HA71 Digital Gas Controller

Technical Manual

TO

• 

C

General Descriptio

• 

n

Basic Operatio

• 

n

Main I/O Interfac

• 

e PCB

System Diagnostic

• 

s

Modbus

• 

®

 RS485 Ports

HA71 Enclosure

• 

s

Ordering Informatio

• 

n

Warrant

• 

y

Содержание HA71

Страница 1: ...HA71 Digital Gas Controller Technical Manual TO C General Descriptio n Basic Operatio n Main I O Interfac e PCB System Diagnostic s Modbus RS485 Ports HA71 Enclosure s Ordering Informatio n Warrant y ...

Страница 2: ...r C22 2 No 152 performance requirements This equipment is suitable for use in Class I Division 2 Groups A B C and D or non hazardous locations only Use a properly rated CERTIFIED AC power mains cable installed as per local or national codes A certified AC power mains disconnect or circuit breaker should be mounted near the controller and installed following applicable local and national codes If a...

Страница 3: ...a From Menu to Set Input Source 2 4 2 2 5 MIN MAX Raw Counts Menu 2 5 2 2 6 Marker Menus 2 6 2 2 7 Linearization Menu 2 6 2 2 8 Configure Menu 2 6 2 2 9 Eunits Measurement Name ASCII Data Fields 2 6 2 2 10 Input Measurement Range 2 7 2 2 11 Decimal Point Resolution 2 7 2 2 12 Turning off Unused Channels 2 7 2 2 13 Copy Data To 2 7 2 2 14 CAL Mode 2 7 2 3 System Configuration Menus 2 8 2 3 1 Common...

Страница 4: ... 5 0 Modbus RS 485 Ports 5 2 5 1 Modbus Slave Register Locations 5 2 6 HA71 Enclosures 6 0 HA71 Enclosures 6 2 6 1 HA71PM Panel Rack Mount Enclosure 6 2 6 2 HA71N4 NEMA 4X Fiberglass Wall Mount Encl 6 3 6 3HA71SSNEMA4X316StainlessSteelWallMountEncl 6 5 6 4 HA71XP NEMA 7 Explosion ProofWall Mount Encl 6 6 6 5 HA71 Main I O and Option PCB Footprint Dimensions 6 7 A Ordering Information A Ordering In...

Страница 5: ...HA71 Digital Gas Controller Revision 3 04 08 HA71 Technical Manual 1 1 1 General Description ...

Страница 6: ... Figure 1 2 displays a 24 hour trend of input data for the channel selected Horizontal tic marks are each hour and vertical tic marks are each 10 of full scale Dashed lines indicate alarm levels The graphic LCD is 240 pixels wide so each pixel represents 1 10 hour or 6 minutes worth of data The trend is 100 pixels high so each represents 1 of full scale in amplitude Since each data point must be c...

Страница 7: ...the AVERAGE buffer overflows and the error message UPDATE appears in the AVERAGE field Exiting this screen resets the buffer and clears the error message 1 2 Specifications 1 2 1 DC Power Supply Requirements Standard HA71 power requirements are 10 30VDC 3 watts applied to terminals 9 11 of TB2 on the standard I O PCB see Section 3 0 Optional features increase power consumption as described below D...

Страница 8: ...e 6 1 NEMA 4X wall mount in fiberglass enclosure weighing 17 lbs Figure 6 2 NEMA 4X wall mount enclosure in 316 stainless steel weighing 30 lbs Figure 6 4 NEMA 7 wall mount suitable for DIV 1 2 Groups B C D weighing 110 lbs Figure 6 5 Includes non intrusive magnetic keypad 1 2 8 Non Intrusive Magnetic Keypad The HA71 operator interface includes five front panel touch keys A magnetic keypad option ...

Страница 9: ...HA71 Digital Gas Controller Revision 3 04 08 HA71 Technical Manual 2 1 2 Basic Operation ...

Страница 10: ...asurement Name 50 PCTLEL S y s t e m C o n t r a s t A u t h o r i z e C h a n n e l 0 C h a n n e l 0 2 C h a n n e l 0 3 C h a n n e l 0 4 C h a n n e l 0 5 C h a n n e l 0 6 C h a n n e l 0 7 C h a n n e l 0 8 C h a n n e l 0 9 C h a n n e l 1 0 C h a n n e l 1 1 C h a n n e l 1 2 C h a n n e l 1 3 C h a n n e l 1 4 C h a n n e l 1 5 C h a n n e l 1 6 U N L O C K E D S E T U P M E N U 1 D i a g...

Страница 11: ... 2 1 Channel Setup Entry Menu The entry menu shown on the left side of Figure 2 2 allows access to all configuration variables for the selected channel These are Alarm 1 Alarm 2 Alarm 3 Data From Linearize Input and Calibrate 2 2 2 Alarm 1 Alarm 2 Horn Relay Set Up Menu Alarms 1 and 2 are identical except A1 may not be acknowledged and front panel LED indicators are yellow while A2 s are red Since...

Страница 12: ...channel alarms LED s optional discrete relays may be set as Alarm 3 level alarms the common relay for this group is always a Fault alarm The fault out of range threshold for the channel is the most recent Fault trip point entered prior to changing the menu to Alarm 3 The following example describes how to configure both the Fault out of range and Alarm 3 level trip points for a channel Example If ...

Страница 13: ...ad out values described in Section 2 2 10 This menu entry is determined by the A D converter resolution of the channel s input For example if the input is a 10 bit Modbus device with zero at 200 counts and 100 at 1000 counts then this menu s MIN should be set at 200 and MAX at 1000 If communicating with the HA71 s optional 12 bit Analog Input PCB the MIN should be 800 and the MAX 4000 If the input...

Страница 14: ... is 0 200 ppm H2 S then 100 ppm is 50 of full scale Zero input will provide a zero output and 100 input a 100 output Nine intermediate points may be entered to define the curve EDIT NEXT Entry menu Channel XX Measurement Name Linearization menu Channel XX Linearization Input Output 10 00 20 00 30 00 40 00 50 00 60 00 70 00 80 00 90 00 10 00 20 00 30 00 40 00 50 00 60 00 70 00 80 00 90 00 Alarm 1 A...

Страница 15: ...ocess inputs applied to this channel and no alarms will be tripped or data displayed Inactive channels have a line drawn through them on the Setup screen as indicated by channels 15 16 in Figure 2 1 If less than 9 channels are to be activated the HA71 may be set for 8 channel mode deactivating channels 9 16 This is done in the System Setup menu described in Section 2 3 The HA71 will only allow 15 ...

Страница 16: ...ombustible sensor is to be spanned with 50 LEL span gas the span set point must be 50 If 45 LEL is to be used later the span set point must be changed to 45 to match the span calibration gas If the reading is only 40 LEL with the 50 gas applied a span calibration is needed Move the pointer to the Set Span entry and press EDIT twice Unity Gain may be used at anytime to cancel incorrect calibrations...

Страница 17: ...gize during alarm conditions and energize when there is no alarm Therefore a power failure forces the relay contact to the alarm position Note the common Fault relay is always failsafe and may be monitored separately to indicate loss of power conditions in many applications A1 and A2 Votes allows creation of logical AND function equations that control common relay 1 common relay 2 Default settings...

Страница 18: ...indicate alarm conditions The common horn relay is similar to the common A1 and A2 common relays Figure 2 13 Common Horn Relay and Local Piezo Menu Turning Piezo Alarm ON causes the audible piezo to duplicate the action of the horn relay This feature may be used to provide a low decibel indication of the status of the system s horn Alarm 1 Alarm 2 menus control how this alarm level from each chann...

Страница 19: ...astrTO master time out and Mastr PR master poll rate menu items affect the HA71 s master Modbus port Time out sets length of time in milliseconds before a communications error Three consecutive timeout errors must occur before a communication error is indicated This item is useful for optimizing throughput to the HA71 from other slave RTU s Poll Rate sets frequency of data requests to the RTU s in...

Страница 20: ...e system the correct 4 digit authorization number must be entered into the Enter Code field Point to the Unlock System entry and press EDIT to complete the unlock procedure It is very important to remember the 4 digit code since the factory must be consulted if it is lost Authorization ENTER NAME Enter Code Lock System Figure 2 16 Authorization Mode Menu 2 5 LCD Contrast Adjustment The Setup menu ...

Страница 21: ...HA71 Digital Gas Controller Revision 3 04 08 HA71 Technical Manual 3 1 3 Main I O Interface PCB ...

