General settings
65
●
Resetting to factory default settings
The instrument settings can be reset to the factory default settings. The calibration data
is deleted but the data of date and time, and the saved data are not deleted. Make sure
there will be no problems before using this function. When this function is used, the
temperature calibration data is also initialized.
1. Press the
key to enter the setting mode.
2. Press the
keys to select "GEN" (general setting) and then press the
key.
3. Press the
keys to select "RESET" (initialize) and then press the
key.
4. Select "YES" to initialize the settings to the factory default settings, or select
"NO" to cancel initialization. And then press the
key to confirm selection.
When "YES" is selected, "END" is displayed after the settings are initialized and then
the instrument is automatically power OFF.
key
key
key
key
key
Selected "YES"
and press the
key