POTENTIAL OPERATING ISSUES
Touchpoint Pro
Pt. No. 2400M2566_1_EN
71
Operating Manual
Error
Code
Message
Possible Cause
Remedial Action
Sys
Fail
Relay
Sys
Fault
Relay
91
Operation failed; system is
running in test mode.
One of the three test
modes is active, and
other operations are
restricted or not
permitted
Wait until testing is complete. If necessary, exit
Test Mode to carry out the required operation.
92
Calculation error
A calculation error
has occurred during
calibration, for
example Zero and
SPAN calibration
done with the same
gas input
Retry the calibration. Ensure that the values for
zero and span gas are entered correctly.
93
Operation failed; module list is
mismatching
The Module list
detected and the
Module list in Backup
file are not the same
Ensure that the correct backup file for the
system is being used. Retry the operation.
94
Operation failed; please perform
ring initialisation first
Faulty CCB has been
replaced by new CCB
and configuration
restore has been
attempted
First initialise the ring network, then try again
95
The username/password is
incorrect, please try again
User credential
mismatch
Verify user name and password and try again.
Usernames and passwords are case sensitive.
96
Not used
—
—
97
Cannot open session because a
channel is being calibrated
A user at the local
Touchscreen is doing
calibration when a PC
Configuration
Software user tries to
open a configuration
session
Wait for the calibration to be completed.
Check with the user at the local controller that it
is acceptable to log in from remote PC.
98
Cannot start calibration because
a configuration session on PC is
in progress
Calibration is
attempted while a
configuration session
is in progress
Close the configuration session and retry
calibration.
Contact the remote user to agree access.
99
Operation failed; Brownout Fault Brown out fault (low
power supply voltage)
is set for a module
and configuration is
attempted for this
module / channels of
module
Check that power supply voltage is correct, and
that current limit is adequate.
Remedy the fault and then do configuration.
X
100
Replace operation cannot be
performed - no free I/O Modules
available. Please add a
compatible I/O Module and
initialize the ring network again.
The replacement
module has not been
detected on the
network
Refer to section Maintenance, Replacing an I/O
Module for the full procedure.
101
Ring initialisation has failed.
Please cycle the power to the
system
Ring Network
Initialisation has failed
1. Power cycle the TPPR system
2. Return to the System Setup screen and retry
the Ring Initialisation operation.
102
Initialisation timeout has
occurred. Please cycle the
power to the TPPR system.
An internal system
timeout has occurred
Power cycle the TPPR system