RDD100 User Guide 

 

 

 

www.honeywell.com                                                                                                                                                             3

 

INTRODUCTION 

 
1.1 

Product Overview 

 
The Honeywell Remote Digital Display, RDD100, produces a digital display of measured pressures and is 
designed for use with RS-232 versions of Honeywell HPA, HPB, PPT, PPTE, and PPTR pressure transducers.  
The 6-digit display features 0.56 inch-tall LED characters that are readable in bright sunlight.  The RDD100 is a 
digital-input instrument, accepting a continuous stream of digital data transmitted in RS-232 format from the 
transducer.  The transducer may be installed at the source of the pressure being measured; a cable up to 50 feet 
in length may separate the transducer from the RDD100.  A wall-mounted AC-to-DC power supply, which powers 
both the RDD100 and the transducer, is supplied with the display.  Alternatively, a user may provide 12V ± 3V DC 
from a system power supply. 

 

The RDD100 has a 9-

pin “D” connector mounted on the rear panel to connect to a personal computer.  The 

computer can perform data collection while the RDD100 is continuously displaying readings.  The computer can 
also be used to send commands to the transducer    (two-way communications). 

 

The RDD100 is housed in a compact enclosure, measuring approximately 3.8 x 2 x 4.4 inches, W x H x D.  The 
instrument may be used on a tabletop or can be easily installed in a panel using the supplied mounting hardware. 

 

1.2 

Pressure Transducer Description 

 
The RDD100 receives pressure readings from an RS-232, Honeywell pressure transducer (HPA, HPB, PPT, 
PPTE and PPTR).   Accuracy of the displayed pressure reading is determined solely by the accuracy of the 
transducer.  Typical PPT accuracy, for example, is ±0.05% of full scale (FS).  Digital output resolutions up to 
0.0011% FS are possible but are dependent upon the pressure display units selected.  Although many of the 
supported transducers also have a 0-5V analog output, only the digital output is used by the RDD100 as it 
provides better resolution and accuracy.  Any of 16 different pressure display units may be selected for use with 
the RDD100.  A list of available pressure units is shown in section 2.4 below.  A more complete description of the 
Honeywell transducers can be found in their respective user manuals.  See http://www.pressuresensing.com. 

 

EQUIPMENT SET UP 

 
2.1 

Cables & Connections 

 
Figure 1 shows connections that must be made between the RDD100 and Honeywell RS-232 pressure 
transducer.  A six foot long cable for this connection is supplied with the RDD100.  Figure 1 also shows 
connections to the 12V DC input power.  To obtain transducer pressure readings, only these two connections are 
needed.  Figure 1 also shows a connection to a personal computer.  This connection may be used for data 
logging or to communicate with the transducer in order to set up and change various parameters (display units, 
integration time, idle count, etc).  A six or ten foot long computer-to-RDD100 cable can be selected when the 
RDD100 is ordered. 
 

2.2 

Personal Computer Set Up

 

 
The RS-232 transducer must be configured to send continuous readings to the RDD100 in order for pressure 
readings to appear.  The display will remain blank and the error LED will be lit continuously when the RDD100 
fails to receive continuous input from a transducer. 
 
Transducers ordered with the RDD100 will be configured at the factory to function properly with the RDD100.  A 
transducer that has not been configured for continuous readings can be configured for use with the RDD100 by 
connecting it to the COM port of a PC as shown in Figure 1.  The HPA/B, PPT, PPTR/E Demo program or a 
terminal emulator program such as 

HyperTerminal,

 may be used to send RS-232 formatted serial data to the 

COM port. 
 
 
 
 

Содержание RDD100

Страница 1: ...Honeywell Pressure Sensor Products RDD100 Remote Digital Display User Manual ...

Страница 2: ...ed Parameters 3 Operation 3 1 Start Up Messages 3 2 Error Indications 3 2 1 Transducer Not Connected 3 2 2 Incorrect Baud Rate 3 2 3 Not Set Up To Continuously Transmit 3 2 4 Wrong Polarity Of DC Power 3 2 5 Transducer Disconnected While Transmitting 3 2 6 Over Under Ranged Transducer Indications 4 Power Requirements 4 1 Voltage 4 2 Current 4 3 Power Connection 5 Operating Environment 5 1 Operatin...

Страница 3: ...dependent upon the pressure display units selected Although many of the supported transducers also have a 0 5V analog output only the digital output is used by the RDD100 as it provides better resolution and accuracy Any of 16 different pressure display units may be selected for use with the RDD100 A list of available pressure units is shown in section 2 4 below A more complete description of the ...

Страница 4: ...te for the transducer is shown below The commands below all use a broadcast address 99 to allow them to operate correctly even if the transducer address has been changed from the factory default setting which is address 01 With the PC and transducer operating at the baud rate found to return a non garbled start up message Type 99we cr Carriage Return Resp 99WE Command Echo Type 99bp n9600 cr Resp ...

Страница 5: ...was in the original psi scale For example when a 10 psia device is switched to DU LCOM the full scale digital pressure output will be 60 000 instead of 10 000 Also a 500 psi device will output full scale 600 00 instead of 500 00 PFS generally provides the highest readout resolution available when using the RDD100 readout 2 5 Continuous Transmit Mode Type 99we cr Resp 99WE Command Echo Type 99mo p2...

Страница 6: ... Power The RDD100 is equipped with the reverse polarity protection diodes at its power input The instrument will operate with either polarity 3 2 5 Transducer Disconnected While Transmitting If the transducer is disconnected while transmitting the last reading received will remain on the display and the error LED will light indicating that no further updates are being made Reestablishing connectio...

Страница 7: ...ed 2 Separate cables are available for transducers with Metal or Plastic electrical connectors Standard cable length is 6 feet 1 8 m Connectors may be purchased from Honeywell to construct custom length cables Due to RS 232 limitations recommended maximum length is 50 ft 15 m 3 12 VDC plug is 2 1 x 5 5 x 10 mm I D x O D x length Center pin may be either positive or negative polarity RDD100 Wall mo...

Страница 8: ...RDD100 User Guide 8 www honeywell com Honeywell International 12001 State Highway 55 Plymouth MN 55441 Tel 1 800 323 8295 www honeywellmps com ADS 14124 December 2012 ...

Отзывы: