Electrical installation
Page 12 Installation guide 854 ATG
External fusing
The 854 ATG is internally fused on secondary side of transformer (fuses are located on GPS
board). Therefore, external fuses must be installed in the mains supply cable to each 854 ATG.
mains voltage
fuse value (in accordance with IEC 127-2-3)
220 Vac or 230 Vac
315 mA slow
110 Vac or 130 Vac
630 mA slow
65 Vac
1 A slow
Cable glands / conduit
Cable glands:
Explosion-proof (Ex-d) or increased safe (Ex-e) cable glands must be used, depending on local
requirements. Refer to the Ex-marking on the identification label of the 854 ATG to determine
which type of cable glands is required.
Note:
Mount the glands according to the supplier
=
s instructions.
Conduit:
If the 854 ATG is installed in a hazardous area, threaded rigid metal conduit or threaded steel
intermediate metal conduit shall be used.
Note:
If the 854 ATG is installed in a hazardous area, stopper boxes
must
be applied within 18
inches (0.45 m) from the 854 ATG to seal the cabling in the conduit.
Depending on the wiring configuration, one to four
:@
NPT threaded cable glands (or rigid
conduits) may be required with the 854 ATG.
Note:
Seal the unused cable inlets with an approved
:@
NPT threaded stopping plug.
Warning
Improper installation of cable glands, conduit or stopping plugs will invalidate
the Ex approval of your 854 ATG.
Содержание Enraf 854 ATG
Страница 18: ...Appendix Page 18 Installation guide 854 ATG Appendix A Dimensions 854 ATG medium pressure and chemical version ...
Страница 19: ...Appendix Installation guide 854 ATG Page 19 Dimensions 854 ATG high pressure version ...
Страница 20: ...Appendix Page 20 Installation guide 854 ATG Appendix B Available adapters calibration chambers ...
Страница 23: ...Notes Installation guide 854 ATG Page 23 ...