Measure timing: 000.0 s Set continuous test time in s.000.1-999.9
Charging delay: 000.0 s set charging time, delay range is
000.1-999.9 s, in single or continuous test mode will be based on
charging time delay.
Discharge delay: 000.0 s set discharge time, delay range is
000.1-999.9 s, in single or continuous test mode will be based on
discharge time delay.
Surface resistance: set surface resistivity parameters.
Volume resistance: set volume resistivity parameters.
Display bit: you can set the test interface resistance display bit 4 or
5.
Response mode: set response mode.
4.4 Clear circuit
1. press [0. ADJ] key to enter the ready zero interface. Remove or
suspend the Input end and the voltage output Output end test clip
before starting to clear.
2. Press the ENTER] key to start clearing ,[ ESC] back to the
measurement interface. The instrument clears zero, and in the
automatic range, the instrument clears all ranges. In manual mode,
the instrument only clears the current range. If the zero clearing is
successful, the zero clearing data will be non-volatile memory.
3. The instrument automatically returns to the test state after