The Sampler 

 

Activating the Sampler   

Activating the Sampler 

  To activate the sampler, press 

C

All 3 LEDs will flash and the audible alarm will emit 4 beeps. 
When the sampler is activated for the first time, it will enter 
calibration mode. See the 

Calibration

 section. 

 

Normal Operation 

During normal operation the green 

ON

 LED will be lit and the 

motor will be running.  

 

Operation 

Guidelines 

When operating the sampler, adhere to the following 
guidelines: 

  Confirm that all hoses are properly attached to the 

sampler 

  Confirm that there are no leaks in the system before 

obtaining a sample. 

For the sample to reach the detector, allow approximately 3 
seconds per 1 ft./0.3 m of hose. Consult your detector 
manual for detector response time. 

www.

optimum

stores

.com/bw

Shop for BW products online at:

Содержание GA-SP01

Страница 1: ...www optimumstores com bw Shop for BW products online at ...

Страница 2: ...uyer promptly notifying BW of any defect and if required promptly making the product available for correction No goods shall be returned to BW until receipt by the buyer of shipping instructions from BW and c the right of BW to require that the buyer provide proof of purchase such as the original invoice bill of sale or packing slip to establish that the product is within the warranty period THE B...

Страница 3: ... Tube Connections 10 Block the Airflow 11 Deactivating the Sampler 11 Alarms 12 Low Flow Alarm 12 Low Battery Alarm 13 System Fault Alarm 13 Calibration 14 Calibration Pass 14 Calibration Fail 14 Maintenance 15 Replacing the Batteries 15 Replacing a Filter 17 Draining the Water Trap 18 If the Sampler Does Not Work 19 Replacement Parts and Accessories 21 Specifications 22 i www optimumstores com bw...

Страница 4: ...The Sampler Users Manual ii www optimumstores com bw Shop for BW products online at ...

Страница 5: ...pler 6 3 LED Descriptions 7 4 Pushbuttons 8 5 Standard Hose Configuration 10 6 Replacing the Batteries 16 7 Replacing a Filter 17 8 Draining the Water Trap 18 9 Troubleshooting Tips 19 10 Replacement Parts and Accessories 21 iii www optimumstores com bw Shop for BW products online at ...

Страница 6: ...The Sampler Users Manual iv www optimumstores com bw Shop for BW products online at ...

Страница 7: ...6 2 LED Descriptions 7 3 Standard Hose Configuration 10 4 Sample Tube Connections 10 5 Replacing the Batteries 16 6 Replacing a Filter 17 7 Draining the Water Trap 18 8 Rear Assembly Screws 20 v www optimumstores com bw Shop for BW products online at ...

Страница 8: ...L ONLY READ AND UNDERSTAND THE USER MANUAL COMPLETELY BEFORE OPERATING OR SERVICING The Sampler Order Number Description GA SP01 Motorized sampling pump with rigid probe and sample tubing GA SP02 Motorized sampling pump with rigid probe and sample tubing Europe vi www optimumstores com bw Shop for BW products online at ...

Страница 9: ...pot sampling device that can be used with most BW single and multiple gas detectors An internal motorized pump transports potentially hazardous gasses from a sample area to a detector The sampler is an accessory for a personal safety device It is your responsibility to respond properly to alarm conditions 1 www optimumstores com bw Shop for BW products online at ...

Страница 10: ...us on the World Wide Web www gasmonitors com ISO 9001 Safety Information Read First Use the sampler only as specified in this manual otherwise the protection provided by the sampler may be impaired International symbols used on the sampler and in this manual are explained in Table 1 Read the Caution statements on the following pages before using the sampler ecNote This instrument contains batterie...

Страница 11: ... using the recommended charger only Do not use any other charger Failure to observe this precaution could lead to fire or explosion Do not change or charge batteries in a hazardous location Doing so will impair the intrinsic safety of the unit and may lead to fire or explosion Read and observe all instructions and precautions in the literature provided with the charger Failure to do so may result ...

Страница 12: ...Meaning n Approved to both U S and Canadian Standards by the Canadian Standards Association ATEX Conforms to European ATEX directives g European Explosive Protection X Conforms to European Union directives 4 www optimumstores com bw Shop for BW products online at ...

Страница 13: ...e 4 65 in 11 81 cm Medium sample tube 10 25 in 26 00 cm Sampling hose 10 ft 3 m Blue stone coarse particulate filter attached Attachment hose 3 ft 1m Particulate filter Qty 2 Quick reference card User Manual To order replacement parts see the section Replacement Parts and Accessories The sampler is shipped with the short sample tube installed The Operation section describes how to install another ...

Страница 14: ...he Sampler Table 2 The Sampler Item Function A Short sample tube inlet B Block trigger C Water Trap D Drain plug E Activation button C F LEDs G Audible alarm H Cowl I Tube fitting outlet J Filter plug 6 www optimumstores com bw Shop for BW products online at ...

Страница 15: ...criptions Item Function A The green ON LED indicates the sampler is operating properly B The yellow BLOCK LED prompts you to block the airflow C The red FAIL LED indicates that the sampler is in an error condition 7 www optimumstores com bw Shop for BW products online at ...

Страница 16: ... There will be a beep the sampler will turn off and upon the next startup the sampler will enter calibration mode To turn off the sampler press C and hold for 2 seconds BLOCK AIRFLOW To block airflow through the sampler firmly depress the block trigger Note Airflow can also be blocked by temporarily kinking the hose or covering the end of the rigid tube 8 www optimumstores com bw Shop for BW produ...

