 57: ...emperature connector into the temperature connector T on the instrument 2 Immerse the electrode into the solution until the temperature sensor is immersed 3 Press the key to enter the setting mode 4 P...

Страница 58: ...ote When initializing temperature calibration data all settings need to be initialized Perform initialization by referring to Resetting to factory default settings page 47 When initialization is perfo...

Страница 59: ...is released and displays instantaneous value Auto hold mode displayed as AH When the criterion for stability judgment is fulfilled during measurement the component icon and the HOLD icon light and the...

Страница 60: ...set the auto hold And then press the key to confirm selection To return to the measurement mode press the key Note In the calibration mode the auto stability mode always works Tip The stability judgme...

Страница 61: ...ere is no key operation for a certain period of time This function is disabled during automatic data memory saving or remote operation using an external device 1 Press the key to enter the setting mod...

Страница 62: ...setting default 30 min 46 4 Enter the automatic power off time and press the key The setting range is 0 min to 30 min 0 indicates the automatic power off is OFF To return to the measurement mode pres...

Страница 63: ...a of date and time and the saved data are not deleted Make sure there will be no problems before using this function When this function is used the temperature calibration data is also initialized 1 P...

Страница 64: ...to the factory default settings or select NO to cancel initialization And then press the key to confirm selection When YES is selected END appears after the settings are initialized and then the inst...

Страница 65: ...y If the setting is incorrect the date and time of saved data becomes incorrect This function is available for DO120 1 Press the key to enter the setting mode 2 Press the keys to select GEN general se...

Страница 66: ...e date and time 50 6 Press the key to confirm To return to the setting mode press the key key Enter a number and press key Enter a number and press key Enter a number and press key Enter a number and...

Страница 67: ...ommunication problem you can perform test printing Connect the instrument and a printer correctly and perform the following procedure for test printing As a result of test printing if the printout is...

Страница 68: ...Press the keys to select PRINT test print and then press the key Printing starts automatically When printing ends the printer icon lights and END appears Press the key or the key To return to the meas...

Страница 69: ...out in the measurement mode Reference For details of how to display measurement data and calibration data refer to the respective section Displaying saved data page 32 Displaying the latest calibrati...

Страница 70: ...Temperature value and temperature setting SAL Coef 5 0 PPT Salinity correction coefficient Pres Coef 101 3 kPa Air pressure correction coefficient Printout format Description Inst model LAQUAact DO120...

Страница 71: ...manual which you can download from our website as well The required PC specifications and recommended PC specifications for using the FD 70 are shown in the following table Note If you are not using...

Страница 72: ...ransfer formats are used a communication error occurs and the online mode does not start up and as a result RS 232C communication cannot be performed Also when the transfer format is changed turn OFF...

Страница 73: ...ficult to remove the dirt wipe it gently with a cloth moistened with alcohol The instrument is made of solvent resistant materials but that does not mean it is resistant to all chemicals Do not dip th...

Страница 74: ...membrane in order to prevent deterioration of the membrane Daily storage of the DO electrode Follow the steps below to store the electrode correctly 9520 10D 1 Clean the DO electrode well with pure wa...

Страница 75: ...vel is empty ERR No Description Definition of error 0001 Memory error Data cannot be read from or written to the internal memory 0002 Empty battery level The battery level is empty 0005 Electrode sens...

Страница 76: ...ctly The settings are wrong Set the correct salinity correction air pressure correction and temperature There is liquid on the DO tip membrane when performing air calibration Let the electrode sit unt...

Страница 77: ...e specified number of items Cause How to solve problem Saving more than specified items of data is attempted The maximum number of savable items of data is 100 DO110 or 1000 DO120 Copy or transfer nec...

Страница 78: ...ence electrode until it is higher than the level of the sample Cause How to solve problem There is a motor or other device causing electrical interference Measure at a place where no influence from in...

Страница 79: ...ortant to select an electrode that is appropriate for the sample Consult your dealer To confirm an electrode that is appropriate for the sample check the pH electrode selection guide in our catalogue...

Страница 80: ...de must be immersed up to the liquid junction As a guide immerse to at least 3 cm from the tip of the electrode The electrode cable is broken Replace the electrode Calibration is not performed or perf...

Страница 81: ...AC adapter option Instrument defect Consult your dealer Cause How to solve problem Using the instrument at high elevation or other location where the air pressure is different from sea level To elimi...

Страница 82: ...M E M O 66...

Страница 83: ...adapter 100 V to 240 V 50 60 Hz 0 37 A option Dimensions Approx 67 80 W 170 D 28 42 H mm The figures in square bracket are maximum thicknesses Mass Approx 270 g without batteries Measurement parameter...

Страница 84: ...Selection item Setting range Default values Common Auto hold AS AH AS Temperature input value 0 0 C to 100 0 C 25 0 C Auto power off time 0 min to 30 min 1 30 min Auto data memory time DO120 only 0 2...

Страница 85: ...mg L Temp C Saturated DO mg L 1 13 77 11 10 67 21 8 68 31 7 42 2 13 40 12 10 43 22 8 53 32 7 32 3 13 04 13 10 20 23 8 39 33 7 22 4 12 70 14 9 97 24 8 25 34 7 13 5 12 37 15 9 76 25 8 11 35 7 04 6 12 0...

Страница 86: ...o the representative With regard to electrodes it is important to select the optimum electrode for the sample you want to measure For details refer to the catalogue or our website or contact your deal...

Страница 87: ...hi Kisshoin Minami ku Kyoto 601 8510 Japan http www horiba com Distributed by ABQ Industrial LP USA Tel 1 281 516 9292 888 275 5772 eFax 1 866 234 0451 Web https www abqindustrial net E mail info abqi...

Отзывы: