ABX Micros
ES60
OT/CT
5–4
User Manual - RAB237AEN
2.2. "QC" tab description
The "
QC
" tab is composed of the following elements:
◆
3 fields, located at the top of the tab, give information on the selected control:
- the "
Name
" field displays the control name.
- the "
Lot number
" field displays the control lot number.
- the "
Expiry date
" field displays the control expiration date.
◆
3 frames containing 1 radio button to select control (See
2.3. Control management, page 5-4
).
It is recommended to use the left frame for the low level control, the middle frame for the normal
level control and the right frame for the high level control.
When the user selects a control, the corresponding control frame is highlighted.
◆
9 icons give the QC results status of each family (See
2.3.1. Control status icons, page 5-4
).
2.3. Control management
This chapter gives information on control status and how to manage them.
◆
2.3.1. Control status icons, page 5-4
.
◆
2.3.2. Control creation, page 5-5
.
◆
2.3.3. Control closing, page 5-8
.
2.3.1. Control status icons
◆
4 different icons inform the user on the control status: See
Tab.5-1 Control status icons, page 5-4
Tab.5-1 Control status icons
Icon
Name
Description
QC expired
The control blood expiration date is reached. Analysis cannot be run on the
expired control but patient samples can be analysed. The control is expired
but its result is valid (until the time limit between two control runs has not
been exceeded).
Inter QC time
limit
The inter QC time limit set has been exceeded between two control runs.
Patient samples can be analysed but the "QCtl" (QC time limit) alarm will
be triggered on the result.
Run an analysis for this control (See Section 4,
).
The instrument automatic lock on Inter QC time limit alarm could be set in
the "
Settings
" menu (See Section 6,
6.5. Analysis running, page 6-14
Invalid QC
The QC is invalid:
- In a parameter family, at least one parameter QC result is out of range.
or
- QC is invalid because flag D, *, $ or ! is triggered.
The instrument automatic lock on invalid QC alarm could be set in the
"
Settings
" menu (See Section 6,
6.5. Analysis running, page 6-14
).
QC passed
The QC has passed.
Содержание ABX MICROS ES60
Страница 1: ...Explore the future ABX Micros ES60 HAN706A 2007 HORIBA ABX User manual P n RAB237AEN...
Страница 2: ......
Страница 6: ...User Manual RAB237AEN...
Страница 68: ...ABX Micros ES60 OT CT 3 12 User Manual RAB237AEN...
Страница 116: ...ABX Micros ES60 OT CT 5 18 User Manual RAB237AEN Diag 5 21 LJ graphs printout example...
Страница 128: ...ABX Micros ES60 OT CT 5 30 User Manual RAB237AEN...
Страница 148: ...ABX Micros ES60 OT CT 6 20 User Manual RAB237AEN...
Страница 196: ...ABX Micros ES60 OT CT 7 48 User Manual RAB237AEN...