Страница 17: ...al operation the green ON LED will be lit and the motor will be running Operation Guidelines When operating the sampler adhere to the following guidelines Confirm that all hoses are properly attached to the sampler Confirm that there are no leaks in the system before obtaining a sample For the sample to reach the detector allow approximately 3 seconds per 1 ft 0 3 m of hose Consult your detector m...

Страница 18: ...The sampler is shipped with the short sample tube installed To remove the installed sample tube grasp the end of the tube and pull straight out Insert the new sample tube through the enclosure and into the gasket opening To ensure a proper seal the end of the tubing should be inserted until firm resistance is felt Figure 4 Sample Tube Connections 10 www optimumstores com bw Shop for BW products on...

Страница 19: ...m beeps while the airflow is blocked then the sampler is operating correctly If it does not reconnect the hose tube to check for leaks Note We recommend that you perform a block test once the sampler is activated Deactivating the Sampler To turn off the sampler press C and hold for 2 seconds The audible alarm beeps 3 times all LEDs light up and then the sampler turns off Note If you reactivate the...

Страница 20: ...pproximately 10 seconds This gives you time to remove any blockage in the line If the condition does not change the sampler will prompt you for a block If the sampler detects the expected airflow it returns to normal operation If the sampler does not detect the expected airflow it returns to the low flow alarm state until the airflow recovers 12 www optimumstores com bw Shop for BW products online...

Страница 21: ...il you replace the batteries or the battery power nears depletion If the battery voltage drops too low the audible alarm beeps once the red FAIL LED lights and the sampler turns off See the section Replacing the Batteries System Fault Alarm If the sampler detects a system fault the red FAIL LED will light and the audible alarm will beep Press C to clear the alarm The sampler will automatically ent...

Страница 22: ...ep the sampler will turn off and upon the next startup the sampler will enter calibration mode Calibration Pass When the sampler passes calibration the audible alarm emits 3 beeps and the sampler enters normal operation mode Calibration Fail If the sampler fails calibration the audible alarm emits 6 beeps the red FAIL LED lights and the sampler turns off See the section If the Sampler Does Not Wor...

Страница 23: ...ce the batteries as soon as the sampler indicates a Low Battery Alarm Use only batteries recommended by BW Technologies to prevent damage or personal injury and maintain intrinsic safety Use only approved AA alkaline or charged AA NiMH batteries properly installed in the sampler See the Specifications section for approved batteries Charge batteries using only a recommended charger Do not use any o...

Страница 24: ...m the body of the sampler 3 Rotate the cowl upwards 4 Remove the two batteries 5 Install new batteries observing polarity 6 Attach the cowl 7 Rotate the cowl downward until it snaps into position Tighten the 4 40 x 3 8 button head screw Table 6 Replacing the Batteries Item Description A 4 40 x 3 8 button head screw may not be included in certain regions B Cowl C Batteries 2 1 3 Figure 5 Replacing ...

Страница 25: ...hat the filter is firmly seated into the plug and the open end of the filter is facing out 3 Insert the filter plug assembly into the sampler ensuring the filter enters the gasket filter opening Table 7 Replacing a Filter Item Description A Filter plug B Filter C Open end of filter D Filter plug assembly E Gasket filter opening 2 1 3 5 4 Figure 6 Replacing a Filter 17 www optimumstores com bw Shop...

Страница 26: ...need to be emptied Pull open the drain plug and shake any water out of the water trap Replace the plug when complete Table 8 Draining the Water Trap Item Function A Water trap B Drain plug 1 2 Figure 7 Draining the Water Trap 18 www optimumstores com bw Shop for BW products online at ...

Страница 27: ...lem Possible Cause Solution Sampler does not turn on No batteries Depleted batteries Damaged or defective sampler Install batteries Replace batteries Contact BW see page 2 Continuous low flow alarm System fault alarm Extreme environmental change has formed condensation in the pump mechanism Wait approximately 15 minutes for the condensation in the sampler to clear Sampler automatically turns off A...

Страница 28: ... are loose Pump has failed Activate the sampler During calibration block the airflow of the sampler when prompted to by the yellow BLOCK LED Empty the water trap Remove the filter plug assembly and orient the filter so the open end is facing away from the filter plug Re insert the filter plug assembly Replace the filter Install the filter assembly and drain plug Tighten rear assembly screws as per...

Страница 29: ...arger and 4 AA NiMH batteries 1 GAMIC C01 K xx 220 VAC 4 port charger and 4 AA NiMH batteries 1 SP T2 S Standard rigid sample tube with barbed fitting 4 65 in 11 81 cm 1 SP T2 M Standard rigid sample tube with barbed fitting 10 25 in 26 00 cm 1 SP T1 2 Rigid sample probe ft 0 3m D4 PF 1 Blue stone coarse particulate filter 1 Model No Description Qty M1806 Flexible sampling hose ft 0 3m SP PF 1 Sam...

Страница 30: ...1500 150 C Energizer E91 163 C NiMH rechargeable Quest Platinum HGAAC1800G 211 4 C Quest HG1600AACS 199 C Energizer NH15 205 C Maha Powerex 1700 mAh MH AA170 192 C Maha Powerex 1800 mAh MH AA180 201 1 C Yuasa Delta 1300 mAh DHA1400AA 209 C Yuasa Delta 1500 mAh DHA1600AAC 204 4 C Worst Case Temperature Code Alkaline 163 C NiMH 212 C Ambient 20 C Ta 50 C Europe Alkaline Temperature Code Duracell MN1...

Страница 31: ...www optimumstores com bw Shop for BW products online at ...