background image

 

 

connecting line. 

5.7  Don't  use  or  have  the  storage  in  the  place  of  sun

high 

temperature

humidity etc.   

5.8 The short circuit time of the testing end should be less than 

60s. 

5.9 The continuous testing time should be less than 3 hours one 

time. 

 

6.Work theory

 

      The  high  voltage  part  of  the  instrument  is  applied  with  the 

Pulse Width Modulate(PWM), via the inner double voltage output 

cathode DC high voltage, with the characteristic of energy saving, 

good voltage linearity, stabilization and small ripple coefficient. 

          The  high  voltage  produced  by  the  instrument  line  terminal 

goes  to  the  instrument  earth  terminal  via  the  load  resistance  Rx, 

then transformation via V/I to display the digit. 

7.Testing 

Danger 

  Before testing, please sure there is no charge existing in the 

tested  circuitry  that  the  high  voltage  instrument  is  connected 

with. 

  Please put on a pair of the high voltage insulation gloves. 

  If  the  battery  cover  is  open,  please  don't  carry  on  any 

measurement. 

7.1 testing end connection 

Insert the plug of the testing line with the testing stick into the L 

terminal socket of the instrument, then insert the plug of the testing 

line with the testing clip to the E terminal socket of the instrument 

and connect well separately. 

7.2 testing connection 

The  connection  of  terminal  E  plug(black  test  lead  with  the  big 

testing clip) is for the Earth, connect the testing clip to one side of 

the  item  ;  the  connection  of  terminal  L  plug(red  test  lead)  is 

circuitry  line,  can  connect  the  testing  clip  to  the  other  side  of  the 

item  ;  the  connection  of  terminal  G  plug  is  shielding  line,  and  is 

connected  to  the  surface  of  the  tested  object  to  avoid  the  surface 

leakiness  which will influence the testing impedance.  Connect the 

testing clip to one end of the tested object, and the small clip of the 

test  lead  to  the  surface  of  the  tested  object  to  avoid  the  current 

leakiness.(Picture A and B) 

 

 

 

 

 

 

 

 

7.3 Rated voltage selection 

Select the rated voltage according to the test requirements: 

HDT20A can choose 250V 500V 1000V 

HDT20B can choose 500V 1000V 2500 V 

7.4 Testing operation 

Please pay attention to safety operation: 

Connect  the  test  lead  of  the  test  stick  to  one  end  of  the  tested 

object, press down the high voltage startup button [TEST], then the 

red indication light HV ON turns on to show that the high voltage 

output  gets  through.  When  the  high  voltage  indicator  is  on,  it 

means  the  instrument  work  normally.  If  it  is  off,  it  shows  maybe 

the battery is low voltage or battery connection is abnormally. 

When the measurement is begun, there is reading on the meter 

header which is the insulation resistance value of the tested object 

under the current voltage.   

For  the  HDT20A,  when  the  test  value  is  less  than  200M

Ω

,suggest pressing the 

                 

 button, using 200M

Ω

  range 

can make the readings more accurate. For the HDT20B, when the 

test  value  is  less  than  2G

Ω

,suggest  pressing  the 

               

 

button,  using  2G

Ω

  range  can  make  the  readings  more  accurate. 

When  the  testing  is  out  of  range  (the  digit  only  display 

1

),  you 

can check the 

                   

or 

               

 button is in the raised 

state or not, if not you can press it to let it in the raised state; if yes 

show  the  measured  value  of  insulation  resistance  measurement  is 

beyond  the  scope  of  this  instrument,  pls  consider  using  a  higher 

resistance of tester. 

If  you  need  more  convenient  readings,  you  can  press  the 

HOLD,  LIGHT

  button,  the  display  digit  will  be  locked  and 

backlight is on for easy reading. Press that button again to restore 

the normal measurement state and the backlight is off. 

7.5 Backlight 

Press      button,  backlight  is  on;  press  the  button  again,  the 

backlight is off.

 

7.6 Power off 

  After  finish  the  measurement,  press  down  the  high  voltage 

button [TEST], HV ON red indicator turns off slowly, showing the 

high voltage output has disconnected. Set the switch to the position 

[OFF], there is no display on the LCD. For the capacitive load, you 

should completely release the residual charge in the tested object to 

avoid the residual charge release electricity to injury people.   

8. AC Voltage Test 

    8.1 Using the rotary switch select the 

ACV

 position 

    8.2 Insert the red test lead into the 

ACV

 jack, insert the black 

test lead into the 

G

 input jack 

    8.3 Contact the two test leads with the both ends of the voltage 

to  be  measured,  then  the  digital  display  show  the  AC  voltage  in 

volt. 

    Note: 

1.

 

The  measured  voltage  can

t  exceed  750V  AC  RMS,  the 

frequency  of  the  ACV  is  40Hz  to  400Hz.  If  exceed  this 

frequency  range,  the  data  errors  will  exceed  the  technical 

specifications of this    instrument. 

2.

 

For the earth voltage to be measured at the both ends can

exceed 750V AC RMS, otherwise will meet the danger of 

electric shock. 

3.

 

AC  voltage  test,  don

t  need  to  transfer  the  high  voltage 

and  range,  so  at  that  time  the 

TEST

  and 

                                       

 button are disabled. 

9.Battery replacement 
   

If there is the symbol "      " on the LCD, it indicates low battery, 

please replace the battery.

 

Before replacing the battery, please sure the testing connection 

has been disconnected and power is cut off. 

Содержание HDT20A

Страница 1: ... sure the instrument is turned off when opening the battery cover Notice Before test please sure the voltage selector switch is on the proper location After test please set the function switch to the range OFF If it is available for a long time please take out the battery for storage Please don t put the instrument in the place of high temperature humidity dew and strong light for a long time Plea...

Страница 2: ... 0 01GΩ 19 99GΩ 10 1000V 0 01GΩ 19 99GΩ 2500V 0 01GΩ 19 99GΩ ACV 750V 0 1 6 AC750V input impedance 1MΩ AC750V frequency response 50 200 Hz 4 Appearance 1 Line terminal L 2 Shield terminal G the end terminal of ACV measurement 3 The positive terminal of the ACV measurement 4 Earth terminal E 5 Backlight button 6 Range switch button 7 High volt startup button 8 External power jack 9 Instrument belt ...

Страница 3: ...the instrument work normally If it is off it shows maybe the battery is low voltage or battery connection is abnormally When the measurement is begun there is reading on the meter header which is the insulation resistance value of the tested object under the current voltage For the HDT20A when the test value is less than 200M Ω suggest pressing the button using 200MΩ range can make the readings mo...

Страница 4: ... Storage 10 1 Set the switch to the OFF position 10 2 When the instrument is available for a long time please take off the battery to avoid the electrolyte discharge resulting in rust 10 3 Please keep the case clean 10 4 instrument storage Storage temperature 10 50 Storage relative humidity 90 RH The place need no dust no causticity gas and well ventilated 11 Accessory HDT20A one plastic cover a p...

Отзывы: