Main specifications
63
Appendix
This section describes the specifications of the instrument, default settings,
measurement principles, and other technical information.
Options for the instrument are also described.
■
Main specifications
・
Specification of each measurement parameter
Item
Description
Model
D-72G
Brand (pet name)
LAQUAact
Series name
Portable pH
Water Quality Meter
Model description
pH/ORP Meter
Measurement
parameters
pH, mV (ORP), temperature
Operating ambient
temperature, humidity
0°C to 45°C
80% or less in relative humidity (no condensation)
Power
AAA alkaline batteries (LR03) or AAA Ni
MH rechargeable
batteries × 2
AC adapter 100 V to 240 V, 50/60 Hz, 0.37 A (option)
Dimensions
Approx. 67 (80) × 28 (42) × 170 mm
(The figures in parentheses are maximum thicknesses.)
Mass
Approx. 270 g (without batteries)
Measurement
parameter
Item
Description
pH
Measuring principle
Glass electrode
Display range
pH
2.00 to pH 16.00
Measuring range
pH 0.00 to pH 14.00
Resolution
0.01 pH
Repeatability
±0.01 pH ±1 digit
mV (ORP)
Measuring range
±2000 mV
Resolution
1 mV
Repeatability
±1 mV ±1 digit
temperature
Measuring principle
Thermistor method
Display range
30.0°C to 130.0°C
Measuring range
0.0°C to 100.0°C
Resolution
0.1°C
Repeatability
±0.1°C ±1 digit