1. Hold the spreader assembly and remove the spreader blade using the thumb and forefinger.
2. Remove the protective paper of the new spreader blade provided with the system.
3. Insert the new spreader blade in the spreader assembly and firmly press.
Check the spread quality after spreader blades replacement.
2.3.
To Replace the Spreader Holders
You need the following elements to perform this procedure:
■
1300029912
The spreader holder kit consists of the following elements:
■
One pair of spreader blades
■
Two spreader holders with attached weights
■
Four "E" rings (two are extras)
1. Remove the two "E" rings holding the spreader holders.
These rings are probably sprung and not reusable.
2. Remove the spreader blades.
3. Place the new spreader holders so that the two "E" rings snap into the groove on the pivot pin.
4. Insert the spreader blades back:
a. Make sure that the previously removed spreader blades are intact. If not, replace with new
ones.
b. If necessary, replace the double-sided tape that holds the blades.
c. Place the spreader blades back to their initial positions.
d. Clean the spreader blades.
Check the spread quality after spreader holders replacement.
Maintenance and Troubleshooting
Replacement Procedures
30
User Manual
Ref: RAB316AEN
Содержание HEMAPREP RAB316AEN
Страница 1: ...Blood Spreading System User Manual Ref RAB316AEN...
Страница 2: ......
Страница 3: ...User Manual User Manual Ref RAB316AEN...
Страница 4: ......
Страница 6: ...ii User Manual Ref RAB316AEN...
Страница 12: ...Foreword Legal Information 8 User Manual Ref RAB316AEN...
Страница 30: ...Workflow Performing the Test 26 User Manual Ref RAB316AEN...
Страница 38: ...Maintenance and Troubleshooting Troubleshooting Procedures 34 User Manual Ref RAB316AEN...
Страница 40: ...36 User Manual Ref RAB316AEN...