pH Electrode
9681-10D Instruction Manual
Thank you for purchasing a HORIBA pH electrode.
Read this manual carefully, before using the electrode.
J
Caution on Use
Safety Precautions
Points of concern
z
Do not shock the electrode.
z
Do not measure chapped or cracked electrode. Replace
such electrode with a new one.
z
Do not allow the connector to come in contact with water
or unclean hands. If not, accurate measurement cannot
be performed.
z
Always use the 3.33 mol/L-KCl solution (model 300)
internal solution for reference electrode.
z
If the responsive glass membrane or liquid junction is
very dirty and cannot be washed clean using pure water,
cleanse it depending on dirt condition.
(Refer to “ Maintenance ” (page 3).)
z
Avoid continuous use of this electrode at acidic or
alkaline concentrations above 0.1 mol/L. Using the
electrode under these conditions may harm
performance and shorten the life of the electrode.
z
Do not measure the solutions which contain fluoride ion,
which may corrode the glass electrode.
z
Do not suddenly immerse the electrode in samples
where the temperature variation is 50
°
C or more.
z
Silver chloride can be detected inside the responsive
glass membrane. However, it will not cause any
performance problem.
z
Do not use the electrode in any place where usable
temperature is out of the specified range.
z
White crystal can adhere to the internal solution filler
port or protective cap after storage of the electrode.
However, it will not cause any performance problem.
z
If the sleeve of the liquid junction is stuck with
crystallized internal solution, immerse the sleeve in hot
water at 30
°
C to 40
°
C to dissolve the crystal, and then
move the sleeve carefully to check the internal solution
discharges. (Use of the ultrasonic cleaner is effective.)
z
Push down the sleeve lightly. If it is pushed down
strongly, potential difference between liquid may be
greater and may cause error of measurement.
J
Packaged Contents
CAUTION
Chemical solution
The internal solution in the electrode uses potassium chloride
(3.33 mol/L-KCI). If the internal material comes in contact with the
hands or skin, wash immediately with water. If the internal solution
comes in contact with the eyes, flush with a large amount of
running water, then seek medical advice.
Glass fragments
Glass fragments can cause injury.
The outer tube and tip of the electrode are made from glass.
Be careful not to break them.
Name
Q’ty
Electrode
9681-10D
1 pc
Attachment holder
1 pc
Instruction manual (this book) 1 booklet
J
Specifications and Parts Description
Part name
Specifications
J
Connecting to pH meter
1. Insert the electrode connector into the connector port
sleeve on the meter, after aligning with the pin. Do not
insert the connector unless it is aligned properly with the
connector port.
2. Press the electrode connector into the connector port on
the meter, while turning the connector to the right.
3. Slide the connector cover over the connector. Then,
push the cover in straight until it comes in light contact
with the meter case. Do not turn the cover.
4. Insert the temperature connector into the jack on the
meter. Insert the connector firmly, until the O-ring on the
connector can no longer be seen.
Note
Be sure to install the temperature connector. If not, accurate
measurement cannot be performed.
Electrode model
9681-10D
Measurement range
pH 0 to 14
Usable temperature range
0
°
C to 60
°
C
Storage temperature range
0
°
C to 50
°
C
Internal electrode
Silver/silver chloride
Reference electrode internal
solution
3.33 mol/L-KCl solution (model 300)
Liquid junction
Sleeve
Liquid junction material
Glass
Wetted material
Glass
Electrode length
203 mm
Ext. diam. of wetted part
12 mm
Liquid junction height
Approximately 26 mm (from the tip of
electrode)
Lead length
1 m
Liquid junction
Protective cap
Stem
Cap
Temp. connector
Electrode connector
Connector cover
Lead
Responsive glass membrane
Internal solution filler port
Knurled nut
①
②
③
④