Страница 22: ...dditional terminals for incoming and outgoing cables For example installing the 2 COM 1 jumpers connects screw terminals 1 and 5 and terminals 3 and 7 Socketed RS 485 terminating resistors R6 COMM 1 and R12 COMM 2 are located on the MAIN I O board These resistors should be removed if communication wire lengths are very short less than 25 feet or if the port is not a the end of the communication li...

Страница 23: ...d IN LO terminals are tied to TB2 2 Bussing allows transmitter power to be brought into the system at a single point TB2 and distributed back out at each channel s EXC IN LO terminals to simplify field wiring Figure 3 3 includes typical wiring to 2 and 3 wire 4 20mA transmitters JP1 determines if the 8 analog inputs are applied to channels 1 8 or channels 9 16 Connecting more than 8 analog inputs ...

Страница 24: ...cient when these boards are in close proximity to each other Caution All relays are rated at 5 Amp for 28 VDC and 250 VAC RESISTIVE loads Appropriatediode DCloads orMOV ACloads snubberdevices must be installed with inductive loads to prevent RFI noise spikes Relay wiring should be kept separate from low level signal wiring TB2 5VDC ONLY ASSY 10 0195 8 CH DISCRETE RELAY TB1 JUMPERS ZONE JP4 U2 J2 J...

Страница 25: ...Setup Catalytic bead sensors vary widely in power requirements and sensitivity It is therefore important to configure each channel to match the sensor with which it will operate Prior to connecting sensors apply power to the system 1 Note This PCB requires 24VDC power be connected to its TB2 terminals 1 and 2 as shown in Figure 3 5 Suitable fused power is available from the Main I O board s TB2 te...

Страница 26: ...LEL 4 20mA jumper to 4 20mA U1 1 3 5 7 9 11 13 15 17 19 21 23 2 4 6 8 10 12 14 16 18 20 22 24 Ch8 16 C A R Ch6 14 C A R Ch4 12 C A R Ch2 10 C A R 1 2 100 ohm Res Sockets JP9 sets for Ch s 1 8 or 9 16 Ch 1 Ch 2 Ch 4 Ch 6 Ch 5 Ch 3 Ch 7 Ch 8 Up to 4 10 0192 Dual CATBEAD input modules may be installed Disconnect power before removing or installing modules CH B CH B CH A CH A 4 2VDC T P LOW HIGH MED G...

Страница 27: ...of RTD sensor readings Each channel must be calibrated individually by either simulating desired zero and span calibration resistance values or by actually placing the channel s RTD into an actual precision temperature generator Ice water is an acceptable method for generating the 0 C zero temperature value Upscale span values are best simulated with an RTD calibrator Since RTD s are stable and re...

Страница 28: ...see Figure 3 1 Since the PCB has 8 channels two are required for 16 channel applications JP1 configures the outputs for channels groups 1 8 or 9 16 TB1 Q4 Q6 Q2 Q10 Q12 Q8 D2 Q16 Q14 HI HI HI HI HI HI HI HI U11 ST 71 mA OUTPUT BOARD Q1 LO LO LO LO LO LO LO LO Q5 Q3 Q7 PCB 0010 1121 Assy 10 0167 C3 High s Milliamp Low s Outputs Rev 0 CR1 C6 R26 C7 U9 Q13 Q11 Q9 C4 C1 J2 C2 J1 D1 U8 Q15 U10 TP1 Q17 ...

Страница 29: ...ta bits no parity one stop bit and FlowControl Hardware Printer diagnostic red LED s indicate printer faults such as out of paper overflowed buffer or loss of communications Green LED s flicker to confirm good communications between the HA71 and printer during print attempts Examples of printed alarm events are shown below The format of each event from left to right is DATE TIME 16 character ASCII...

Страница 30: ...s immediate printing of all channel variables such as channel ID s Engineering Units etc PRINTER READY ERROR indicates the functional status of the printer Figure 3 9 Clock Printer Setup Menu 3 1 9 Optional 24VDC 150 Watt Power Supplies The HA71 Controller may be powered from 10 30VDC However many applications require 24VDC power for the monitors or transmitters providing inputs to the HA71 A 150 ...

Страница 31: ...HA71 Digital Gas Controller Revision 3 04 08 HA71 Technical Manual 4 1 4 System Diagnostics ...

