TROUBLESHOOTING

If any abnormality arises during use, please check the following points:

Note:  

  

If the unit still does not work, contact HoMedics Consumer Relations. Under no circumstance 
should you disassemble or attempt to repair the unit by yourself. Contact information for 
HoMedics Consumer Relations Department can be found on the warranty page.

SYMPTOMS

POSSIBLE  

CAUSES

CORRECTION

Unit does not turn on when 
the START/STOP button is 
pushed.

Batteries have run down.

Replace them with four new “AA” alkaline 
batteries.

Battery polarities (+/-) have been posi-
tioned incorrectly.

Re-insert the batteries in the correct posi-
tions. See page 11.

EE measurement error symbol 
shown on display or the blood 
pressure value is displayed 
excessively low (or high).

The UltraSoft™ Upper Arm Cuff has been 
positioned on the arm incorrectly.

Re-wrap the cuff properly so that it is posi-
tioned correctly. Take new measurement. 

Did you talk or move during measurement? Keep arm steady during measurement. 

Measure again. Refer to “Measurement 
Procedure” instructions, see page 21. 

Shaking of the arm with the cuff on.

E1 error symbol shown on display. Air circuit abnormality. Cuff tube may 

not be plugged into monitor correctly.

Check cuff connection. See page 18. 
Take new measurement.

E2 error symbol shown on display. Inflation pressure exceeding 300 

mmHg.

Switch the unit off and then turn back 
ON. Take new measurement.

E3 error symbol shown on display. Error determining measurement data.

Re-wrap the cuff properly so that it is posi-
tioned correctly. Take new measurement.

E4 - Data transmission error 

symbol shown on the display.

Bluetooth function is turned OFF, on 

your mobile device.

Turn Bluetooth ON, on your mobile 

device, and try again.

Bluetooth function is turned OFF, on 

the blood pressure monitor.

Turn the Bluetooth function ON, on your 

blood pressure monitor (see page 15), 

then try again. 

Mobile device does not pair with the 

blood pressure monitor.

Make sure Bluetooth is ON, on the mobile 

device and the blood pressure monitor, 

and try again.
Requires compatible mobile device.  This 

blood pressure monitor is compatible 

with mobile devices that are running iOS 

5.0 or newer, or Android

TM

 4.3 and newer.

For a complete list, go to  

www.walgreens.com/connectapp.

The App on the mobile device is not open. Make sure the app is open on the mobile 

device, and try again.

The blood pressure monitor and mobile 

device are out of transmitting range.

Make sure the mobile device and blood 

pressure monitor are within the acceptable 

range of 32 feet (10 meters).

Unexpected loss of electrical/

mechanical integrity.

Make sure your last reading is stored in

memory and the App is open and try 

again. See page 26.
Remove batteries, re-insert, and try again.
Return the device to your local distributor 

or importer.

EP error symbol shown on 

display.

System Error.

Remove batteries. After 1 minute, re-insert the 

batteries.  Take new measurement.

Power Source:

(4) 1.5V DC “AA” alkaline batteries or 6V DC AC 

adapter 

Measurement Method:

Oscillometric

Measurement Range:

Pressure: 40~280 millimeters Mercury (mmHg) 

Pulse: 40~199 beats/minute

Accuracy:

Pressure: ±3 mmHg

Pulse: ±5% of reading

Pressure Sensor:

Semi-conductor

Inflation:

Pump driven

Deflation:

Automatic Pressure release valve

Memory Capacity:

120 memories for each User (240 total)

Operation 

Environment:

Temperature: 41°F~104°F (5°C ~40°C)

Humidity: 15% ~ 93% RH

Storage/Transportation 

Environment:

Temperature: -13°F~158°F (-25°C~70°C)

Humidity: Less Than 93% RH

Weight:

0.74 lb (330 g) (without batteries) 

Arm Circumference:

Ultrasoft™ Upper Arm Cuff Size: 9”- 17”  

(23-43cm)

Dimensions:

4.65”(L) x 6.46”(W) x 1.89”(H) 

118mm(L) x 164mm(W) x 48mm(H)

Accessories:

(4) “AA” alkaline batteries, 6V DC AC adapter, 

instruction manual, 1 arm cuff with tube
Internally Powered equipment

Type BF Equipment

IP22- Degrees of protection provided by  

enclosures.

Not suitable for use in presence of flammable 

anesthetic mixture with air, oxygen or  

nitrous oxide. 

Continuous operation with short-time loading.
Follow instructions for use.

SPECIFICATIONS

Note:

  These specifications are subject to change without notice.

Содержание Walgreens Premium Arm WGNBPA-960BT

Страница 1: ...PREMIUM ARM Blood Pressure Monitor 5 Year Limited Warranty WGNBPA 960BT El manual en espa ol empieza en la p gina 46...

