36
"range automatic: ON" on the screen. The instrument will
automatically select the most suitable range for measurement
through the following table.
Range number, current range and range variation
Range
Number
Current
range
Alarm
Rise range
Drop range
1
2mA
>2mA alarm
R: range over
2.2mA
220uA
22uA
2.2uA
220nA
22nA
180uA
18uA
1.8uA
180nA
18nA
1.8nA
2
200uA
>200uA no alarm
R: range over
3
20uA
>20uA no alarm
R: range over
4
2uA
>2uA no alarm
R: range over
5
200nA
>200nA no alarm
R: range over
6
20nA
>20nA no alarm
R: range over
7
2nA
>2nA no alarm
R: range over
4.3 Setting test parameters
Click [PAGE] on the test page to switch to the settings page, using
the direction key to select the parameters to be set;