2. Replacement Procedures
2.1.
Consumables
Here is the list of consumables you may need on your system:
Designation
Reference
SPREADER BLADES
1300029910
SPREADER HOLDER REPAIR KIT
1300029912
SLIDE TRAYS DOMESTICS
1300029913
SLIDE TRAYS INTERNATIONAL
1300029914
HEMAPREP is made of rugged material and does not require periodic adjustment or lubrication.
For more detailed information, please contact your local HORIBA Medical representative.
2.2.
To Replace the Spreader Blades
You need the following elements to perform this procedure:
■
1300029910
The spreader blades used in the system are manufactured of high-strength glass and designed for
long term operation (several months).
Replace the spreader blades if nicked or cracked because it may cause streaks in the smears.
Maintenance and Troubleshooting
Replacement Procedures
User Manual
Ref: RAB316AEN
29