Preparation
16
●
Connecting an electrode
To perform measurement, it is necessary to use the appropriate electrode for
measurement parameters. Recommended electrodes for each measured sample are
listed in our product catalog and on our website. Refer to them when preparing
electrodes. Use the following procedure to correctly connect the electrode to the
instrument.
• pH or ORP electrode
1. Insert the electrode connector by fitting
its groove with the connector pin of the
CH1 measurement electrode (G).
2. Turn the electrode connector clockwise
by following the groove.
3. Put the connector cover on the
connector.
4. When using a combination electrode
equipped with a temperature sensor,
insert the temperature connector into
the jack of the CH1 temperature
electrode (T).
• Conductivity electrode
Connect conductivity electrode to CH2 in the same way as “ pH or ORP electrode ”.
Tip
If the temperature connector is unconnected or the connection is wrong, temperature
selected for MTC (manual temperature compensation) is displayed as the sample
temperature.
Preparation for measurement is complete.
For details of the measurement operation, refer to the following pages.
CH1 measurement electrode (G)
CH1 temperature electrode (T)