Section 6 - Troubleshooting
6-15
Event
Number
Event List
Desc. (Short)
Event History
Description
Data Field
Problem
Possible Cause
Recovery
233
FAIL OPT
VERIFY
Optics
Verification
Failed
Adj Cnts
Optics Verification
Failed
Chemcassette leader
not tight or improperly
positioned during
light gray to dark gray
calibration
Reload Chemcassette
Bad RFID tag
Load new Chemcassette
Dirty optics block
Clean and recalibrate
Bad optics PCB set
Service or replace
analyzer
234
GRAY FAILED
Light Gray Read
Failed
Optics Cnts
Optics values of this
field were outside limits
Chemcassette leader
not tight or improperly
positioned during
light gray to dark gray
calibration
Reload Chemcassette
Bad RFID tag
Load new Chemcassette
Dirty optics block
Clean and recalibrate
Bad optics PCB set
Service or replace
analyzer
235
AZ SW VER
WRONG
Analyzer
software version
mismatch
0
Software mismatch
Reload program to all
analyzers
Contact Honeywell
service
237
AZ VACUUM
LOW
Inadequate
Analyzer Vacuum
Observed
vacuum in in
Hg
Individual analyzer
observes inadequate
vacuum, stops
monitoring
Fault 219 has triggered.
Recover from Fault 219,
then reset Fault 237
Circular tubing harness
disconnected
Connect circular tubing
harness
Tubing pinched
Correct tubing dress in
cable carrier
Contact Honeywell
Service
238
OP SW
MISMATCH
Optics block
software
mismatch.
0
The two optics blocks
are not running the same
software version.
Program update done
incorrectly.
Reload program to both
optics blocks.
239
OPTICS COMM
FAIL
Optics block
communication
failure
Error code
Analyzer is not able to
communicate with the
optics block.
Cable disconnected.
Software anomaly.
Defective optics block.
Check the 14 pin
cable connection on
the sensor board.
Cycle power to analyzer.
Contact Honeywell service.
240
OPTICS REF
FAIL
Optics reference
limit exceeded
Reference
value signal
The signal detected at
the reference photodiode
is out of range.
Degraded optics block.
Contact Honeywell service.
Содержание Vertex M
Страница 8: ...1 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook 1 Introduction ...
Страница 25: ...Vertex MTM 24 Point Continuous Monitor Section 1 Introduction 1 18 ...
Страница 26: ...2 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook 2 Installation ...
Страница 35: ...Vertex MTM 24 Point Continuous Monitor Section 2 Installation 2 10 ...
Страница 36: ...3 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook 3 Startup ...
Страница 40: ...Vertex MTM 24 Point Continuous Monitor Section 3 Startup 3 5 Figure 3 1 Vertex M Main Screen ...
Страница 77: ...4 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook 4 Operation ...
Страница 93: ...Vertex MTM 24 Point Continuous Monitor Section 4 Operation 4 17 Figure 4 9 Filter Options ...
Страница 109: ...Vertex MTM 24 Point Continuous Monitor Section 4 Operation 4 33 ...
Страница 125: ...5 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook 5 Maintenance ...
Страница 134: ...Vertex MTM 24 Point Continuous Monitor Section 5 Maintenance 5 10 ...
Страница 140: ...Vertex MTM 24 Point Continuous Monitor Section 5 Maintenance 5 16 ...
Страница 141: ...6 1 Vertex M TM Technical Handbook 6 Troubleshooting ...
Страница 164: ...Section 6 Troubleshooting 6 24 ...
Страница 165: ...A 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook A Installation Drawings ...
Страница 166: ...Vertex MTM 24 Point Continuous Monitor Appendix A Installation Drawings A 2 A 1 Introduction ...
Страница 167: ...Vertex MTM 24 Point Continuous Monitor Appendix A Installation Drawings A 3 A 2 Floor Space Requirements ...
Страница 168: ...Vertex MTM 24 Point Continuous Monitor Appendix A Installation Drawings A 4 A 3 General Considerations and Cooling ...
Страница 169: ...Vertex MTM 24 Point Continuous Monitor Appendix A Installation Drawings A 5 A 4 General Wiring ...
Страница 170: ...Vertex MTM 24 Point Continuous Monitor Appendix A Installation Drawings A 6 ...
Страница 171: ...Vertex MTM 24 Point Continuous Monitor Appendix A Installation Drawings A 7 A 5 4 Port Manifold Option ...
Страница 172: ...Vertex MTM 24 Point Continuous Monitor Appendix A Installation Drawings A 8 A 6 Floor Mounting Option ...
Страница 173: ...Vertex MTM 24 Point Continuous Monitor Appendix A Installation Drawings A 9 ...
Страница 174: ...B 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook A Specifications ...
Страница 178: ...C 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook A Detectable Gases ...
Страница 190: ...D 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook A Replacement and Consumable Items ...
Страница 193: ...Vertex MTM 24 Point Continuous Monitor Appendix A Replacement and Consumable Items D 4 ...
Страница 194: ...E 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook A Optional Relay Specifications ...
Страница 198: ...F 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook A Network Interface and Options ...
Страница 226: ...Vertex MTM 24 Point Continuous Monitor Appendix A Network Interface and Options F 29 Figure A 18 Custom Permissions ...
Страница 237: ...G 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook A 4 20mA Analog Output Option ...
Страница 240: ...Vertex MTM 24 Point Continuous Monitor Appendix A 4 20mA Analog Output Option G 4 ...
Страница 241: ...H 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook A Line Integrity Test Option ...
Страница 247: ...I 1 Vertex MTM 24 Point Continuous Monitor Vertex M TM Technical Handbook A Warranty Statement ...
Страница 250: ...Vertex MTM 24 Point Continuous Monitor Appendix A Warranty Statement I 4 ...