4 CALIBRATION
22
4.3.3
Setting the AIC sequence
To set the AIC sequence, go to the AIC SEQUENCE screen.
1.
Press the [MENU] key on the MEAS. screen.
2.
Press the [
] or [
] key to display the MENU/SETTING screen.
Fig. 28 MENU/SETTING screen
3.
Press the [AIC SEQUENCE] button. The AIC SEQUENCE screen will be displayed.
Fig. 29 AIC SEQUENCE screen
Item
Settable range
Description
WAIT
0 min to 999 min
Set the waiting time for stabilization after changing gas.
The recommended setting is 20 minutes or longer.
HOLD 0 min to 999 min
Set the calibration validation time (to check the readouts on the recorder after finishing the
calibration).
CAL.
YES/NO
Specify whether or not to perform calibration.
YES: Calibration is performed. NO: Calibration is not performed.