Страница 2: ...EMC 38 Troubleshooting 41 Specifications 42 5 Year Limited Warranty 43 TABLE OF CONTENTS INTRODUCTORY INFORMATION Important Product Notices and Safety Instructions 3 About Blood Pressure 5 Blood Press...

Страница 3: ...red from a stroke Excessive use may result in blood overflow interference which is likely to cause uncomfortable sensations such as partial subcutaneous hemorrhage or temporary numbness to your arm In...

Страница 4: ...the arteries The pressure measured when the heart contracts and sends blood out of the heart is systolic highest blood pressure The pressure measured when the heart dilates with blood flowing back int...

Страница 5: ...matically finds where your measurement results fall on the NIH s National Heart Lung and Blood Institute s table and provides a cue if your reading falls into one of the stages that could potentially...

Страница 6: ...riple Check Function ON The REST symbol will flash among 3 measurements during countdown Memory Average Displays average of last 3 readings AM PM Averaging Indicates the reading being displayed is an...

Страница 7: ...tery compartment according to the indications inside the compartment 3 Close the battery cover by pushing in the top end of the battery door Replace the batteries if The weak battery symbol appears Wh...

Страница 8: ...buttons on the monitor If setting the date time by syncing with your mobile device it is important that this is done prior to taking any measurements to ensure the date time are accurate To set using...

Страница 9: ...connection with the monitor Bluetooth Function OFF Bluetooth Function ON Triple Check Rest Assure ON ON OFF OFF USING REST ASSURE FUNCTION This blood pressure monitor features the Rest Assure Functio...

Страница 10: ...hown below A R R O W M U S T FA LL W IT H IN B LU E B A N D Solid Colored Line Colored Arrow BLUE BLUE A R R O W M U S T FA LL W IT H IN B LU E B A N D D ring End of cuff AR RO W MUS T FA LL WITHIN BL...

Страница 11: ...8 Lay your arm on a table palm upward so the cuff is at the same height as your heart Make sure the tube is not kinked Note This device should not be used when your arm is injured or wounded If it is...

Страница 12: ...body temperature During Measurement Do not talk or move your arm or hand muscles Do not cross your legs Sit with feet flat on the floor Do not touch cuff or monitor during measurement If you are using...

Страница 13: ...ly after a button is pressed Using Triple Check Measurement 1 Please make sure the Triple Check switch located on the back of the unit is in the ON position 2 Press the USER SELECT button After the us...

Страница 14: ...s Connect App that is free to download on the App StoreSM and on Google Play Make sure you have the Walgreens Connect App downloaded and open on your mobile device before trying to transmit your blood...

Страница 15: ...efore taking measurements Measurements transmitted to the App cannot be edited For a complete list of compatible mobile devices please visit www walgreens com connectapp Fig 2 EARN BALANCE REWARDS POI...

Страница 16: ...ected during measurement Usually this is not a cause for concern However if the symbol appears often we recommend you seek medical advice Please note that the device does not replace a cardiac examina...

Страница 17: ...the button to review individual readings that make up the Triple Check average Note A Triple Check average and the 3 measurements that make this average count as 4 readings when stored in memory To us...

Страница 18: ...our physiology The measurement results that users receive are for reference only If users have any blood pressure concerns please consult a physician Once inflation reaches 300 mmHg the unit will defl...

Страница 19: ...s contact information can be found on the warranty page Do not immerse the unit in water as this will result in damage to the unit Do not subject the monitor or cuff to extreme temperatures humidity m...

Страница 20: ...ELECTROMAGNETIC COMPATIBILITY EMC Guidance and manufacturer s declaration electromagnetic emissions The device is intended for use in the electromagnetic environments listed below and should only be...

Страница 21: ...he transmitter Recommended separation distance d 1 2 d 1 2 80 MHz to 800 MHz d 2 3 800 MHz to 2 5 GHz where P is the maximum output power rating of the transmit ter in watts W according to the transmi...

Страница 22: ...tor Make sure Bluetooth is ON on the mobile device and the blood pressure monitor and try again Requires compatible mobile device This blood pressure monitor is compatible with mobile devices that are...

Страница 23: ...modifications or adop tion to enable it to operate in any other country than the country for which it was designed manufactured approved and or authorized or repair of products damaged by these modifi...

Страница 24: ...45 DE PRIMERA CALIDAD PARA EL BRAZO Monitor de presi n arterial Garant a limitada por 5 a os WGNBPA 960BT...

Страница 25: ...icaciones 88 Garant a limitada por 5 a os 89 NDICE INFORMACI N INICIAL Avisos importantes del producto e instrucciones de seguridad 49 Acerca de la presi n arterial 51 Valores est ndar de la presi n a...

