Self-Test Pass
After the self-test has been completed and any alarms have
been acknowledged, the detector begins normal operation.
The LCD then displays the current ambient gas readings.
The detector begins recording immediately. It records the
•
maximum (MAX) gas exposure,
•
records the short-term exposure levels (STEL), and
•
calculates the time-weighted average (TWA)
exposures.
When the air is at acceptable levels, and if the safe option is
enabled,
SAFE
scrolls across the screen continually while in
normal operating mode.
SAFE
does not display while
functions are being performed.
Note
This option is disabled in stealth mode.
Self-Test Fail
If the detector fails the self-test, refer to
Battery Test
The detector tests the batteries upon activation and
continuously thereafter. The battery power icon
I
displays
at all times while in normal operation mode. If the battery
voltage is low, the detector displays the low battery icon
LOW
and activates the low battery alarm (beeps every 10
seconds if the confidence beep is disabled).
If the confidence beep option is enabled, the confidence
beep stops if the battery power is low. For more information,
refer to
.
If the battery voltage is extremely low, the detector performs
an automatic shutdown.
Datalogger Operation
(optional)
The datalogger operation is automatic and requires no
settings.
When the MultiMediaCard (MMC) is inserted in the detector,
the MMC card icon
S
displays continuously while in normal
operation mode.
14
Содержание BW GasAlertMicro
Страница 6: ...GasAlertMicro User Manual iv ...
Страница 8: ...GasAlertMicro User Manual vi ...
Страница 71: ...GasAlertMicro Maintenance Figure 7 Replacing a Sensor or Sensor Filter 61 ...
Страница 82: ...GasAlertMicro User Manual 72 ...