Страница 32: ...analog I O It is exited manually by pressing NEXT and automatically if no keys are pressed for 5 minutes Caution It is very important to understand that CHANNEL INPUT DATA IS NOT PROCESSED DURING THE DIAGNOSTICS MODE It is possible to miss important input values while utilizing this mode and appropriate safeguards should be in place However the Diagnostics Mode can prove invaluable when testing I ...

Страница 33: ...HA71 Digital Gas Controller Revision 3 04 08 HA71 Technical Manual 4 3 Figure 4 2 System Diagnostic Mode Part 2 ...

Страница 34: ...HA71 Digital Gas Controller Revision 3 04 08 HA71 Technical Manual 5 1 5 Modbus RS 485 Ports ...

Страница 35: ... Last Read FC Write FC Notes Channel Tag 1 40401 40408 3 n a 2 Characters per Register Channel Tag 2 40409 40416 3 n a 2 Characters per Register Channel Tag 3 40417 40424 3 n a 2 Characters per Register Channel Tag 4 40425 40432 3 n a 2 Characters per Register Channel Tag 5 40433 40440 3 n a 2 Characters per Register Channel Tag 6 40441 40448 3 n a 2 Characters per Register Channel Tag 7 40449 404...

Страница 36: ...rs per Channel EUNITS 7 40547 40549 3 n a 2 Characters per Register 3 Registers per Channel EUNITS 8 40550 40552 3 n a 2 Characters per Register 3 Registers per Channel EUNITS 9 40553 40555 3 n a 2 Characters per Register 3 Registers per Channel EUNITS 10 40556 40558 3 n a 2 Characters per Register 3 Registers per Channel EUNITS 11 40559 40561 3 n a 2 Characters per Register 3 Registers per Channe...

Страница 37: ...egisters per Channel ASCII Value 10 40604 40606 3 n a 2 Characters per Register 3 Registers per Channel ASCII Value 11 40607 40609 3 n a 2 Characters per Register 3 Registers per Channel ASCII Value 12 40610 40612 3 n a 2 Characters per Register 3 Registers per Channel ASCII Value 13 40613 40615 3 n a 2 Characters per Register 3 Registers per Channel ASCII Value 14 40616 40618 3 n a 2 Characters p...

Страница 38: ...t integer see bit by bit definition below Bit by Bit Definition Alarm 1 Trip bit0 1 Low 0 High Alarm 1 Horn Drive bit1 1 On 0 Off Alarm 3 Type bit2 1 Level 0 Fault Alarm 2 Horn Drive bit3 1 On 0 Off Linearize bit4 1 On 0 Off Alarm 3 Trip bit5 1 Low 0 High Input In Cal bit6 1 1 5mA Cal Detected 0 Normal Mode Channel Disable bit7 1 Disabled 0 Enabled Controller Channel In Cal bit8 1 Cal Mode 0 Norma...

Страница 39: ...mon Relay status bits register 31036 are as follows Relay 1 bit0 Relay 2 bit1 Fault Relay bit2 Horn Relay bit3 Alarm LED flashing status bit 1 indicates LED is flashing Acknowledge clears all to 0 Type Channel First Last Read FC Write FC Notes Alarm 1 Status 1 16 31049 n a 4 n a packed 16bit integer Alarm 2 Status 1 16 31050 n a 4 n a packed 16bit integer Alarm 3 Status 1 16 31051 n a 4 n a packed...

Страница 40: ...r 1000 for decimal position of 1 2 3 or 4 where 123 4 would return the value 10 Type Channel First Last Read FC Write FC Notes Zero Real 1 16 41001 16 n a 4 n a zero real w o decimal point Zero DP 1 16 41017 32 n a 4 n a zero real divisor Span Real 1 16 41033 48 n a 4 n a span real w o decimal point Span DP 1 16 41049 64 n a 4 n a span real divisor Alarm 1 Real 1 16 41065 80 n a 4 n a alarm 1 real...

Страница 41: ...eiving an update command from the Modbus master Type Channel First Last Read FC Write FC Notes Update MIN n a 2065 n a n a 5 Moves 24 hour MIN data trend to trend data base Update AVG n a 2066 n a n a 5 Moves 24 hour MIN data trend to trend data base Update MAX n a 2067 n a n a 5 Moves 24 hour AVG data trend to trend data base This update requires several seconds Therefore a data ready register is...