Страница 26: ...cular El uso excesivo puede provocar interferencia del flujo sangu neo lo que probablemente produzca sensaciones inc modas como hemorragia parcial subcut nea o adormecimiento temporal del brazo En gen...

Страница 27: ...fluye por las arterias La presi n que se mide cuando el coraz n se contrae y env a la sangre es la presi n arterial sist lica la m s alta La presi n que se mide cuando el coraz n se dilata y la sangre...

Страница 28: ...ran sus resultados de medici n en la tabla del Instituto Nacional del Coraz n los Pulmones y la Sangre de los NIH y brinda una gu a si su lectura entra en una de las etapas que podr a indicar un riesg...

Страница 29: ...ra es un promedio de las ltimas 3 mediciones de la ma ana o las ltimas 3 mediciones de la noche S mbolo Multi Read varias lecturas Aparece cuando est encendido el sistema de medici n de triple control...

Страница 30: ...er odos prolongados Limpie los contactos de las pilas y del compartimiento de las pilas con un pa o seco suave cada vez que instale las pilas Las bater as son desechos peligrosos No las deseche junto...

Страница 31: ...izando los botones SET configuraci n en el monitor Si configura la fecha y hora sincroniz ndolas con su dispositivo m vil es importante que esto se haga antes de tomar cualquier medici n para asegurar...

Страница 32: ...vil para establecer una conexi n con el monitor Funci n de Bluetooth APAGADA Funci n de Bluetooth ENCENDIDA USO DE LA FUNCI N REST ASSURE DESCANSE TRANQUILO Este monitor de presi n arterial cuenta con...

Страница 33: ...ue se muestra a continuaci n A R R O W M U S T FA LL W IT H IN B LU E B A N D BLUE BLUE L nea de color s lido Flecha de color 1 Si el brazalete no est armado pase el extremo del brazalete que se encue...

Страница 34: ...8 Apoye el brazo en una mesa con la palma hacia arriba de modo que el brazalete quede a la misma altura que el coraz n Aseg rese de que el tubo no est plegado Nota El dispositivo no debe ser utilizado...

Страница 35: ...de la mano No cruce las piernas Si ntese con los pies planos apoyados en el piso No toque el brazalete ni el monitor durante la medici n Si usted est usando este monitor de presi n arterial por primer...

Страница 36: ...riple control 1 Aseg rese de que el interruptor de medici n de Triple Check triple control ubicado en la parte trasera de la unidad est en posici n ON encendido 2 Presione el bot n de SELECCI N DE USU...

Страница 37: ...de descargarse de forma gratuita en App Store y Google Play Aseg rese de haber descargado la aplicaci n Walgreens Connect App y brala en su dispositivo m vil antes de intentar transmitir sus medicione...

Страница 38: ...aci n no se pueden modificar Para obtener una lista completa de los dispositivos m viles compatibles visite www walgreens com connectapp Fig 2 GANE PUNTOS BALANCE REWARDS Gane puntos Balance Rewards p...

Страница 39: ...es aparece un cono que lo indica Generalmente esto no es causa de preocupaci n Sin embargo si el s mbolo aparece con frecuencia recomendamos que busque asesoramiento m dico Tenga en cuenta que este di...

Страница 40: ...e el bot n para revisar las lecturas individuales que forman el promedio de triple control Nota Un promedio de triple control y las 3 mediciones que forman este promedio se cuentan como 4 lecturas cua...

Страница 41: ...de las mediciones que los usuarios obtienen son nicamente para referencia Si los usuarios tienen inquietudes en cuanto a la presi n arterial deber n consultar a un m dico Una vez que la unidad se inf...

Страница 42: ...laciones con el consumidor de HoMedics encontrar la informaci n de contacto en la p gina de la garant a No sumerja la unidad en agua ya que puede da arla No exponga el monitor ni el brazalete a temper...

Страница 43: ...en los entornos electromagn ticos que se detallan a continuaci n y nicamente en tales entornos Prueba de emisiones Cumplimiento Gu a sobre entornos electromagn ticos Emisiones de RF CISPR 11 Grupo 1...

Страница 44: ...etos y personas Gu a y declaraci n del fabricante respecto de la inmunidad electromagn tica Se espera que el dispositivo se utilice en los entornos electromagn ticos que se detallan a continuaci n y n...

Страница 45: ...rial Aseg rese de que el Bluetooth est activado en el dispositivo m vil y en el monitor de presi n arterial e int ntelo de nuevo Se requiere un dispositivo m vil compatible Este monitor de presi n art...

Страница 46: ...ual se compr el producto Un producto que requiera modificaciones o adaptaci n para habilitar su funcionamiento en cualquier pa s que no sea aqu l para el que fue dise ado fabricado aprobado y o autori...

Отзывы: