Symptom
Cause
Corrective Action
Chemcassette
tape will not ad-
vance
No power
Check the cable connection
Chemcassette
tape continues to
advance
Encoder
Check the cable connection
Hall effect sensor
Check the cable connection
Chemcassette gate will
not open
No power
Check the cable connection
Cannot remove Chem-
cassette cartridge
Gate open command not sent
Use either Change Chemcassette wizard or Maintenance / Open/Close gate.
Chemcasette cartridge
door will not latch closed
Door is not properly closed or
aligned
Make sure Chemcassette cartridge is fully seated.
Make sure USB stick is not interfering with door.
Make sure door is fully seated on both sides – latches should hook under the tabs on the side of the case easily.
Make sure all four latches are properly secured.
Pump will not turn on
No power
Check the cable connection
Cannot achieve target
flow rate
Inlet or outlet tube is blocked
Confirm inlet and outlet tube are free of restrictions or kinks, and that all filters are clean.
Gate is not sealing correctly
Confirm Chemcassette cartridge is fully seated.
Exit and enter monitor to pull a fresh part of the tape.
Pump is old
Replace pump. Contact HA.
Filters are clogged
Check filters (internal and external) and replace as needed
Tubing connections are not
properly seated
Confirm all tubing connections are properly seated (internal and external).
Detector will not connect
properly or stay on
mounting plate
Incorrectly installed mounting
plate
Make sure the mounting plate is installed the correct direction so that the mounting heads on the back of the SPM
Flex can slot in and slide down to secure.
Mounting screws not installed Confirm the mounting screws have been installed on the back of the SPM Flex
Gas readings are not as
expected
Flow system issue
Confirm all tubing connections are properly inserted.
Confirm correct filter type is used for the target gas. (Refer to
.)
Confirm target flow rate is achieved.
Chemcassette cartridge issue
Confirm correct Chemcassette cartridge is selected for the target gas.
Confirm Chemcassette cartridge is within operating age, sampling conditions, and has been correctly stored.
If generating a gas bump test gas to confirm performance, refer to gas generation document located in the online
High Tech Technical Library.
Refer to individual Chemcassette type technical notes for specific performance information.
Buzzer does not activate No power
Check the cable connection
Detector won’t turn off
Power button on top left / menu
option not available
Verify that the user is logged in with the correct security level
Detecttor unintentionally
turns on during transport
Physical power switch is on,
unintentionally pushing power
switch by display
Put the physical power switch located behind the Chemcassette door in the off position. Place switch back in the on
position when ready to use again.
Detector will not charge Faulty connection
Check the cable connection
*corrective actions vary with security levels
Содержание SPM Flex
Страница 1: ...SPM Flex Single Point Monitor Gas Detector User Manual ...
Страница 2: ......
Страница 27: ...27 SPM Flex Gas Detector Setup Installation drawing Figure 18 Installation drawing ...
Страница 70: ...70 SPM Flex Gas Detector ...
Страница 73: ......
Страница 74: ......
Страница 75: ......
Страница 76: ...SPM Flex Gas Detector User Manual 1998M0845 Revision 07 October 2016 2016 Honeywell Analytics ...
Страница 77: ......