Страница 42: ...HA71 Digital Gas Controllers Revision 3 04 08 HA71 Technical Manual 6 1 6 HA71 Enclosures ...

Страница 43: ...I O option PCB s such as analog input or discrete relays may be mounted directly to the back of the enclosure Additional 8 channel I O option PCB s must be located external from the assembly on another mounting plate A 3 foot length of I2 C cable is also supplied for this purpose Weight is approximately 7 pounds Properly ground the enclosure and follow national and local electrical codes RACK PANE...

Страница 44: ...ng outdoors but an above mounted weather deflector shield is recommended Weight is approximately 17 pounds Figure 6 3 provides important warning information concerning correct grounding procedures for non metallic enclosures Conduit entries are not provided so installers may place entries as needed Bottom or lower side areas are recommended Care must be taken to avoid drilling into circuit boards ...

Страница 45: ...HA71 Digital Gas Controller Revision 3 04 08 HA71 Technical Manual 6 4 Figure 6 3 Grounding Illustration ...

Страница 46: ...y be mounted inside this enclosure It is suitable for mounting outdoors but an above mounted weather deflector shield is recommended Weight is approximately 30 pounds Conduit entries are not provided so installers may place entries as needed Bottom or lower side areas are recommended Care must be taken to avoid drilling into circuit boards mounted inside the enclosure Properly ground the enclosure...

Страница 47: ... or discrete relays may be mounted inside this enclosure It is suitable for mounting outdoors but an above mounted weather deflector shield is recommended Weight is approximately 110 pounds Properly ground the enclosure and follow national and local electrical codes VERTICAL POSTIION 7 W 21 H HORIZONTAL POSITION 15 W 12 7 8 H MOUNTING HOLE DIMENSION FRONT VIEW HORIZ POSITION MOUNTING FOOT TYP 4 PL...

Страница 48: ...and 4 more with the 10 0180 expansion plate since in panel rack mount installations 10 0180 s must be mounted in user space behind panels or inside racks multiple 10 0180 s may be incorporated to support the required option positions HA71N4 NEMA 4X Wall Mount supports 3 option positions as standard and 4 more with the 10 0180 expansion plate If more than 7 option positions are required the 10 0178...

Страница 49: ...HA71 Digital Gas Controller Revision 3 04 08 HA71 Technical Manual A 1 A Ordering Information ...

Страница 50: ...71SM within a rack panel mount enclosure HA71PM a NEMA 4X enclosure HA71N4 and a NEMA 7 explosion proof enclosure HA71XP is also available These package styles support addition of up to 7 I2 C option board footprints 3 w o 10 0180 expansion kit option inside each respective enclosure Additional options may be added externally or in other enclosures Optional Equipment cont d PART NUMBER DESCRIPTION...

Страница 51: ...NUMBER DESCRIPTION 10 0142 Replacement Main I O PCB for HA71 XP 10 0213 Replacement Main I O PCB for HA71PM ribbon connector on circuit side of PCB 10 0006 Replacement HA71 Flat Panel LCD Module 10 0185 Replacement Complete HA71 Front Panel Electronic Nest Assembly Including 10 0006 LCD 10 0231 Replacement HA71 Front Graphic Panel 1000 1992 Replacement Main I O PCB Fuse 20mm 2 amp 0010 1110 I2 C 6...

Страница 52: ...HA71 Digital Gas Controller Revision 3 04 08 HA71 Technical Manual B 1 B Warranty ...

Страница 53: ...under these warranty terms Time and travel expenses for on site warranty services will be charged at Honeywell Analytics normal billing rates Contact your Honeywell Analytics representative for information on available Service Contracts Honeywell Analytics shall not be liable for any loss or damage whatsoever or howsoever occasioned which may be a direct or indirect result of the use or operation ...

Страница 54: ...oneywell com www honeywell com Asia Pacific Honeywell Analytics 508 Kolon Science Valley l 187 10 Guro Dong Guro Gu Seoul 152 050 Korea Tel 82 2 2025 0300 Fax 82 2 2025 0329 analytics ap honeywell com 1998M0650 Rev 3 April 2008 2008 Honeywell Analytics Please Note While every effort has been made to ensure accuracy in this publication no responsibility can be accepted for errors or omissions Data ...