4 CALIBRATION
25
4.3.4
Starting the AIC sequence with the [AIC] key
1.
Press the [CAL.] key on the MEAS. screen. The CAL. screen will be displayed.
Fig. 33 CAL. screen
2.
Press the [AIC] key. The AIC start message will be displayed.
Fig. 34 AIC start message
3.
Press the [YES] button. The preset AIC sequence will start.
While the AIC sequence is in progress, the CAL. screen is displayed again and the AIC
mode icon blinks. Pressing the [AIC] key in this state displays the AIC abort message.
Fig. 35 AIC abort message
The keys allow you to perform the following operations.
Note
For the AIC sequence and its setting, see “4.3.1 AIC setting” (page 16) to “4.3.3 Setting the AIC
sequence” (page 22).
[YES]:
The ongoing AIC sequence will be aborted.
[NO]:
The ongoing AIC sequence will be continued.