
Table 1. Programmable Parameters
Parameter
Display
Default
Min
Max
Fahrenheit or Centigrade
F
-
C
F
F
C
Remote Probe Enable
rEN
NO
NO
YES
Display Local or Remote
temperature
DSP
LCL
LCL
RPR
Set Local Sensor Low Limit
LLO
40
32
I40
Set Local Sensor High Limit
LHI
I40
32
I40
Set Remote Probe Low Limit
RLO
40
-
40
I40
Set Remote Probe High Limit
RHI
I40
-
40
I40
Select Output Mode (both
relays)
1
Out
NC
NC
NO
Set Output Map for Output 1
2
OuI
I
I
I
Set Output Map for Output 2
2
Ou2
2
2
2
Set Alarm Delay for Output 1
DLI
0
0
255
Set Alarm Delay for Output 2
dL2
0
0
255
Enable Alarm Audible
Annunciator
AUD
YES
NO
YES
Audible Alarm Silence
Timeout
SIL
0
0
255
Hysteresis
3
Degrees of
HYS
0
0
I0
NOTES:
1
If programmed for Normally Closed (NC) it is supervised, if Normally Open (NO) it is unsupervised.
2
The Output map is a graphical representation of the possible choices.
The left character indicates which relay you are mapping, 1 or 2.
The middle character is the local sensor map and the right character is the remote probe map. See Table 2 for
map designations.
3
Hysteresis provides a buffer so that temperature that is shifting back and forth around the alarm limit point will
not create multiple alarms. You set the # of degrees that the temperature will have to rise
above
the
temperature low limit set point or drop
below
the high limit set point for the alarm to turn off. For example, if
you set the HYS at 3 degrees and your low limit is 40 degrees, you will go into an alarm condition when the
temperature reaches 39 degrees and restore from alarm would not occur until the temperature rose to 43
degrees. Hysteresis is limited to ½ the difference in degrees between the low temperature limit and high
temperature limit
or
the programmed HYS setting (max of 10
E
) whichever is least.
4
The setup will
NOT
be lost in the event that power is lost at the panel.
5
Options that relate to the remote probe will only appear is you have enabled the remote probe in selection 2
of the table parameters.
6
High/Low limits shown in table are degrees Fahrenheit. Equivalent Centigrade temperatures apply when
Centigrade has been selected.
7
Hysteresis is always programmed in degrees Fahrenheit, even if Centigrade has been selected to set limits.
-2-