APT4000PH
80
Relay test (manual test of contacts)
• Press
CONF
, enter code 5557
The relays are frozen. This state is indicated in the display.
The 4 digits in the display correspond to the 4 relays (as on
terminal plate):
1st digit: R1
2nd digit: R2
3rd digit: AL
4th digit: CLN
Function test using arrow keys – see left column.
When exiting the function (
ENTER
), the relays are set corre-
sponding to the measured value.
Select a relay
Specify current at output 1
• Press
CONF
, enter code 5555
The current indicated in the main display for output 1 can
be edited.
Select with
key,
edit number with
key,
proceed with
ENTER
The actually measured current is shown in the secondary
display. The APT4000PH is in Hold mode.
Press
ENTER
to return to measurement (Hold remains
active for another 20 sec).
Diagnostics functions
These functions are used for testing the connected peripherals.
Display
Action / Remarks
Specify current at output 2
• Press
CONF
, enter code 5556
The current indicated in the main display for output 2 can
be edited.
Select with
key,
edit number with
key,
proceed with
ENTER
The actually measured current is shown in the secondary
display. The APT4000PH is in Hold mode.
Press
ENTER
to return to measurement.
Test 0/1
Return to
measurement
81
The arrows indi-
cate which relay
(valve) is active:
Controller
characteristic
+100 %
-100 %
Relay 2 active
(Meas. value >
setpoint)
Relay 1 active
(Meas. value <
setpoint)
Setpoint
Controller output
Momentary controller output (adjusted value has not been
stored yet)
Controller test (manual specification of controller
output)
• Press
CONF
, enter code 5559
After function activation “Ctrl” is displayed for approx.
3 sec.
With controller turned off, “OFF” is displayed in addition,
then return to measuring mode.
The function is used to start up control loops or check the
actuators.
For bumpless changeover to automatic operation
(exiting this function), configure an I-action component
(reset time).
Specify value:
Select with
key,
edit number with
key,
proceed with
ENTER
APT4000PH is in Hold mode.
Press
ENTER
to return to measurement
(Hold remains active for another 20 sec).
Controller output -100 to 0 %:
Relay 2 active
Controller output 0 to +100 %:
Relay 1 active
Display
Action / Remarks