HERMetic Sampler A2
6.4 Operating the SPOT SAMPLING BOTTLE
(O = OPTION)
ND
TS
DESCRIPTION
O
20255
20137
Spot bottle 0.40 l. FKM
1.
Remove the 2” cover from the valve.
2.
Install the sampler on top of the valve.
3. Open the valve.
4.
Lower the spot bottle to the level where the sample is to be taken.
5.
Stop the bottle at this level and shake it rapidly up and down about 10 times on a 200 mm
stroke.
This movement has a pumping effect as the ball opens and closes the bottom of bottle.
6.
Lift the bottle back into the sampler.
7.
Close the valve.
8.
Remove the cover of the sampler together with the spot bottle. Mind not to open the bottom
valve inadvertently.
9.
Put the spot bottle vertically into a laboratory bottle with a minimu
m 2” neck diameter.
10. When the bottle spot cover hits the bottom of the laboratory bottle the liquid is transferred.
11. Reassemble the spot bottle and the cover to the sampler.
12. Remove the sampler from the valve.
13.
Reinstall the 2” cover on top of the valve.
14. Clean the equipment after use and check it for proper functioning.