pH Electrode
9632-10D Instruction Manual
This manual describes the operation of the pH electrode,
9632-10D. Be sure to read this manual carefully, before
using the electrode.
Conformable standards
Authorised representative in EU
HORIBA UK Limited
2 Dalston Gardens, Stanmore, Middx HA7 1BQ, UK
Caution on use
Safety precautions
Points of concern
Do not shock the electrode.
Do not measure chapped or cracked electrode. Replace
such electrode with a new one.
Do not allow the connector to come in contact with water
or unclean hands. If not, accurate measurement cannot
be performed.
Always use the 3.33 mol/L-KCl solution (model 300)
internal solution for reference electrode.
If the responsive glass membrane or liquid junction is
very dirty and cannot be washed clean using pure water
(or deionized water), clean it depending on dirt
condition. (Refer to “ Maintenance ” (page 3).)
Do not use this electrode at acidic concentrations above
0.1 mol/L. Using the electrode under these conditions
may harm performance and shorten the life of the
electrode.
Do not measure the solutions which contain hydrofluoric
acid, which may corrode the glass electrode.
Do not suddenly immerse the electrode in the sample
solution where the temperature variation is 50
C or
more.
Silver chloride can be detected inside the responsive
glass membrane. However, it will not cause any
performance problem.
Do not use the electrode in any place where usable
temperature is out of the specified range.
White crystal can adhere to the internal solution filler
port or protective cap after storage of the electrode.
However, it will not cause any performance problem.
When disposing of the product, follow the related laws
and/or regulations of your country for disposal of the
product.
This equipment conforms to the following standards:
RoHS:
EN50581
9. Monitoring and control instruments
CAUTION
Chemical solution
The internal solution in the electrode uses potassium chloride
(3.33 mol/L-KCI). If the internal material comes in contact with the
hands or skin, wash immediately with water. If the internal solution
comes in contact with the eyes, flush with a large amount of
running water, then seek medical advice.
Glass fragments
Glass fragments can cause injury.
The outer tube and tip of the electrode are made from glass.
Be careful not to break them.
Packaged contents
Specifications and parts description
Part name
Specifications
Note
Body material is plastic. Do not use organic solvents (such as
acetone or toluene).
Name
Q’ty
Electrode
9632-10D
1 pc
Instruction manual (this book) 1 booklet
Electrode model
9632-10D
Measurement range
pH 0 to 14
Usable temperature range
0
C to 100
C (the part from the internal
solution filler port to the tip of the
electrode)
Storage temperature range
0
C to 50
C
Internal electrode
Silver/silver chloride
Reference electrode internal
solution
3.33 mol/L-KCl solution (model 300)
Liquid junction material
Porous ceramic
Wetted material
Glass, butyl rubber, ceramic, polysulfone,
polyolefin, polyamide
Electrode length
150 mm
Ext. diam. of wetted part
16 mm
Liquid junction height
Approximately 15 mm (from the tip of
electrode)
Lead length
1 m